Development of a glucose reference material in human serum for clinical assay standardization

Zhichao Wang, Yuanyuan Du, Qiaoxuan Zhang, Jun Yan, Min Zhan, Pengwei Zhang, Zemin Wan, Jianbing Wang, Xianzhang Huang, Liqiao Han

Journal of Pharmaceutical and Biomedical Analysis · 2023

DOI: 10.1016/j.jpba.2023.115285 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Pharmaceutical and Biomedical Analysis, 2023. Verfasst von Zhichao Wang und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine (Guangdong Provincial Hospital of Chine…. Hauptthemen laut MEDLINE: Blood Glucose, Glucose. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 15 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Blood glucose is an important monosaccharide functioning as the main source of energy for the human body. The accurate measurement of blood glucose is crucial for the screening, diagnosis, and monitoring of diabetes and diabetes-associated diseases. To assure the reliability and traceability of blood glucose measurements, we developed a reference material (RM) for use in human serum at two different concentrations, which were certified by the National Institute of Metrology (NIM) as GBW(E)091040 and GBW(E)091043. METHODS: Raw serum samples were collected from residual samples after clinical testing, filtered, and repackaged under mild stirring. The homogeneity and stability of the samples were examined according to ISO Guide 35: 2017. Commutability was evaluated in compliance with CLSI EP30-A. Value assignment was carried out in six certified reference laboratories using the JCTLM-listed reference method for serum glucose. Moreover, the RMs was further applied in a trueness verification program. RESULTS: The developed RMs was homogeneous and commutable enough for clinical use. They were also stable for 24 h at 2-8 ℃ or 20-25 ℃ and for at least 4 years at - 70 ℃. The certified values were 5.20 ± 0.18 mmol/L and 8.18 ± 0.19 mmol/L (k = 2) for GBW(E)091040 and GBW(E)091043, respectively. The pass rates were evaluated by bias, coefficient of variation (CV), and total error (TE) for 66 clinical laboratories in the trueness verification program were 57.6%, 98.5%, and 89.4% of GBW(E)091040, and 51.5%, 98.5%, and 90.9% of GBW(E)091043, respectively. CONCLUSION: The developed RM could be used for the standardization of reference and clinical systems with satisfactory performance and traceable values, providing strong support for the accurate measurement of blood glucose.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood GlucoseGlucoseHumansReproducibility of ResultsReference StandardsSerum

Stichwörter der Autor:innen

Blood glucoseCommutabilityHomogeneityReference materialStabilityTraceability

Substanzen

Blood GlucoseGlucose

Autor:innen und Institutionen

  • Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine (Guangdong Provincial Hospital of Chinese Medicine), Guangzhou 510120, China; Second Clinical Medical College…
  • Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine (Guangdong Provincial Hospital of Chinese Medicine), Guangzhou 510120, China.
  • Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine (Guangdong Provincial Hospital of Chinese Medicine), Guangzhou 510120, China. Electronic address: wangjianbin…
  • Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine (Guangdong Provincial Hospital of Chinese Medicine), Guangzhou 510120, China. Electronic address: huangxz020@…
  • Department of Laboratory Medicine, The Second Affiliated Hospital of Guangzhou University of Chinese Medicine (Guangdong Provincial Hospital of Chinese Medicine), Guangzhou 510120, China. Electronic address: hanliqiao@g…

Zitieren

APA

Wang, Z., Du, Y., Zhang, Q., Yan, J., Zhan, M., Zhang, P., Wan, Z., Wang, J., Huang, X., & Han, L. (2023). Development of a glucose reference material in human serum for clinical assay standardization. Journal of Pharmaceutical and Biomedical Analysis. https://doi.org/10.1016/j.jpba.2023.115285

Vancouver

Wang Z, Du Y, Zhang Q, Yan J, Zhan M, Zhang P, et al. Development of a glucose reference material in human serum for clinical assay standardization. Journal of Pharmaceutical and Biomedical Analysis. 2023. doi:10.1016/j.jpba.2023.115285

BibTeX
@article{Wang2023,
  author = {Wang, Zhichao and Du, Yuanyuan and Zhang, Qiaoxuan and Yan, Jun and Zhan, Min and Zhang, Pengwei and Wan, Zemin and Wang, Jianbing and Huang, Xianzhang and Han, Liqiao},
  title = {Development of a glucose reference material in human serum for clinical assay standardization},
  journal = {Journal of Pharmaceutical and Biomedical Analysis},
  year = {2023},
  doi = {10.1016/j.jpba.2023.115285}
}
Kostenlos starten →