Exploring Dental Anatomy: An Ex Vivo Comparative Study Between Photon‐Counting Detector CT and Cone‐Beam CT
Clinical and Experimental Dental Research · 2025
DOI: 10.1002/cre2.70266 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Clinical and Experimental Dental Research, 2025. Verfasst von Adib Al-Haj Husain und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cranio-Maxillofacial and Oral Surgery, University Hospital Zurich, University of Zurich, Zurich, Switzerland. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Mandible, Tooth. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
The aim of this ex vivo study was to assess the performance of photon‐counting detector computed tomography (PCD‐CT) compared with cone‐beam computed tomography (CBCT) at equivalent radiation doses for detecting and evaluating mandibular and dental anatomical structures in porcine cadavers. Material and Methods This intermodal comparative study evaluated imaging protocols at three radiation dose levels (high: 360 μSv, standard: 145 μSv, low: 20 μSv) in six porcine cadaver heads, analyzing 12 CBCT and 18 PCD‐CT volumes. Two blinded observers assessed image quality, artifact susceptibility, and diagnostic interpretability using a 5‐point Likert scale (5 = highest, 1 = lowest). Statistical analysis included descriptive statistics and interobserver reliability, assessed by weighted kappa (κ) analysis.
Results
PCD‐CT either matched or outperformed CBCT at standard‐ and low‐dose protocols, demonstrating superior anatomy coverage, density, contrast, and less artifact susceptibility. High‐ and standard‐dose protocols achieved perfect scores 5 (SD = 0). At low‐dose levels, PCD‐CT showed slightly lower scores but still outperformed CBCT. PCD‐CT showed minimal artifacts, with no significant artifacts in high and standard doses. Inter‐ and intra‐reader reliability was higher for PCD‐CT (κ: 0.694–1; p < 0.001) compared to CBCT (0.55–0.916; p < 0.001), with the difference being largest at low doses.
Conclusion
PCD‐CT outperforms CBCT in diagnostic interpretability and artifact reduction across various radiation dose levels, offering a promising alternative for dentomaxillofacial imaging that aligns with the ALADAIP principle.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cranio-Maxillofacial and Oral Surgery, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
- Clinic of Cranio-Maxillofacial and Oral Surgery, Center of Dental Medicine, University of Zurich, Zurich, Switzerland.
- Department of Cranio-Maxillofacial Surgery, GROW School for Oncology and Reproduction, Maastricht University Medical Centre, Maastricht, the Netherlands.
- Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
- Department of Neuroradiology, Clinical Neuroscience Center, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
- Oral Radiology, Department of Dentistry and Oral Health, Aarhus University, Aarhus, Denmark.
ORCID-Profile: Adib Al-Haj Husain, Rubens Spin-Neto
Zitieren
APA
Al-Haj Husain, A., Benz, F., Mergen, V., Valdec, S., Winklhofer, S., Alkadhi, H., Essig, H., Frauenfelder, T., Spin-Neto, R., & Stadlinger, B. (2025). Exploring Dental Anatomy: An Ex Vivo Comparative Study Between Photon‐Counting Detector CT and Cone‐Beam CT. Clinical and Experimental Dental Research. https://doi.org/10.1002/cre2.70266
Vancouver
Al-Haj Husain A, Benz F, Mergen V, Valdec S, Winklhofer S, Alkadhi H, et al. Exploring Dental Anatomy: An Ex Vivo Comparative Study Between Photon‐Counting Detector CT and Cone‐Beam CT. Clinical and Experimental Dental Research. 2025. doi:10.1002/cre2.70266
BibTeX
@article{AlHajHusain2025,
author = {Al-Haj Husain, Adib and Benz, Fabian and Mergen, Victor and Valdec, Silvio and Winklhofer, Sebastian and Alkadhi, Hatem and Essig, Harald and Frauenfelder, Thomas and Spin-Neto, Rubens and Stadlinger, Bernd},
title = {Exploring Dental Anatomy: An Ex Vivo Comparative Study Between Photon‐Counting Detector CT and Cone‐Beam CT},
journal = {Clinical and Experimental Dental Research},
year = {2025},
doi = {10.1002/cre2.70266}
}
EUVIMED