Safety and performance of the Medtronic 3830 lead in His-bundle and Left bundle branch area pacing: A single-center experience

Alejandro Sanchez-Nadales, Abdullah Sarkar, Jose Sleiman, Andres Sanchez-Nadales, Mileydis Alonso, John Bibawy, Marcelo Helguera, Sergio Pinski, Jose Baez-Escudero

Journal of Interventional Cardiac Electrophysiology · 2025

DOI: 10.1007/s10840-025-02070-3 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Interventional Cardiac Electrophysiology, 2025. Verfasst von Alejandro Sanchez-Nadales und 8 weiteren Autor:innen. Beteiligte Einrichtung: Robert and Suzanne Tomsich Department of Cardiology, Department of Cardiovascular Disease, Cleveland Clinic Florida, Weston Campus, Weston, FL, 33324…. Hauptthemen laut MEDLINE: Bundle of His, Cardiac Pacing, Artificial, Bundle-Branch Block, Pacemaker, Artificial. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Conduction system pacing (CSP) using His Bundle Pacing (HBP) and Left Bundle Branch Area Pacing (LBAP) is an evolving alternative to traditional right ventricular pacing (RVP), promising better physiological outcomes. This study evaluates the safety, feasibility, and performance of HBP and LBAP with Medtronic SelectSecure 3830 leads.

Methods

We conducted a single-center retrospective analysis of 490 patients undergoing HBP or LBAP. The study assessed implant success rates, pacing thresholds, device longevity, and complication rates over an average follow-up of 28 months for HBP and 14 months for LBAP.

Results

The implantation success rate was 85% for HBP and 97.4% for LBAP. LBAP demonstrated lower and more stable pacing thresholds, with initial values of 0.8V at 0.5 ms rising slightly to 0.9V at 0.5 ms, and fewer device revisions compared to HBP, whose initial pacing threshold of 1.3V at 0.8ms increased to 1.68 V at 0.7ms. Complications were minimal and similar across both groups. The need for fewer device revisions and potential for prolonged device life highlighted LBAP as potentially more cost-effective. Cardiac function measured by LVEF remained stable across both groups.

Conclusions

Both HBP and LBAP are safe and feasible with comparable safety profiles. LBAP may offer advantages in terms of stability, fewer revisions, and extended device longevity. The study underscores the need for further research into optimal lead positioning and long-term outcomes of CSP, particularly for LBAP. Graphical Abstract

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bundle of HisCardiac Pacing, ArtificialBundle-Branch BlockPacemaker, ArtificialElectrodes, ImplantedHumansMaleRetrospective StudiesFemaleAgedTreatment OutcomeMiddle AgedEquipment DesignFeasibility StudiesRisk Assessment

Stichwörter der Autor:innen

Conduction system pacingHis bundle pacingLeft bundle branch area pacingSelectSecure 3830

Autor:innen und Institutionen

  • Robert and Suzanne Tomsich Department of Cardiology, Department of Cardiovascular Disease, Cleveland Clinic Florida, Weston Campus, Weston, FL, 33324, USA. asanc86@uic.edu.
  • Robert and Suzanne Tomsich Department of Cardiology, Department of Cardiovascular Disease, Cleveland Clinic Florida, Weston Campus, Weston, FL, 33324, USA.

Zitieren

APA

Sanchez-Nadales, A., Sarkar, A., Sleiman, J., Sanchez-Nadales, A., Alonso, M., Bibawy, J., Helguera, M., Pinski, S., & Baez-Escudero, J. (2025). Safety and performance of the Medtronic 3830 lead in His-bundle and Left bundle branch area pacing: A single-center experience. Journal of Interventional Cardiac Electrophysiology. https://doi.org/10.1007/s10840-025-02070-3

Vancouver

Sanchez-Nadales A, Sarkar A, Sleiman J, Sanchez-Nadales A, Alonso M, Bibawy J, et al. Safety and performance of the Medtronic 3830 lead in His-bundle and Left bundle branch area pacing: A single-center experience. Journal of Interventional Cardiac Electrophysiology. 2025. doi:10.1007/s10840-025-02070-3

BibTeX
@article{SanchezNadales2025,
  author = {Sanchez-Nadales, Alejandro and Sarkar, Abdullah and Sleiman, Jose and Sanchez-Nadales, Andres and Alonso, Mileydis and Bibawy, John and Helguera, Marcelo and Pinski, Sergio and Baez-Escudero, Jose},
  title = {Safety and performance of the Medtronic 3830 lead in His-bundle and Left bundle branch area pacing: A single-center experience},
  journal = {Journal of Interventional Cardiac Electrophysiology},
  year = {2025},
  doi = {10.1007/s10840-025-02070-3}
}
Kostenlos starten →