EUVIMEDEUVIMED entdecken
Crossref

Active detection of melanoma via a dual-extraction-driven flexible microneedles sensor

Min Wu, Songlong Jiao, Lei Liu

Biosensors and Bioelectronics · 2026-05 · 300 · 118451

DOI: 10.1016/j.bios.2026.118451 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2026-05. Verfasst von Min Wu und 2 weiteren Autor:innen. Beteiligte Einrichtung: School of Mechanical Engineering, Southeast University, Nanjing, 211189, People's Republic of China. Hauptthemen laut MEDLINE: Biosensing Techniques, Melanoma, MicroRNAs. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 53 Literaturangaben. Gefördert von: Science and Technology Program of Suzhou, National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Wearable microneedles (MNs) biosensors offer significant potential for minimally invasive biomarker detection in interstitial fluid (ISF). However, suboptimal fluid collection efficiency, dermal irritation due to rigid structural components, and inadequate long-term operational stability collectively represent critical barriers to their clinical translation for early disease diagnosis. Here, we propose a fully integrated dual-extraction-driven flexible microneedles (DFMN) sensor for active detection of melanoma-associated miRNA. The DFMN sensor comprises hydrogel MNs, hollow Au-Ag NPs functionalized flexible electrodes, and a miniaturized wireless module. The synergistic combination of passive diffusion driven by hydrogel MN swelling and active transport via reverse iontophoresis enhanced ISF extraction volume by 1.6-fold within 5 min. This dual-extraction mechanism enabled real-time wireless detection of miRNA-221 with an ultra-low limit of detection (LOD) of 43 aM. Furthermore, to enhance reliability in complex biological environments, an artificial neural network algorithm was implemented, achieving a correlation coefficient (R2) of 0.985. The DFMN sensor holds broad application prospects in interdisciplinary fields such as human-machine interaction, smart healthcare, and artificial intelligence, providing an innovative solution for real-time sensitive detection of wearable flexible sensors.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Biosensing TechniquesMelanomaMicroRNAsHumansNeedlesGoldEquipment DesignLimit of DetectionExtracellular FluidWearable Electronic DevicesMetal NanoparticlesElectrochemical TechniquesHydrogelsMicroneedle Drug Delivery

Stichwörter der Autor:innen

Electrochemical biosensorFlexible biosensorMachine learningMiRNA detectionMicroneedles

Substanzen

MicroRNAsGoldHydrogels

Autor:innen und Institutionen

  • School of Mechanical Engineering, Southeast University, Nanjing, 211189, People's Republic of China.
  • School of Automation, Nanjing University of Information Science and Technology, Nanjing, 211189, People's Republic of China.
  • School of Mechanical Engineering, Southeast University, Nanjing, 211189, People's Republic of China. Electronic address: liulei@seu.edu.cn.

Förderung

  • Science and Technology Program of Suzhou
  • National Natural Science Foundation of China

Zitieren

APA

Wu, M., Jiao, S., & Liu, L. (2026). Active detection of melanoma via a dual-extraction-driven flexible microneedles sensor. Biosensors and Bioelectronics, 300, 118451. https://doi.org/10.1016/j.bios.2026.118451

Vancouver

Wu M, Jiao S, Liu L. Active detection of melanoma via a dual-extraction-driven flexible microneedles sensor. Biosensors and Bioelectronics. 2026;300:118451. doi:10.1016/j.bios.2026.118451

BibTeX
@article{Wu2026,
  author = {Wu, Min and Jiao, Songlong and Liu, Lei},
  title = {Active detection of melanoma via a dual-extraction-driven flexible microneedles sensor},
  journal = {Biosensors and Bioelectronics},
  year = {2026},
  volume = {300},
  pages = {118451},
  doi = {10.1016/j.bios.2026.118451}
}
Kostenlos starten →