Implementation of lead V8 as a simple non-invasive tool to improve patient selection in CRT and guide LV lead placement

Jacqueline Joza iD, Johan van Koll iD, Vadivelu Ramalingam iD, Peter Tai, Vidal Essebag iD, Tomy A Hadjis, Theo A R Lankveld, Antonius M W van Stipdonk iD, Jesse H J Rijks iD, Uyên C Nguyên iD, Justin Luermans iD, Atul Verma iD und 1 weitere

Europace · 2025

DOI: 10.1093/europace/euaf194 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Europace, 2025. Verfasst von Jacqueline Joza und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medicine, McGill University Health Center, Montréal, Québec, Canada H4A 3J1. Hauptthemen laut MEDLINE: Cardiac Resynchronization Therapy, Heart Failure, Patient Selection, Cardiac Resynchronization Therapy Devices. Die Arbeit wurde bisher 1-mal zitiert (Crossref). Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Aims CRT improves outcomes in heart failure patients with electrical dyssynchrony. The QLV interval, a marker of delayed left ventricular (LV) activation, independently predicts CRT response. This study aimed to determine whether activation time measured in surface ECG lead V8 reflects postero-lateral LV delay by correlating with the QLV interval during biventricular (BiV) CRT implantation.

Methods

Consecutive patients with wide QRS (non-RBBB) undergoing BiV-CRT or generator change were prospectively enrolled. Surface ECGs were recorded with the V5 electrode repositioned to the V8 location. Local wavefront activation was determined using the steepest negative downslope of the precordial leads, referred to as the negative derivative activation time (NDAT). The LV lead was connected to the EP recording system, and NDAT and QLV measurements were obtained.

Results

Forty-three patients were included. The NDAT in V8 strongly correlated with the QLV interval (r = 0.895; mean difference 1 ± 11 ms) across both LBBB and IVCD patterns. In contrast, NDAT in V6 showed weaker correlation (r = 0.592; mean difference 21 ± 24 ms). During LV-only pacing, a QS morphology in V8 consistently indicated postero-lateral lead placement. More anterior/apical lead positions showed larger QLV–NDAT V8 differences and non-QS morphologies, indicating suboptimal positioning.

Conclusion

NDAT in lead V8 is a non-invasive marker of delayed activation in the postero-lateral LV, outperforming V6 in identifying late activation. When combined with LV-paced QRS morphology, V8 facilitates accurate lead localization without additional tools. These findings may support the use of lead V8 to refine patient selection and optimize CRT delivery.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cardiac Resynchronization TherapyHeart FailurePatient SelectionCardiac Resynchronization Therapy DevicesElectrocardiographyVentricular Function, LeftHumansMaleFemaleAgedMiddle AgedProspective StudiesTreatment OutcomeAction PotentialsPredictive Value of TestsTime FactorsHeart Ventricles

Stichwörter der Autor:innen

Biventricular pacingCardiac resynchronization therapyConduction system pacingIntraventricular conduction delayLOT-CRTLeft bundle branch blockQLV

Autor:innen und Institutionen

  • Department of Medicine, McGill University Health Center, Montréal, Québec, Canada H4A 3J1.
  • Department of Cardiology, Cardiovascular Research Institute Maastricht (CARIM), Maastricht University Medical Center, Maastricht, The Netherlands.

ORCID-Profile: Jacqueline Joza, Johan van Koll, Vadivelu Ramalingam, Vidal Essebag, Antonius M W van Stipdonk, Jesse H J Rijks, Uyên C Nguyên, Justin Luermans, Atul Verma, Kevin Vernooy

Zitieren

APA

Joza, J., van Koll, J., Ramalingam, V., Tai, P., Essebag, V., Hadjis, T. A., Lankveld, T. A. R., van Stipdonk, A. M. W., Rijks, J. H. J., Nguyên, U. C., Luermans, J., Verma, A., & Vernooy, K. (2025). Implementation of lead V8 as a simple non-invasive tool to improve patient selection in CRT and guide LV lead placement. Europace. https://doi.org/10.1093/europace/euaf194

Vancouver

Joza J, van Koll J, Ramalingam V, Tai P, Essebag V, Hadjis TA, et al. Implementation of lead V8 as a simple non-invasive tool to improve patient selection in CRT and guide LV lead placement. Europace. 2025. doi:10.1093/europace/euaf194

BibTeX
@article{Joza2025,
  author = {Joza, Jacqueline and van Koll, Johan and Ramalingam, Vadivelu and Tai, Peter and Essebag, Vidal and Hadjis, Tomy A and Lankveld, Theo A R and van Stipdonk, Antonius M W and Rijks, Jesse H J and Nguyên, Uyên C and Luermans, Justin and Verma, Atul and Vernooy, Kevin},
  title = {Implementation of lead V8 as a simple non-invasive tool to improve patient selection in CRT and guide LV lead placement},
  journal = {Europace},
  year = {2025},
  doi = {10.1093/europace/euaf194}
}
Kostenlos starten →