EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Asymmetrical Expansion of Balloon-Expandable Transcatheter Aortic Valve Prostheses

Annette Maznyczka, Dierik Heg, Daijiro Tomii, Masaaki Nakase, Pernille Steen Baekke, Jonas Lanz, Stefan Stortecky, David Reineke, Stephan Windecker, Thomas Pilgrim

JACC: Cardiovascular Interventions · 2024-09 · 17 (17) · 2011-2022

DOI: 10.1016/j.jcin.2024.05.039 ↗

Kurz gefasst

Fachartikel in der Zeitschrift JACC: Cardiovascular Interventions, 2024-09. Verfasst von Annette Maznyczka und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Electronic address: https://twitter.com/AMazn…. Hauptthemen laut MEDLINE: Hemodynamics, Transcatheter Aortic Valve Replacement, Heart Valve Prosthesis, Prosthesis Design. Die Arbeit wurde bisher 28-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Asymmetrical expansion of transcatheter heart valves (THVs), manifesting as stent frame deformation, is an occasional fluoroscopic finding in transcatheter aortic valve replacement (TAVR). OBJECTIVES: The aim of this study was to investigate the impact of asymmetrical expansion of balloon-expandable THVs on hemodynamic valve performance and clinical outcomes. METHODS: In a prospective registry, TAVR asymmetry index was measured using freeze-frame fluoroscopic images and was defined as the ratio of THV heights: [(longer height/shorter height) - 1] × 100. THV hemodynamic performance was measured using echocardiography before hospital discharge. Impaired hemodynamic valve performance was defined as a mean residual THV gradient ≥20 mm Hg and/or moderate or greater paravalvular regurgitation. RESULTS: Among 1,216 patients undergoing transfemoral TAVR for native severe aortic valve stenosis with contemporary balloon-expandable THVs between February 2014 and June 2022, asymmetry index was an excellent predictor of impaired hemodynamic valve performance (area under the receiver-operating characteristic curve: 0.88; 95% CI: 0.84-0.92; P 5.5% (sensitivity 77%, specificity 86%) and occurred in 17% of patients. Higher asymmetry index, as a continuous variable, was associated with impaired hemodynamic valve performance independent of total aortic valve calcium, bicuspid anatomy, balloon-expandable valve prosthesis type or size, and Society of Thoracic Surgeons Predicted Risk of Mortality score (OR: 1.37; 95% CI: 1.29-1.46; P < 0.001). High TAVR asymmetry index was not associated with all-cause mortality during a median of 376 days of follow-up (HR: 1.00; 95% CI: 0.71-1.41; P = 0.989). CONCLUSIONS: Asymmetrical expansion of balloon-expandable THVs was associated with impaired hemodynamic valve performance but not with clinical outcomes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HemodynamicsTranscatheter Aortic Valve ReplacementHeart Valve ProsthesisProsthesis DesignAortic Valve StenosisAortic ValveBalloon ValvuloplastyRegistriesHumansMaleFemaleAged, 80 and overTreatment OutcomeAgedRisk FactorsTime FactorsProspective StudiesSeverity of Illness IndexRecovery of FunctionAortic Valve Insufficiency

Stichwörter der Autor:innen

asymmetrybioprosthetic valve functionclinical outcomestranscatheter aortic valve replacement

Autor:innen und Institutionen

  • Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Electronic address: https://twitter.com/AMaznyczka.
  • Department of Clinical Research, University of Bern, Bern, Switzerland.
  • Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Electronic address: https://twitter.com/DaijiroTomii.
  • Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Electronic address: https://twitter.com/masaaki0825.
  • Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Electronic address: https://twitter.com/pbaekke.
  • Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.

Zitieren

APA

Maznyczka, A., Heg, D., Tomii, D., Nakase, M., Baekke, P. S., Lanz, J., Stortecky, S., Reineke, D., Windecker, S., & Pilgrim, T. (2024). Asymmetrical Expansion of Balloon-Expandable Transcatheter Aortic Valve Prostheses. JACC: Cardiovascular Interventions, 17(17), 2011-2022. https://doi.org/10.1016/j.jcin.2024.05.039

Vancouver

Maznyczka A, Heg D, Tomii D, Nakase M, Baekke PS, Lanz J, et al. Asymmetrical Expansion of Balloon-Expandable Transcatheter Aortic Valve Prostheses. JACC: Cardiovascular Interventions. 2024;17(17):2011-2022. doi:10.1016/j.jcin.2024.05.039

BibTeX
@article{Maznyczka2024,
  author = {Maznyczka, Annette and Heg, Dierik and Tomii, Daijiro and Nakase, Masaaki and Baekke, Pernille Steen and Lanz, Jonas and Stortecky, Stefan and Reineke, David and Windecker, Stephan and Pilgrim, Thomas},
  title = {Asymmetrical Expansion of Balloon-Expandable Transcatheter Aortic Valve Prostheses},
  journal = {JACC: Cardiovascular Interventions},
  year = {2024},
  volume = {17},
  number = {17},
  pages = {2011--2022},
  doi = {10.1016/j.jcin.2024.05.039}
}
Kostenlos starten →