Development and multi-institutional evaluation of a new phantom for verifying beam-positioning errors at off-isocenter positions
Physica Medica · 2023
DOI: 10.1016/j.ejmp.2023.102645 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Physica Medica, 2023. Verfasst von Takahisa Kido und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, Kyoto, Japan. Hauptthemen laut MEDLINE: Radiosurgery, Brain Neoplasms. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 22 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: Single-isocenter stereotactic radiotherapy for multiple brain metastases requires highly accurate treatment delivery at off-isocenter positions (off-iso). This study aimed to verify the beam-positioning errors at off-iso using a newly developed phantom tested at multiple institutions. METHODS: The off-iso phantom comprised five stainless-steel balls with a 3-mm diameter placed at the center and at four peripheral positions on a diagonal line. Each ball was placed 3.5 cm apart along each of the three axes. Two patterns of the phantom setup were defined as 0° and 90° phantom rotations to evaluate the beam-positioning error, which is the distance between the center of the ball and the irradiated field on the electronic portal imaging device. Furthermore, the reproducibility of the beam-positioning errors was verified by evaluating their standard deviation (SD) at a single institution, which included five measurements for two treatment machines. The errors were evaluated at multiple institutions using eight treatment machines. RESULTS: The measurement time from setup to image acquisition was approximately 20 min for two patterns. The SD of the beam-positioning errors in the reproducibility tests was 0.41 mm. In the multi-institutional evaluation, the beam-positioning error at the isocenter position was within 1.00 mm of the AAPM-RSS tolerance, with the exception of two linacs. The largest beam-positioning error (1.36 mm) was observed 7.5 cm away from the isocenter in three directions at a gantry angle of 180°. CONCLUSIONS: The developed phantom can be applied as a new tool for establishing beam-positioning errors in single-isocenter stereotactic radiotherapy at off-isocenter positions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, Kyoto, Japan.
- Department of Radiation Oncology and Image-Applied Therapy, Kyoto University, Kyoto, Japan.
- Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, Kyoto, Japan. Electronic address: m_nkmr@kuhp.kyoto-u.ac.jp.
- Department of Radiation Oncology, Division of Medical Physics, Japanese Red Cross Wakayama Medical Center, Japan.
- Department of Technology, Division of Medical Technology, Medical Physics Office, Yamatotakada Municipal Hospital, Japan.
- Department of Radiation Oncology, Shiga General Hospital, Japan.
Zitieren
APA
Kido, T., Ono, T., Nakamura, M., Ishihara, Y., Itoh, H., Matsugi, K., Yoshimoto, A., Kishigami, Y., & Mizowaki, T. (2023). Development and multi-institutional evaluation of a new phantom for verifying beam-positioning errors at off-isocenter positions. Physica Medica. https://doi.org/10.1016/j.ejmp.2023.102645
Vancouver
Kido T, Ono T, Nakamura M, Ishihara Y, Itoh H, Matsugi K, et al. Development and multi-institutional evaluation of a new phantom for verifying beam-positioning errors at off-isocenter positions. Physica Medica. 2023. doi:10.1016/j.ejmp.2023.102645
BibTeX
@article{Kido2023,
author = {Kido, Takahisa and Ono, Tomohiro and Nakamura, Mitsuhiro and Ishihara, Yoshitomo and Itoh, Hiroyuki and Matsugi, Kiyotomo and Yoshimoto, Atsushi and Kishigami, Yukako and Mizowaki, Takashi},
title = {Development and multi-institutional evaluation of a new phantom for verifying beam-positioning errors at off-isocenter positions},
journal = {Physica Medica},
year = {2023},
doi = {10.1016/j.ejmp.2023.102645}
}
EUVIMED