Agreement between cardiac output estimation with a wireless, wearable pulse decomposition analysis device and continuous thermodilution in post cardiac surgery intensive care unit patients
Journal of Clinical Monitoring and Computing · 2024
DOI: 10.1007/s10877-023-01059-5 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Monitoring and Computing, 2024. Verfasst von Ashish K Khanna und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology, Section on Critical Care Medicine, School of Medicine, Wake Forest University, Atrium Health Wake Forest Baptist Medica…. Hauptthemen laut MEDLINE: Pulmonary Artery, Cardiac Surgical Procedures. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 31 Literaturangaben. Gefördert von: NCATS NIH HHS.
Themen: Intensivpflege
Abstract (Originalsprache)
PURPOSE: Pulse Decomposition Analysis (PDA) uses integration of the systolic area of a distally transmitted aortic pulse as well as arterial stiffness estimates to compute cardiac output. We sought to assess agreement of cardiac output (CO) estimation between continuous pulmonary artery catheter (PAC) guided thermodilution (CO-CCO) and a wireless, wearable noninvasive device, (Vitalstream, Caretaker Medical, Charlottesville, VA), that utilizes the Pulse Decomposition Analysis (CO-PDA) method in postoperative cardiac surgery patients in the intensive care unit. METHODS: CO-CCO measurements were compared with post processed CO-PDA measurements in prospectively enrolled adult cardiac surgical intensive care unit patients. Uncalibrated CO-PDA values were compared for accuracy with CO-CCO via a Bland-Altman analysis considering repeated measurements and a concordance analysis with a 10% exclusion zone. RESULTS: 259.7 h of monitoring data from 41 patients matching 15,583 data points were analyzed. Mean CO-CCO was 5.55 L/min, while mean values for the CO-PDA were 5.73 L/min (mean of differences +- SD 0.79 ± 1.11 L/min; limits of agreement - 1.43 to 3.01 L/min), with a percentage error of 37.5%. CO-CCO correlation with CO-PDA was moderate (0.54) and concordance was 0.83. CONCLUSION: Compared with the CO-CCO Swan-Ganz, cardiac output measurements obtained using the CO-PDA were not interchangeable when using a 30% threshold. These preliminary results were within the 45% limits for minimally invasive devices, and pending further robust trials, the CO-PDA offers a noninvasive, wireless solution to complement and extend hemodynamic monitoring within and outside the ICU.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anesthesiology, Section on Critical Care Medicine, School of Medicine, Wake Forest University, Atrium Health Wake Forest Baptist Medical Center, Winston-Salem, NC, USA. akhanna@wakehealth.edu.
- Outcomes Research Consortium, Cleveland, OH, USA. akhanna@wakehealth.edu.
- Perioperative Outcomes and Informatics Collaborative (POIC), Winston-Salem, NC, USA. akhanna@wakehealth.edu.
- Wake Forest University School of Medicine, Winston-Salem, NC, USA.
- Perioperative Outcomes and Informatics Collaborative (POIC), Winston-Salem, NC, USA.
- Department of Anesthesiology, Wake Forest University School of Medicine, Atrium Health Wake Forest Baptist Medical Center, Winston-Salem, NC, USA.
ORCID-Profile: Ashish K Khanna
Förderung
- NCATS NIH HHS
Zitieren
APA
Khanna, A. K., Garcia, J. O., Saha, A. K., Harris, L., Baruch, M., & Martin, R. S. (2024). Agreement between cardiac output estimation with a wireless, wearable pulse decomposition analysis device and continuous thermodilution in post cardiac surgery intensive care unit patients. Journal of Clinical Monitoring and Computing. https://doi.org/10.1007/s10877-023-01059-5
Vancouver
Khanna AK, Garcia JO, Saha AK, Harris L, Baruch M, Martin RS. Agreement between cardiac output estimation with a wireless, wearable pulse decomposition analysis device and continuous thermodilution in post cardiac surgery intensive care unit patients. Journal of Clinical Monitoring and Computing. 2024. doi:10.1007/s10877-023-01059-5
BibTeX
@article{Khanna2024,
author = {Khanna, Ashish K and Garcia, Julio O and Saha, Amit K and Harris, Lynnette and Baruch, Martin and Martin, R Shayn},
title = {Agreement between cardiac output estimation with a wireless, wearable pulse decomposition analysis device and continuous thermodilution in post cardiac surgery intensive care unit patients},
journal = {Journal of Clinical Monitoring and Computing},
year = {2024},
doi = {10.1007/s10877-023-01059-5}
}
EUVIMED