Oxygen Detection in Dermal Interstitial Fluid with Microneedles Based on Platinum Nanoparticles
ACS Sensors · 2026
DOI: 10.1021/acssensors.5c04533 ↗Kurz gefasst
Fachartikel in der Zeitschrift ACS Sensors, 2026. Verfasst von Qikun Wei und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Chemistry, KTH Royal Institute of Technology, Teknikringen 30, SE-114 28Stockholm, Sweden. Hauptthemen laut MEDLINE: Platinum, Oxygen, Needles, Extracellular Fluid. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ministerio de Ciencia, Innovación y Universidades, Vetenskapsrådet, European Regional Development Fund.
Abstract (Originalsprache)
Herein, we have developed a microneedle-based oxygen sensor wearable patch (O2 MN) designed for real-time, minimally invasive detection of dermal ISF oxygen levels. This work pioneers the use of platinum nanoparticles as the sensing element for oxygen detection within a microneedle platform, enabling high electrochemical sensitivity at a mild applied potential. The wearable patch integrates the nanoparticle-based microneedle as the working electrode and a Ag/AgCl ink-modified microneedle as the counter/reference electrode, operating at an applied potential of −0.3 V to detect a wide range of oxygen (10–254 μM) with high sensitivity (−0.51 nA μM−1) and a rapid response time (9 s). The analytical assessment revealed excellent performance, including high reversibility (RSD = 3.4%), repeatability (RSD = 5.2%), medium-term stability (1.9% deviation over 90 min), acceptable reproducibility (6.6%), and mechanical reliability during skin insertion. Notably, to the best of our knowledge, this study represents the first comprehensive evaluation of the performance of an oxygen microneedle sensor, including systematic in vitro, on-body, and in vivo experiments. The sensor’s accuracy was validated with a Pearson correlation coefficient of 0.99 for in vitro tests, and on-body oxygen measurements in euthanized rats highlighted the reliability of the O2 MN patch for dermal ISF oxygen detection. In vivo studies further revealed a strong positive correlation (Pearson correlation coefficient = 0.83) between dermal ISF and blood oxygen levels, underscoring the O2 MN patch’s potential for biomedical applications through real-time, minimally invasive monitoring.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Chemistry, KTH Royal Institute of Technology, Teknikringen 30, SE-114 28Stockholm, Sweden.
- UCAM-SENS, Universidad Católica San Antonio de Murcia, UCAM HiTech, Avda. Andres Hernandez Ros 1, 30107Murcia, Spain.
- Group of Metabolism and Gene Regulation, Universidad Católica San Antonio de Murcia, UCAM HiTech, Avda. Andres Hernandez Ros 1, 30107Murcia, Spain.
- Section of Neuropharmacology and Addiction Research, Department of Pharmaceutical Biosciences, Uppsala University, SE-751 05Uppsala, Sweden.
ORCID-Profile: Daniel Rojas, Gastón A Crespo, María Cuartero
Förderung
- Ministerio de Ciencia, Innovación y Universidades
- Vetenskapsrådet
- European Regional Development Fund
- Agencia Estatal de Investigación
Zitieren
APA
Wei, Q., Wang, Q., Molinero-Fernández, Á., Rojas, D., Zapata-Pérez, R., Konradsson-Geuken, Å., Crespo, G. A., & Cuartero, M. (2026). Oxygen Detection in Dermal Interstitial Fluid with Microneedles Based on Platinum Nanoparticles. ACS Sensors. https://doi.org/10.1021/acssensors.5c04533
Vancouver
Wei Q, Wang Q, Molinero-Fernández Á, Rojas D, Zapata-Pérez R, Konradsson-Geuken Å, et al. Oxygen Detection in Dermal Interstitial Fluid with Microneedles Based on Platinum Nanoparticles. ACS Sensors. 2026. doi:10.1021/acssensors.5c04533