Cone-beam versus photon-counting detector CT: Influence of dose variations on the detection of simulated mandibular osseous lesions
Journal of Cranio-Maxillofacial Surgery · 2025-08 · 53 (8) · 1236-1245
DOI: 10.1016/j.jcms.2025.04.023 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Cranio-Maxillofacial Surgery, 2025-08. Verfasst von Adib Al-Haj Husain und 9 weiteren Autor:innen. Beteiligte Einrichtung: Clinic of Cranio-Maxillofacial and Oral Surgery, Center of Dental Medicine, University of Zurich, Zurich, Switzerland; Department of Cranio-Maxillofa…. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Radiation Dosage, Mandibular Diseases, Tomography, X-Ray Computed. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Bayer, Siemens.
Abstract (Originalsprache)
This study compared cone-beam computed tomography (CBCT) and photon-counting detector computed tomography (PCD-CT) for detecting simulated mandibular osseous lesions, emphasizing lesion detectability, quantitative accuracy, and inter-reader reliability at equivalent radiation doses. Three types of simulated osseous lesions (sequestrum/fracture, cystic lesion, and extended periodontal gap) were evaluated by two readers, across three radiation dose levels: high (360 μSv), standard (122 μSv), and low (20 μSv). Image quality, artifact presence, and lesion detectability were qualitatively evaluated using a 5-point Likert scale (5 = highest to 1 = lowest rating). Lesion dimensions were quantitatively measured. Statistical analysis included descriptive statistics and weighted kappa for interobserver reliability. High-dose PCD-CT consistently provided the highest image quality, lesion detectability, and minimal artifacts across all lesion types (5 ± 0), achieving perfect inter-reader agreement (100 %). PCD-CT outperformed CBCT in lesion detectability, with the greatest difference at lower doses (3.87 ± 0.35 to 4.83 ± 0.39 vs. 3.13 ± 0.53 to 4.5 ± 0.52). PCD-CT demonstrated superior inter-reader reliability, with perfect agreement for high-dose (κ = 1.0; p < 0.001) and good agreement for standard- and low-dose protocols (κ = 0.909 and κ = 0.744; p < 0.001). Quantitative measurements showed minimal intermodal differences ranging from 0.15 to 0.34 mm, confirming PCD-CT's clinical precision even at low-dose protocols. These findings highlight PCD-CT's potential in dentomaxillofacial imaging, enhancing diagnostic capabilities and reducing radiation exposure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Clinic of Cranio-Maxillofacial and Oral Surgery, Center of Dental Medicine, University of Zurich, Zurich, Switzerland; Department of Cranio-Maxillofacial and Oral Surgery, University Hospital Zurich, University of Zuric…
- Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland. Electronic address: victor.mergen@usz.ch.
- Clinic of Cranio-Maxillofacial and Oral Surgery, Center of Dental Medicine, University of Zurich, Zurich, Switzerland. Electronic address: silvio.valdec@zzm.uzh.ch.
- Department of Neuroradiology, Clinical Neuroscience Center, University Hospital Zurich, University of Zurich, Zurich, Switzerland. Electronic address: sebastian.winklhofer@hirslanden.ch.
- Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland. Electronic address: hatem.alkadhi@usz.ch.
- Department of Cranio-Maxillofacial and Oral Surgery, University Hospital Zurich, University of Zurich, Zurich, Switzerland. Electronic address: harald.essig@usz.ch.
Förderung
- Bayer
- Siemens
Zitieren
APA
Al-Haj Husain, A., Mergen, V., Valdec, S., Winklhofer, S., Alkadhi, H., Essig, H., Frauenfelder, T., Kessler, P., Lie, S. A. N., & Stadlinger, B. (2025). Cone-beam versus photon-counting detector CT: Influence of dose variations on the detection of simulated mandibular osseous lesions. Journal of Cranio-Maxillofacial Surgery, 53(8), 1236-1245. https://doi.org/10.1016/j.jcms.2025.04.023
Vancouver
Al-Haj Husain A, Mergen V, Valdec S, Winklhofer S, Alkadhi H, Essig H, et al. Cone-beam versus photon-counting detector CT: Influence of dose variations on the detection of simulated mandibular osseous lesions. Journal of Cranio-Maxillofacial Surgery. 2025;53(8):1236-1245. doi:10.1016/j.jcms.2025.04.023
BibTeX
@article{AlHajHusain2025,
author = {Al-Haj Husain, Adib and Mergen, Victor and Valdec, Silvio and Winklhofer, Sebastian and Alkadhi, Hatem and Essig, Harald and Frauenfelder, Thomas and Kessler, Peter and Lie, Suen An Nynke and Stadlinger, Bernd},
title = {Cone-beam versus photon-counting detector CT: Influence of dose variations on the detection of simulated mandibular osseous lesions},
journal = {Journal of Cranio-Maxillofacial Surgery},
year = {2025},
volume = {53},
number = {8},
pages = {1236--1245},
doi = {10.1016/j.jcms.2025.04.023}
}
EUVIMED