How to avoid baseplate failure: the effect of compression and reverse shoulder arthroplasty baseplate design on implant stability

Miguel A Diaz, Adam J Hutchinson, Eric T Ricchetti, Jason E Hsu, Grant E Garrigues, Sergio Gutiérrez, Mark A Frankle

Journal of Shoulder and Elbow Surgery · 2024

DOI: 10.1016/j.jse.2023.07.043 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Shoulder and Elbow Surgery, 2024. Verfasst von Miguel A Diaz und 6 weiteren Autor:innen. Beteiligte Einrichtung: Foundation for Orthopaedic Research & Education, Tampa, FL, USA. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Shoulder, Shoulder Joint. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 50 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Failure to achieve fixation of the glenoid baseplate will lead to clinical failure. The fixation of the baseplate to the scapula must be able to withstand sufficient shear forces to allow bony ingrowth. The importance of compression to neutralize the forces at the baseplate-bone interface has been assumed to be critical in limiting excessive micromotion. The purpose of this study is to determine the effect of compression on implant stability with different baseplate designs. METHODS: Various baseplate designs (1-piece monolithic central screw [1P], 2-piece locking central screw [2PL], and 2-piece nonlocking center screw [2PNL]) were investigated at 3 different compressive forces (high [810 N], medium [640 N], and low [530 N]). Synthetic bone cylinders were instrumented, and peripheral screws were used in all models. The combination of 1 locking and 3 nonlocking peripheral screw fixation was selected as worst-case scenario. Dynamic testing protocol followed the ASTM F2028-17 standard. The baseplate micromotion at high compression was compared to low compression. Additionally, the baseplate micromotion for each design was compared at baseline (first 50 cycles) and at 10,000 cycles for the 3 different compressive forces where motion above 150 μm was defined as failure. RESULTS: Baseplate micromotion was found to negatively correlate with compression (rpb = -0.83, P < .0001). At baseline, all baseplate designs were considered stable, regardless of compression. With high compression, average micromotion at the glenoid baseplate-bone interface remained below the 150-μm threshold for all baseplate designs at 10,000 cycles (1P: 50 ± 10 μm; 2PL: 78 ± 32 μm; 2PNL: 79 ± 8 μm; P = .060). With medium compression, average micromotion at 10,000 cycles for all 3 designs remained below the 150-μm threshold (1P: 88 ± 22 μm; 2PL: 132 ± 26 μm; 2PNL: 107 ± 39 μm). The 2PL design had the highest amount of micromotion (P = .013). With low compression, both 2-piece designs had an average micromotion above the 150-μm threshold whereas the 1-piece design did not (1P: 133 ± 35 μm; 2PL: 183 ± 21 μm; 2PNL: 166 ± 39 μm). The 2PL design had significantly higher micromotion when compared to 1P design (P = .041). DISCUSSION: The stability of a central screw baseplate correlates with the amount of compression obtained and is affected by implant design. For the same amount of compression, more micromotion is observed in a 2-piece design than a 1-piece design.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, ShoulderShoulder JointHumansArthroplastyScapulaMotionBiomechanical Phenomena

Stichwörter der Autor:innen

Baseplate designRSA baseplatebaseplate compressionbaseplate micromotionstability

Autor:innen und Institutionen

  • Foundation for Orthopaedic Research & Education, Tampa, FL, USA.
  • Department of Orthopaedic Surgery, Cleveland Clinic Foundation, Cleveland, OH, USA.
  • Department of Orthopaedics and Sports Medicine, University of Washington, Seattle, WA, USA.
  • Midwest Orthopaedics at Rush, Rush University Medical Center, Chicago, IL, USA.
  • Department of Orthopaedics & Sports Medicine, University of South Florida, Tampa, FL, USA; Florida Orthopaedic Institute, Tampa, FL, USA. Electronic address: mfrankle@floridaortho.com.

Zitieren

APA

Diaz, M. A., Hutchinson, A. J., Ricchetti, E. T., Hsu, J. E., Garrigues, G. E., Gutiérrez, S., & Frankle, M. A. (2024). How to avoid baseplate failure: the effect of compression and reverse shoulder arthroplasty baseplate design on implant stability. Journal of Shoulder and Elbow Surgery. https://doi.org/10.1016/j.jse.2023.07.043

Vancouver

Diaz MA, Hutchinson AJ, Ricchetti ET, Hsu JE, Garrigues GE, Gutiérrez S, et al. How to avoid baseplate failure: the effect of compression and reverse shoulder arthroplasty baseplate design on implant stability. Journal of Shoulder and Elbow Surgery. 2024. doi:10.1016/j.jse.2023.07.043

BibTeX
@article{Diaz2024,
  author = {Diaz, Miguel A and Hutchinson, Adam J and Ricchetti, Eric T and Hsu, Jason E and Garrigues, Grant E and Gutiérrez, Sergio and Frankle, Mark A},
  title = {How to avoid baseplate failure: the effect of compression and reverse shoulder arthroplasty baseplate design on implant stability},
  journal = {Journal of Shoulder and Elbow Surgery},
  year = {2024},
  doi = {10.1016/j.jse.2023.07.043}
}
Kostenlos starten →