EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Mechanical load applied by Intraosseous Transcutaneous Amputation Prosthesis (ITAP) during walking on level and sloped treadmill: A case study

K Ahmed, M Thornton, S J G Taylor

Medical Engineering & Physics · 2024

DOI: 10.1016/j.medengphy.2023.104097 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Engineering & Physics, 2024. Verfasst von K Ahmed und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics and Musculoskeletal Science, Division of Surgery and Interventional Science, University College London, Brockley Hill, Sta…. Hauptthemen laut MEDLINE: Bone-Anchored Prosthesis, Artificial Limbs, Amputees. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This proof of concept study presents a method to collect and analyse kinetic data from one participant with a transfemoral amputation fitted with a percutaneous osseointegrated implant walking on a level and sloped treadmill. We describe the construction of and results from a bespoke wireless six axis load cell built into one participant's prosthetic assembly. The load cell does not clinically compromise the participant in any way and is an initial milestone in the development of a light-weight wireless load cell for use with percutaneous osseointegrated implants. In this case, it is the first time that kinetic data from a participant fitted with an Intraosseous Transcutaneous Amputation Prosthesis has been published. We propose that the data can be used to model the load transfer to the host bone, with several clinically significant applications. The raw dynamic data are made available and quasi-static load cases for each functional phase of gait are presented. Peak forces obtained in the medio-lateral (X), cranio-caudal (Y) and antero-posterior (Z) axes over level ground respectively were -243.8 N (0.24 BW), 1321.5 N (1.31 BW) and -421.8 N (0.42 BW); uphill were -141.0 N (0.14 BW), 1604.2 N (1.59 BW), -498.1 N (0.49 BW); downhill were -206.0 N (0.20 BW), 1103.9 N (1.09 BW), -547.2 N (0.54 BW). The kinetics broadly followed able bodied gait patterns with some gait strategies consistent in participants with other implant designs or prosthetic socket connections, for example offloading the artificial limb downhill.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bone-Anchored ProsthesisArtificial LimbsAmputeesHumansWalkingGaitAmputation, SurgicalProsthesis DesignBiomechanical Phenomena

Stichwörter der Autor:innen

Amputation biomechanicsBone-anchored prosthesisIntraosseous transcutaneous amputation prosthesisOsseointegrated implantSkeletally anchored amputation implantTransfemoral bone anchorTransfemoral kinetics

Autor:innen und Institutionen

  • Department of Orthopaedics and Musculoskeletal Science, Division of Surgery and Interventional Science, University College London, Brockley Hill, Stanmore, Middlesex, HA7 4LP, UK; Center for Bionics and Pain Research, M…
  • Department of Orthopaedics and Musculoskeletal Science, Division of Surgery and Interventional Science, University College London, Brockley Hill, Stanmore, Middlesex, HA7 4LP, UK; Motor Learning Laboratory, Royal Nation…
  • Department of Orthopaedics and Musculoskeletal Science, Division of Surgery and Interventional Science, University College London, Brockley Hill, Stanmore, Middlesex, HA7 4LP, UK.

Zitieren

APA

Ahmed, K., Thornton, M., & Taylor, S. J. G. (2024). Mechanical load applied by Intraosseous Transcutaneous Amputation Prosthesis (ITAP) during walking on level and sloped treadmill: A case study. Medical Engineering & Physics. https://doi.org/10.1016/j.medengphy.2023.104097

Vancouver

Ahmed K, Thornton M, Taylor SJG. Mechanical load applied by Intraosseous Transcutaneous Amputation Prosthesis (ITAP) during walking on level and sloped treadmill: A case study. Medical Engineering & Physics. 2024. doi:10.1016/j.medengphy.2023.104097

BibTeX
@article{Ahmed2024,
  author = {Ahmed, K and Thornton, M and Taylor, S J G},
  title = {Mechanical load applied by Intraosseous Transcutaneous Amputation Prosthesis (ITAP) during walking on level and sloped treadmill: A case study},
  journal = {Medical Engineering & Physics},
  year = {2024},
  doi = {10.1016/j.medengphy.2023.104097}
}