Multicenter End-to-End Evaluation of Single-Isocenter Multitarget Stereotactic Radiosurgery Approaches: Accuracy and Plan Quality

Benedikt Thomann, Tobias Fechter, Johannes Fischer, Armin Runz, Ute Ludwig, Ilias Sachpazidis, Oliver Blanck, Julian Roers, Melanie Grehn, Maximilian Grohmann, Christian Ziemann, Michael Judge und 26 weitere

International Journal of Radiation Oncology*Biology*Physics · 2026-10 · 126 (2) · 520-532

DOI: 10.1016/j.ijrobp.2026.03.055 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift International Journal of Radiation Oncology*Biology*Physics, 2026-10. Verfasst von Benedikt Thomann und 37 weiteren Autor:innen. Beteiligte Einrichtung: Division of Medical Physics, Department of Radiation Oncology, Medical Center - University of Freiburg, Freiburg, Germany; Faculty of Medicine, Unive…. Hauptthemen laut MEDLINE: Radiosurgery, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Brain Neoplasms. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: German Cancer Aid.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: We conducted a multicenter study on single-isocenter multitarget stereotactic radiosurgery to dosimetrically assess end-to-end test results and identify approaches and techniques influencing spatial accuracy and treatment plan quality. METHODS AND MATERIALS: An anthropomorphic head phantom with radiochromic film and polymer gel inserts was used with a reference structure set of 5 brain metastases. End-to-end tests were performed on-site at 23 centers in Germany, Austria, and Switzerland, each following its own single-isocenter multitarget stereotactic radiosurgery protocol. Spatial accuracy was quantified by comparing planned and measured prescription isodose-volume centroids. Plan quality was assessed from treatment planning system calculations using the Paddick gradient index (GI) and Paddick conformity index. Statistical analyses, including a generalized linear model, correlated results with protocol parameters to identify favorable systems and techniques. RESULTS: The mean spatial offset between measured and calculated prescription isodose centroids across all centers and targets was 0.9 ± 0.4(1σ) mm. Offsets above 1 mm were observed in 33% of centers. Better imaging-to-radiation isocenter consistency (ICC) yielded significantly higher accuracy (P = .002): dICC

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryPhantoms, ImagingRadiotherapy Planning, Computer-AssistedBrain NeoplasmsHumansAustriaGermanySwitzerlandRadiotherapy DosageLinear Models

Autor:innen und Institutionen

  • Division of Medical Physics, Department of Radiation Oncology, Medical Center - University of Freiburg, Freiburg, Germany; Faculty of Medicine, University of Freiburg, Freiburg, Germany; German Cancer Consortium (DKTK),…
  • Faculty of Medicine, University of Freiburg, Freiburg, Germany; Division of Medical Physics, Department of Diagnostic and Interventional Radiology, University Medical Center Freiburg, Freiburg, Germany.
  • Medical Physics in Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, National Center for Radiation Research in Oncology (NCRO), Heidelberg Institute for Radiation Oncology (HIRO), Heidelberg, Germany.
  • Department of Radiation Oncology, University Medical Center Schleswig-Holstein, Campus Kiel, Saphir Radiosurgery Center Frankfurt and Northern Germany, Kiel, Germany.
  • Department of Radiotherapy and Radiation Oncology, University Hospital Münster, Münster, Germany.
  • Department of Radiotherapy and Radiation Oncology, University Hospital Hamburg-Eppendorf, Hamburg, Germany.

Förderung

  • German Cancer Aid

Zitieren

APA

Thomann, B., Fechter, T., Fischer, J., Runz, A., Ludwig, U., Sachpazidis, I., Blanck, O., Roers, J., Grehn, M., Grohmann, M., Ziemann, C., Judge, M., Baus, W., Grahle, M., Walke, M., Bathen, B., Köhn, J., Käthner, P., Shariff, M., …, & Baltas, D. (2026). Multicenter End-to-End Evaluation of Single-Isocenter Multitarget Stereotactic Radiosurgery Approaches: Accuracy and Plan Quality. International Journal of Radiation Oncology*Biology*Physics, 126(2), 520-532. https://doi.org/10.1016/j.ijrobp.2026.03.055

Vancouver

Thomann B, Fechter T, Fischer J, Runz A, Ludwig U, Sachpazidis I, et al. Multicenter End-to-End Evaluation of Single-Isocenter Multitarget Stereotactic Radiosurgery Approaches: Accuracy and Plan Quality. International Journal of Radiation Oncology*Biology*Physics. 2026;126(2):520-532. doi:10.1016/j.ijrobp.2026.03.055

BibTeX
@article{Thomann2026,
  author = {Thomann, Benedikt and Fechter, Tobias and Fischer, Johannes and Runz, Armin and Ludwig, Ute and Sachpazidis, Ilias and Blanck, Oliver and Roers, Julian and Grehn, Melanie and Grohmann, Maximilian and Ziemann, Christian and Judge, Michael and Baus, Wolfgang and Grahle, Michelle and Walke, Matthias and Bathen, Bastian and Köhn, Janett and Käthner, Paul and Shariff, Maya and Wegner, Nils and Fleckenstein, Jens and Karle, Heiko and Streller, Tino and Howitz, Simon and Priegnitz, Marlen and Weigel, Rocco and Konrad, Thomas and Schmitt, Daniela and Beck, Jürgen and Machein, Marcia and Popp, Ilinca and Reiner, Michael and Moustakis, Christos and Karger, Christian P and Pappas, Evangelos and Grosu, Anca-Ligia and Bock, Michael and Baltas, Dimos},
  title = {Multicenter End-to-End Evaluation of Single-Isocenter Multitarget Stereotactic Radiosurgery Approaches: Accuracy and Plan Quality},
  journal = {International Journal of Radiation Oncology*Biology*Physics},
  year = {2026},
  volume = {126},
  number = {2},
  pages = {520--532},
  doi = {10.1016/j.ijrobp.2026.03.055}
}
Kostenlos starten →