Magnetic Rolling Circle Amplification-Assisted Automated Electrochemiluminescence Immunoassay for Ultrasensitive Determination of Protein Biomarkers
Analytical Chemistry · 13.01.2026 · 98 (1) · 729-738
DOI: 10.1021/acs.analchem.5c05778 ↗Kurz gefasst
Fachartikel in der Zeitschrift Analytical Chemistry, online erschienen am 20.12.2025. Verfasst von Danyang Zhang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Key Laboratory of Analytical Chemistry for Life Science of Shaanxi Province, School of Chemistry and Chemical Engineering, Shaanxi Normal University,…. Hauptthemen laut MEDLINE: Luminescent Measurements, Troponin I, Electrochemical Techniques, Nucleic Acid Amplification Techniques. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 46 Literaturangaben. Gefördert von: Fundamental Research Funds for the Central Universities, National Natural Science Foundation of China, Youth Innovation Team of Shaanxi Universities.
Abstract (Originalsprache)
Electrochemiluminescence immunoassay (ECLIA), recognized for its good sensitivity and selectivity, has been widely commercialized in the field of in vitro diagnostics. Despite its advancements, the determination of protein biomarkers at concentrations below pg/mL remains a significant challenge in automated ECLIA. Here, a new magnetic rolling circle amplification-assisted automated ECLIA (ECLIA-RCA) method was proposed for ultrasensitive determination of protein biomarkers. The presence of protein biomarkers could specifically drive the formation of magnetic immunocomplexes between the magnetic bead (MB)-based capture antibody, the target protein, and the single-stranded DNA-tagged secondary antibody. Subsequent triggering of the RCA reaction and signal labeling enables the hybridization of a substantial number of Ru(bpy)32+ derivative-labeled single-stranded DNA to the elongated DNA linkers attached to each magnetic immunocomplex, thereby significantly amplifying the ECL signal. By integrating RCA with MB separation, the developed ECLIA-RCA method realized sensitive determination of cardiac troponin I (cTnI) with a detection limit of 0.1 pg/mL. The method also demonstrates excellent selectivity with minimal interference from nontarget proteins. Validation using 19 clinical human serum samples revealed a good linear correlation of the measured cTnI concentrations between the ECLIA-RCA results and clinical results (R2 = 0.9471). Additionally, the versatility of this automated ECLIA-RCA approach was successfully demonstrated through the quantification of alpha-fetoprotein, achieving a detection limit of 0.3 pg/mL. The ECLIA-RCA method holds great promise for the ultrasensitive determination of protein biomarkers, offering significant potential for early disease diagnosis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Key Laboratory of Analytical Chemistry for Life Science of Shaanxi Province, School of Chemistry and Chemical Engineering, Shaanxi Normal University, Xi'an 710062, P. R. China.
- Lifotronic Technology Co.,Ltd., Shenzhen 518110, P. R. China.
ORCID-Profile: Xiaolin Yang, Chengxiao Zhang, Honglan Qi
Förderung
- Fundamental Research Funds for the Central Universities
- National Natural Science Foundation of China
- Youth Innovation Team of Shaanxi Universities
Zitieren
APA
Zhang, D., Yang, X., Ming, P., Yang, N., Zhang, C., & Qi, H. (2026). Magnetic Rolling Circle Amplification-Assisted Automated Electrochemiluminescence Immunoassay for Ultrasensitive Determination of Protein Biomarkers. Analytical Chemistry, 98(1), 729-738. https://doi.org/10.1021/acs.analchem.5c05778
Vancouver
Zhang D, Yang X, Ming P, Yang N, Zhang C, Qi H. Magnetic Rolling Circle Amplification-Assisted Automated Electrochemiluminescence Immunoassay for Ultrasensitive Determination of Protein Biomarkers. Analytical Chemistry. 2026;98(1):729-738. doi:10.1021/acs.analchem.5c05778
BibTeX
@article{Zhang2026,
author = {Zhang, Danyang and Yang, Xiaolin and Ming, Peng and Yang, Ning and Zhang, Chengxiao and Qi, Honglan},
title = {Magnetic Rolling Circle Amplification-Assisted Automated Electrochemiluminescence Immunoassay for Ultrasensitive Determination of Protein Biomarkers},
journal = {Analytical Chemistry},
year = {2026},
volume = {98},
number = {1},
pages = {729--738},
doi = {10.1021/acs.analchem.5c05778}
}