Smartphone-based particle tracking velocimetry for the in vitro assessment of coronary flows

Elena Torta, Bianca Griffo, Giuseppe C A Caridi, Giuseppe De Nisco, Claudio Chiastra, Umberto Morbiducci, Diego Gallo

Medical Engineering & Physics · 2024

DOI: 10.1016/j.medengphy.2024.104144 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Engineering & Physics, 2024. Verfasst von Elena Torta und 6 weiteren Autor:innen. Beteiligte Einrichtung: Polito(BIO)Med Lab, Department of Mechanical and Aerospace Engineering, Politecnico di Torino, Turin, Italy. Hauptthemen laut MEDLINE: Smartphone, Cardiovascular Physiological Phenomena. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Italian Ministry of Education, University and Research, MUR.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The present study adopts a smartphone-based approach for the experimental characterization of coronary flows. Technically, Particle Tracking Velocimetry (PTV) measurements were performed using a smartphone camera and a low-power continuous wave laser in realistic healthy and stenosed phantoms of left anterior descending artery with inflow Reynolds numbers approximately ranging from 20 to 200. A Lagrangian-Eulerian mapping was performed to convert Lagrangian PTV velocity data to a Eulerian grid. Eulerian velocity and vorticity data obtained from smartphone-based PTV measurements were compared with Particle Image Velocimetry (PIV) measurements performed with a smartphone-based setup and with a conventional setup based on a high-power double-pulsed laser and a CMOS camera. Smartphone-based PTV and PIV velocity flow fields substantially agreed with conventional PIV measurements, with the former characterized by lower average percentage differences than the latter. Discrepancies emerged at high flow regimes, especially at the stenosis throat, due to particle image blur generated by smartphone camera shutter speed and image acquisition frequency. In conclusion, the present findings demonstrate the feasibility of PTV measurements using a smartphone camera and a low-power light source for the in vitro characterization of cardiovascular flows for research, industrial and educational purposes, with advantages in terms of costs, safety and usability.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SmartphoneCardiovascular Physiological PhenomenaRheologyBlood Flow VelocityPhantoms, Imaging

Stichwörter der Autor:innen

Coronary arteriesLow-power laserParticle image velocimetryParticle tracking velocimetrySmartphone

Autor:innen und Institutionen

  • Polito(BIO)Med Lab, Department of Mechanical and Aerospace Engineering, Politecnico di Torino, Turin, Italy.
  • Institute of Fluid Mechanics and Heat Transfer, TU Wien, Vienna, Austria.
  • Polito(BIO)Med Lab, Department of Mechanical and Aerospace Engineering, Politecnico di Torino, Turin, Italy. Electronic address: diego.gallo@polito.it.

Förderung

  • Italian Ministry of Education, University and Research
  • MUR

Zitieren

APA

Torta, E., Griffo, B., Caridi, G. C. A., De Nisco, G., Chiastra, C., Morbiducci, U., & Gallo, D. (2024). Smartphone-based particle tracking velocimetry for the in vitro assessment of coronary flows. Medical Engineering & Physics. https://doi.org/10.1016/j.medengphy.2024.104144

Vancouver

Torta E, Griffo B, Caridi GCA, De Nisco G, Chiastra C, Morbiducci U, et al. Smartphone-based particle tracking velocimetry for the in vitro assessment of coronary flows. Medical Engineering & Physics. 2024. doi:10.1016/j.medengphy.2024.104144

BibTeX
@article{Torta2024,
  author = {Torta, Elena and Griffo, Bianca and Caridi, Giuseppe C A and De Nisco, Giuseppe and Chiastra, Claudio and Morbiducci, Umberto and Gallo, Diego},
  title = {Smartphone-based particle tracking velocimetry for the in vitro assessment of coronary flows},
  journal = {Medical Engineering & Physics},
  year = {2024},
  doi = {10.1016/j.medengphy.2024.104144}
}
Kostenlos starten →