EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

Effect of lead body and helix design variables on implantation success, insertion depth, and muscle torque in the left bundle branch area: Insights from an ex vivo porcine model

Ankur R Shah, Alex Puccio, Kyoichiro Yazaki, Emmanuel Offei, Martha Sofia Ruiz Castilo, Surachat Jaroonpipatkul, Ava Yaktaeian Vaziri, Muhammad S Khan, Ravi Ranjan, Robert Hitchcock, Derek J Dosdall

Heart Rhythm · 2026

DOI: 10.1016/j.hrthm.2026.02.034 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Heart Rhythm, 2026. Verfasst von Ankur R Shah und 10 weiteren Autor:innen. Beteiligte Einrichtung: Nora Eccles Harrison Cardiovascular Research and Training Institute. Hauptthemen laut MEDLINE: Bundle of His, Pacemaker, Artificial, Cardiac Pacing, Artificial, Electrodes, Implanted. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NHLBI NIH HHS, National Heart Lung and Blood Institute, Heart Rhythm Society.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Lumenless and stylet-driven leads used for left bundle branch area pacing differ in design and have a significant implantation learning curve. Although previous studies examined longer helices for deep-septal pacing, the influence of other design variables remains unclear. OBJECTIVE: This study aimed to evaluate how helix design and axial force affect interventricular septum insertion efficacy. METHODS: Rigid leads were developed using helical coils with variable outer diameter, number of turns, and pitch. Porcine septa (n = 16) were clamped perpendicularly for insertion using an optimized rotation-response system. Leads were tested under low (30g) and high (60g) axial force conditions to represent lumenless and stylet-driven leads, respectively, and a fixed number of rotations were delivered at a constant rate. Each helix design (n = 8) was tested 3× per axial force at 3 septal sites. Insertion depth, muscle torque, and visual feedback were recorded. Insertion was successful if the depth exceeded coil length without surface entanglement. Effects of design factors were compared. RESULTS: At 30g, more helix turns significantly improved insertion success (P = .04), whereas fewer turns frequently produced entangled failure (P = .04) marked by high torque variability (P < .001). Smaller-pitch helices trended toward higher torque and success, whereas a larger pitch achieved greater depth (P = .05). Larger outer diameters also trended toward higher torque and improved success at 30g. At 60g, the influence of helix design variable diminished and consistently yielded higher values than at 30g. CONCLUSION: An optimized lead rotation-to-translation system elucidates how helix geometry and axial force interact during septal insertion. These interactions are explainable using an intuitive mechanical framework, which helps optimize lead design.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bundle of HisPacemaker, ArtificialCardiac Pacing, ArtificialElectrodes, ImplantedAnimalsSwineTorqueEquipment Design

Stichwörter der Autor:innen

Ex vivoHelix designInterventricular septumLead behaviorLeft bundle branch areaPacing leads

Autor:innen und Institutionen

  • Nora Eccles Harrison Cardiovascular Research and Training Institute.
  • Department of Biomedical Engineering.
  • Nora Eccles Harrison Cardiovascular Research and Training Institute; Department of Biomedical Engineering.
  • Nora Eccles Harrison Cardiovascular Research and Training Institute; Division of Cardiothoracic Surgery, Department of Surgery.
  • Nora Eccles Harrison Cardiovascular Research and Training Institute; Department of Biomedical Engineering; Division of Cardiovascular Medicine, Department of Internal Medicine, The University of Utah, Salt Lake City, Ut…
  • Nora Eccles Harrison Cardiovascular Research and Training Institute; Department of Biomedical Engineering; Division of Cardiothoracic Surgery, Department of Surgery; Division of Cardiovascular Medicine, Department of In…

Förderung

  • NHLBI NIH HHS
  • National Heart Lung and Blood Institute
  • Heart Rhythm Society
  • Nora Eccles Treadwell Foundation
  • American Heart Association Inc

Zitieren

APA

Shah, A. R., Puccio, A., Yazaki, K., Offei, E., Ruiz Castilo, M. S., Jaroonpipatkul, S., Yaktaeian Vaziri, A., Khan, M. S., Ranjan, R., Hitchcock, R., & Dosdall, D. J. (2026). Effect of lead body and helix design variables on implantation success, insertion depth, and muscle torque in the left bundle branch area: Insights from an ex vivo porcine model. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2026.02.034

Vancouver

Shah AR, Puccio A, Yazaki K, Offei E, Ruiz Castilo MS, Jaroonpipatkul S, et al. Effect of lead body and helix design variables on implantation success, insertion depth, and muscle torque in the left bundle branch area: Insights from an ex vivo porcine model. Heart Rhythm. 2026. doi:10.1016/j.hrthm.2026.02.034

BibTeX
@article{Shah2026,
  author = {Shah, Ankur R and Puccio, Alex and Yazaki, Kyoichiro and Offei, Emmanuel and Ruiz Castilo, Martha Sofia and Jaroonpipatkul, Surachat and Yaktaeian Vaziri, Ava and Khan, Muhammad S and Ranjan, Ravi and Hitchcock, Robert and Dosdall, Derek J},
  title = {Effect of lead body and helix design variables on implantation success, insertion depth, and muscle torque in the left bundle branch area: Insights from an ex vivo porcine model},
  journal = {Heart Rhythm},
  year = {2026},
  doi = {10.1016/j.hrthm.2026.02.034}
}
Kostenlos starten →