Task-Based Assessment of Post-processed Image Quality in Full-Field Digital Mammography Using the Task Transfer Function

C Murata, T Hara, A Urikura, A Kameyama

Radiography · 2026-12 · 32 (8) · 103609

DOI: 10.1016/j.radi.2026.103609 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiography, 2026-12. Verfasst von C Murata und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Technology, Nakatsugawa Municipal General Hospital, 1522-1 Komanba, Nakatsugawa, Gifu 508-8502, Japan. Die Arbeit nennt 20 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

INTRODUCTION: To develop a task-based method for evaluating the spatial resolution characteristics of post-processed full-field digital mammography (FFDM) images and to investigate the influence of image-processing algorithms on image quality. METHODS: Post-processed images were acquired using a commercially available FFDM system. Spatial resolution was evaluated by calculating the task transfer function (TTF) from images of a 10-mm acrylic disk embedded in oil, simulating the contrast between a breast mass and the surrounding glandular tissue. TTFs obtained with six image-processing algorithms were compared with the presampled modulation transfer function (pMTF) derived from the raw image data. Additionally, the contrast-to-noise ratio (CNR) was calculated for each algorithm using the same phantom. RESULTS: The TTFs of the post-processed images exceeded unity (>1.0) at low spatial frequencies and showed significant suppression at higher spatial frequencies. These characteristics varied according to the image-processing algorithm: The 50% TTF values of the post-processed images were approximately 20%-35% lower than those of the pMTF, and the TTF response decreased to below 40% at 5 cycles/mm. The CNR ranged from approximately 2.2 to 3.7 and showed no consistent relationship with the TTF results. CONCLUSION: TTF analysis based on the proposed method demonstrated that the spatial-frequency characteristics of post-processed FFDM images differed from those reflected by the pMTF derived from raw image data. Furthermore, no consistent correlation was observed between TTF and CNR across the evaluated image-processing algorithms. These findings suggest that TTF reflects the effective spatial-frequency characteristics of post-processed FFDM images and, when combined with CNR, provides complementary information relevant to task performance. IMPLICATIONS FOR PRACTICE: TTF and CNR may provide complementary information for evaluating post-processed mammographic image quality in clinical practice.

Schlagwörter

Stichwörter der Autor:innen

BreastMammographyPost processed imageQuality assessmentSpatial resolutionTask transfer function

Autor:innen und Institutionen

  • Department of Medical Technology, Nakatsugawa Municipal General Hospital, 1522-1 Komanba, Nakatsugawa, Gifu 508-8502, Japan.
  • Department of Medical Technology, Nakatsugawa Municipal General Hospital, 1522-1 Komanba, Nakatsugawa, Gifu 508-8502, Japan. Electronic address: hara_tnk2@ybb.ne.jp.
  • Graduate School of Health Sciences, Ibaraki Prefectural University of Health Sciences, 4669-2 Ami, Ami-machi, Inashiki-gun, Ibaraki 300-0394, Japan.

Zitieren

APA

Murata, C., Hara, T., Urikura, A., & Kameyama, A. (2026). Task-Based Assessment of Post-processed Image Quality in Full-Field Digital Mammography Using the Task Transfer Function. Radiography, 32(8), 103609. https://doi.org/10.1016/j.radi.2026.103609

Vancouver

Murata C, Hara T, Urikura A, Kameyama A. Task-Based Assessment of Post-processed Image Quality in Full-Field Digital Mammography Using the Task Transfer Function. Radiography. 2026;32(8):103609. doi:10.1016/j.radi.2026.103609

BibTeX
@article{Murata2026,
  author = {Murata, C and Hara, T and Urikura, A and Kameyama, A},
  title = {Task-Based Assessment of Post-processed Image Quality in Full-Field Digital Mammography Using the Task Transfer Function},
  journal = {Radiography},
  year = {2026},
  volume = {32},
  number = {8},
  pages = {103609},
  doi = {10.1016/j.radi.2026.103609}
}
Kostenlos starten →