Ventricular Intramyocardial Navigation for Tachycardia Ablation Guided by Electrograms (VINTAGE)
JACC: Clinical Electrophysiology · 2024
DOI: 10.1016/j.jacep.2024.04.002 ↗Kurz gefasst
Fachartikel in der Zeitschrift JACC: Clinical Electrophysiology, 2024. Verfasst von Rim N Halaby und 11 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Branch, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland, U…. Hauptthemen laut MEDLINE: Tachycardia, Ventricular, Catheter Ablation, Electrophysiologic Techniques, Cardiac. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Intramural NIH HHS, National Heart, Lung, and Blood Institute, NHLBI DIR.
Abstract (Originalsprache)
BACKGROUND: Deep intramural ventricular tachycardia substrate targets are difficult to access, map, and ablate from endocardial and epicardial surfaces, resulting in high recurrence rates. OBJECTIVES: In this study, the authors introduce a novel approach called ventricular intramyocardial navigation for tachycardia ablation guided by electrograms (VINTAGE) to access and ablate anatomically challenging ventricular tachycardia from within the myocardium. METHODS: Guidewire/microcatheter combinations were navigated deep throughout the extravascular myocardium, accessed directly from the right ventricle cavity, in Yorkshire swine (6 naive, 1 infarcted). Devices were steered to various intramyocardial targets including the left ventricle summit, guided by fluoroscopy, unipolar electrograms, and/or electroanatomic mapping. Radiofrequency ablations were performed to characterize ablation parameters and reproducibility. Intramyocardial saline irrigation began 1 minute before ablation and continued throughout. Lesions were analyzed on cardiac magnetic resonance and necropsy. RESULTS: VINTAGE was feasible in all animals within naive and infarcted myocardium. Forty-three lesions were created, using various guidewires and power settings. Forty-one (95%) lesions were detected on cardiac magnetic resonance and 38 (88%) on necropsy; all undetected lesions resulted from intentionally subtherapeutic ablation energy (10 W). Larger-diameter guidewires yielded larger size lesions. Lesion volumes on necropsy were significantly larger at 20 W than 10 W (178 mm3 [Q1-Q3: 104-382 mm3] vs 49 mm3 [Q1-Q3: 35-93 mm3]; P = 0.02). Higher power (30 W) did not create larger lesions. Median impedance dropped with preablation irrigation by 12 Ω (Q1-Q3: 8-17 Ω), followed by a further 15-Ω (Q1-Q3: 11-19 Ω) drop during ablation. Intramyocardial navigation, ablation, and irrigation were not associated with any complications. CONCLUSIONS: VINTAGE was safe and effective at creating intramural ablation lesions in targets traditionally considered inaccessible from the endocardium and epicardium, both naive and infarcted. Intramyocardial guidewire irrigation and ablation at 20 W creates reproducibly large intramural lesions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiovascular Branch, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland, USA.
- Cardiovascular Branch, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland, USA; Emory University Hospital, Atlanta, Georgia, USA.
- Division of Cardiology, The Johns Hopkins University School of Medicine, Baltimore, Maryland, USA.
- Emory University Hospital, Atlanta, Georgia, USA.
- Cardiovascular Branch, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland, USA; MedStar Washington Hospital Center, Washington, DC, USA.
- Cardiovascular Branch, Division of Intramural Research, National Heart Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland, USA; St Francis Hospital, Roslyn, New York, USA.
Förderung
- Intramural NIH HHS
- National Heart, Lung, and Blood Institute
- NHLBI DIR
Zitieren
APA
Halaby, R. N., Bruce, C. G., Kolandaivelu, A., Bhatia, N. K., Rogers, T., Khan, J. M., Yildirim, D. K., Jaimes, A. E., O'Brien, K., Babaliaros, V. C., Greenbaum, A. B., & Lederman, R. J. (2024). Ventricular Intramyocardial Navigation for Tachycardia Ablation Guided by Electrograms (VINTAGE). JACC: Clinical Electrophysiology. https://doi.org/10.1016/j.jacep.2024.04.002
Vancouver
Halaby RN, Bruce CG, Kolandaivelu A, Bhatia NK, Rogers T, Khan JM, et al. Ventricular Intramyocardial Navigation for Tachycardia Ablation Guided by Electrograms (VINTAGE). JACC: Clinical Electrophysiology. 2024. doi:10.1016/j.jacep.2024.04.002
BibTeX
@article{Halaby2024,
author = {Halaby, Rim N and Bruce, Christopher G and Kolandaivelu, Aravindan and Bhatia, Neal K and Rogers, Toby and Khan, Jaffar M and Yildirim, D Korel and Jaimes, Andi E and O'Brien, Kendall and Babaliaros, Vasilis C and Greenbaum, Adam B and Lederman, Robert J},
title = {Ventricular Intramyocardial Navigation for Tachycardia Ablation Guided by Electrograms (VINTAGE)},
journal = {JACC: Clinical Electrophysiology},
year = {2024},
doi = {10.1016/j.jacep.2024.04.002}
}
EUVIMED