EUVIMEDOpdag EUVIMED
Open access · CC BY 4.0

Capturing Current Practices and Characterizing Measurement Reproducibility for Seized Drug Analysis Using Ambient Ionization Mass Spectrometry: An Interlaboratory Study

Edward Sisco iD, Dennis D Leber, Arun S Moorthy iD

Journal of the American Society for Mass Spectrometry · 2025

DOI: 10.1021/jasms.5c00213 ↗

At a glance

Journal article in Journal of the American Society for Mass Spectrometry, 2025. Written by Edward Sisco and 2 co-authors. Institution: National Institute of Standards and Technology, Gaithersburg, Maryland 20899, United States. Major MEDLINE topics: Illicit Drugs, Mass Spectrometry, Spectrometry, Mass, Electrospray Ionization. Cited 4 times so far (Crossref); 22 references. The full text is freely available (open access, CC BY 4.0).

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

The use of ambient ionization mass spectrometry (AI-MS) to aid in the preliminary screening of seized drug evidence has steadily increased over the past two decades. Unlike gas chromatography-mass spectrometry (GC-MS), where electron ionization using a single quadrupole analyzer is commonplace, a wide range of ionization sources and mass spectrometers can be used in AI-MS. Differences in instrument configuration can lead to substantial variability in the mass spectral data obtained. An interlaboratory study, consisting of 35 participants from 17 laboratories, was conducted to begin to understand the landscape and the differences in the data that are produced. Laboratories analyzed a series of 21 solutions across multiple days using their own instrumental methods. Mass spectra were extracted and compared to understand operator, within-lab, and between-lab reproducibility for common compounds and mixtures observed in seized drug analysis. In addition, five participants analyzed the 21 solutions using prescribed method parameters to measure reproducibility improvements when using identical instrumental conditions. Mass spectral reproducibility, measured through pairwise cosine similarity, was found to be generally quite high, regardless of sample type, instrument type, method, or operator. Low-fragmentation spectra showed the lowest variability, as they were dominated by intact protonated molecule peaks. Several potential issues that increased variability were identified, including carryover from mass calibrants, poor sample introduction, and mass spectrometer inlets that required cleaning. The use of uniform method parameters was shown to increase the reproducibility of mass spectra across laboratories, most notably at higher in-source collision-induced dissociation energies. This study provides initial insights into the current landscape of AI-MS in seized drug analysis and lays the foundation for future studies that can provide needed data for the development of documentary standards, standard methods, and possibly the establishment of error rates.

Subjects

MeSH terms (MEDLINE)

Illicit DrugsMass SpectrometrySpectrometry, Mass, Electrospray IonizationReproducibility of ResultsLaboratories

Author keywords

ASAP-MSAmbient Ionization Mass SpectrometryDART-MSInterlaboratory StudySeized DrugsSpectral Reproducibility

Substances

Illicit Drugs

Authors and institutions

  • National Institute of Standards and Technology, Gaithersburg, Maryland 20899, United States.
  • Department of Forensic Science, Trent University, Peterborough, Ontario K9L 0G2, Canada.

ORCID profiles: Edward Sisco, Arun S Moorthy

Cite

APA

Sisco, E., Leber, D. D., & Moorthy, A. S. (2025). Capturing Current Practices and Characterizing Measurement Reproducibility for Seized Drug Analysis Using Ambient Ionization Mass Spectrometry: An Interlaboratory Study. Journal of the American Society for Mass Spectrometry. https://doi.org/10.1021/jasms.5c00213

Vancouver

Sisco E, Leber DD, Moorthy AS. Capturing Current Practices and Characterizing Measurement Reproducibility for Seized Drug Analysis Using Ambient Ionization Mass Spectrometry: An Interlaboratory Study. Journal of the American Society for Mass Spectrometry. 2025. doi:10.1021/jasms.5c00213

BibTeX
@article{Sisco2025,
  author = {Sisco, Edward and Leber, Dennis D and Moorthy, Arun S},
  title = {Capturing Current Practices and Characterizing Measurement Reproducibility for Seized Drug Analysis Using Ambient Ionization Mass Spectrometry: An Interlaboratory Study},
  journal = {Journal of the American Society for Mass Spectrometry},
  year = {2025},
  doi = {10.1021/jasms.5c00213}
}
Start for free →