Guided implantation of a leadless left ventricular endocardial electrode and acoustic transmitter using computed tomography anatomy, dynamic perfusion and mechanics, and predicted activation pattern

Baldeep S Sidhu, Angela W C Lee, Justin Gould, Bradley Porter, Benjamin Sieniewicz, Mark K Elliott, Vishal S Mehta, Nadeev Wijesuriya, Abdoul A Amadou, Gernot Plank, Ulrike Haberland, Ronak Rajani und 2 weitere

Heart Rhythm · 2023

DOI: 10.1016/j.hrthm.2023.07.007 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Heart Rhythm, 2023. Verfasst von Baldeep S Sidhu und 13 weiteren Autor:innen. Beteiligte Einrichtung: School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom; Guy's and St Thomas' NHS Foundation Trust, Lond…. Hauptthemen laut MEDLINE: Heart Failure, Cardiac Resynchronization Therapy. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: British Heart Foundation, Centre For Medical Engineering, King’s College London, Abbott Laboratories.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: The WiSE-CRT System (EBR systems, Sunnyvale, CA) permits leadless left ventricular pacing. Currently, no intraprocedural guidance is used to target optimal electrode placement while simultaneously guiding acoustic transmitter placement in close proximity to the electrode to ensure adequate power delivery. OBJECTIVE: The purpose of this study was to assess the use of computed tomography (CT) anatomy, dynamic perfusion and mechanics, and predicted activation pattern to identify both the optimal electrode and transmitter locations. METHODS: A novel CT protocol was developed using preprocedural imaging and simulation to identify target segments (TSs) for electrode implantation, with late electrical and mechanical activation, with ≥5 mm wall thickness without perfusion defects. Modeling of the acoustic intensity from different transmitter implantation sites to the TSs was used to identify the optimal transmitter location. During implantation, TSs were overlaid on fluoroscopy to guide optimal electrode location that were evaluated by acute hemodynamic response (AHR) by measuring the maximal rate of left ventricular pressure rise with biventricular pacing. RESULTS: Ten patients underwent the implantation procedure. The transmitter could be implanted within the recommended site on the basis of preprocedural analysis in all patients. CT identified a mean of 4.8 ± 3.5 segments per patient with wall thickness < 5 mm. During electrode implantation, biventricular pacing within TSs resulted in a significant improvement in AHR vs non-TSs (25.5% ± 8.8% vs 12.9% ± 8.6%; P < .001). Pacing in CT-identified scar resulted in either failure to capture or minimal AHR improvement. The electrode was targeted to the TSs in all patients and was implanted in the TSs in 80%. CONCLUSION: Preprocedural imaging and modeling data with intraprocedural guidance can successfully guide WiSE-CRT electrode and transmitter implantation to allow optimal AHR and adequate power delivery.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart FailureCardiac Resynchronization TherapyHumansCardiac Resynchronization Therapy DevicesElectrodesTomography, X-Ray ComputedPerfusionTreatment OutcomeHeart Ventricles

Stichwörter der Autor:innen

Cardiac resynchronization therapyEndocardial pacingGuiding endocardial pacingHemodynamic assessmentWiSE-CRT

Autor:innen und Institutionen

  • School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom; Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom. Electronic address: Baldeep.sidhu@nhs.net.uk.
  • School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom.
  • School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom; Guy's and St Thomas' NHS Foundation Trust, London, United Kingdom.
  • Siemens Healthcare GmbH, Forchheim, Germany.
  • Medical University of Graz, Graz, Austria.
  • School of Biomedical Engineering and Imaging Sciences, King's College London, London, United Kingdom; Siemens Healthcare GmbH, Forchheim, Germany.

Förderung

  • British Heart Foundation
  • Centre For Medical Engineering, King’s College London
  • Abbott Laboratories
  • National Institute for Health and Care Research

Zitieren

APA

Sidhu, B. S., Lee, A. W. C., Gould, J., Porter, B., Sieniewicz, B., Elliott, M. K., Mehta, V. S., Wijesuriya, N., Amadou, A. A., Plank, G., Haberland, U., Rajani, R., Rinaldi, C. A., & Niederer, S. A. (2023). Guided implantation of a leadless left ventricular endocardial electrode and acoustic transmitter using computed tomography anatomy, dynamic perfusion and mechanics, and predicted activation pattern. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2023.07.007

Vancouver

Sidhu BS, Lee AWC, Gould J, Porter B, Sieniewicz B, Elliott MK, et al. Guided implantation of a leadless left ventricular endocardial electrode and acoustic transmitter using computed tomography anatomy, dynamic perfusion and mechanics, and predicted activation pattern. Heart Rhythm. 2023. doi:10.1016/j.hrthm.2023.07.007

BibTeX
@article{Sidhu2023,
  author = {Sidhu, Baldeep S and Lee, Angela W C and Gould, Justin and Porter, Bradley and Sieniewicz, Benjamin and Elliott, Mark K and Mehta, Vishal S and Wijesuriya, Nadeev and Amadou, Abdoul A and Plank, Gernot and Haberland, Ulrike and Rajani, Ronak and Rinaldi, Christopher A and Niederer, Steven A},