EUVIMEDEUVIMED entdecken
Crossref

Novel Approach to Lead Extraction, Utilizing Vibration Energy

Orr Tomer, Yair Elitzur, Andrzej Przybylski, Janusz Romanek, Uri Stein, Roey Shafrir, Ayelet Shauer, Charles J Love, Paul A Friedman, Abhishek J Deshmukh, David Luria

JACC: Clinical Electrophysiology · 2026-07 · 12 (7) · 1684-1693

DOI: 10.1016/j.jacep.2026.02.039 ↗

Kurz gefasst

Fachartikel in der Zeitschrift JACC: Clinical Electrophysiology, 2026-07. Verfasst von Orr Tomer und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Hadassah Medical Center and Faculty of Medicine, Hebrew University of Jerusalem, Jerusalem, Israel. Electronic address: orr…. Hauptthemen laut MEDLINE: Vibration, Device Lead Extraction, Defibrillators, Implantable, Pacemaker, Artificial. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND Lead extraction procedures typically use "over-the-lead" cutting tools (sheaths) to separate leads from the surrounding fibrotic, vascular, and cardiac tissue. These methods carry significant risks of collateral damage and major bleeding complications. OBJECTIVES The goal of this study was to evaluate the safety and efficacy of a novel vibration-based lead extraction technique that eliminates the need for cutting tools. The study included comprehensive preclinical testing and initial human experience. METHODS Development included 3 phases: 1) bench experiments; 2) animal studies (N = 13) with varying lead implant durations; and 3) a first-in-human study enrolling 10 consecutive patients requiring a total of 18 lead extractions. The primary endpoints were complete procedural success and absence of major complications. The secondary endpoint was successful extraction of all leads without conventional tools. RESULTS The bench experiments established optimal vibration parameters for lead extraction. The animal studies confirmed the safety of high-frequency vibration application to an implanted pacing lead. In 9 animals with long-term implants (up to 5 years), all leads (6 pacing and 3 implantable cardiac-defibrillator) were successfully extracted without significant tissue damage on anatomical and histologic examination. In the human study, complete extraction was achieved in all 10 cases (18 leads [12 pacing, 4 implantable cardiac-defibrillator, and 2 coronary sinus]) with a mean dwell time of 48.6 ± 21.74 months (21-78 months). Mean lead vibration time was 40 seconds (6-176 seconds). No device-related complications occurred during the procedure or at 1-month follow-up. CONCLUSIONS This novel vibration-based extraction technique exhibited excellent safety and efficiency profiles in preclinical and initial human study, potentially offering a simple, rapid, and safe alternative to conventional extraction methods.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

VibrationDevice Lead ExtractionDefibrillators, ImplantablePacemaker, ArtificialDevice RemovalHumansAnimalsFemaleMaleMiddle AgedAgedAdult

Stichwörter der Autor:innen

novel extraction methodtransvenous lead extractionvibration energy

Autor:innen und Institutionen

  • Department of Cardiology, Hadassah Medical Center and Faculty of Medicine, Hebrew University of Jerusalem, Jerusalem, Israel. Electronic address: orrtomer@gmail.com.
  • Department of Cardiology, Hadassah Medical Center and Faculty of Medicine, Hebrew University of Jerusalem, Jerusalem, Israel.
  • Clinical Department of Cardiology with the Acute Coronary Syndromes Subdivision, Clinical Provincial Hospital No. 2, Rzeszów, Faculty of Medicine, University of Rzeszow, Rzeszów, Poland.
  • Xtrac Medical O.S. Ltd, Raanana, Israel.
  • Johns Hopkins School of Medicine, Baltimore, Maryland, USA.
  • Department of Cardiovascular Medicine, Mayo Clinic, Rochester, Minnesota, USA.

Zitieren

APA

Tomer, O., Elitzur, Y., Przybylski, A., Romanek, J., Stein, U., Shafrir, R., Shauer, A., Love, C. J., Friedman, P. A., Deshmukh, A. J., & Luria, D. (2026). Novel Approach to Lead Extraction, Utilizing Vibration Energy. JACC: Clinical Electrophysiology, 12(7), 1684-1693. https://doi.org/10.1016/j.jacep.2026.02.039

Vancouver

Tomer O, Elitzur Y, Przybylski A, Romanek J, Stein U, Shafrir R, et al. Novel Approach to Lead Extraction, Utilizing Vibration Energy. JACC: Clinical Electrophysiology. 2026;12(7):1684-1693. doi:10.1016/j.jacep.2026.02.039

BibTeX
@article{Tomer2026,
  author = {Tomer, Orr and Elitzur, Yair and Przybylski, Andrzej and Romanek, Janusz and Stein, Uri and Shafrir, Roey and Shauer, Ayelet and Love, Charles J and Friedman, Paul A and Deshmukh, Abhishek J and Luria, David},
  title = {Novel Approach to Lead Extraction, Utilizing Vibration Energy},
  journal = {JACC: Clinical Electrophysiology},
  year = {2026},
  volume = {12},
  number = {7},
  pages = {1684--1693},
  doi = {10.1016/j.jacep.2026.02.039}
}
Kostenlos starten →