Safety evaluation of smart scales, smart watches, and smart rings with bioimpedance technology shows evidence of potential interference in cardiac implantable electronic devices
Heart Rhythm · 2023
DOI: 10.1016/j.hrthm.2022.11.026 ↗Kurz gefasst
Fachartikel in der Zeitschrift Heart Rhythm, 2023. Verfasst von Gia-Bao Ha und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, University of Utah, Salt Lake City, Utah. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Defibrillators, Implantable. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 24 Literaturangaben. Gefördert von: NHLBI NIH HHS, National Heart, Lung, and Blood Institute, National Heart Lung and Blood Institute Division of Intramural Research.
Abstract (Originalsprache)
BACKGROUND: Smart scales, smart watches, and smart rings with bioimpedance technology may create interference in patients with cardiac implantable electronic devices (CIEDs). OBJECTIVES: The purpose of this study was to determine interference at CIEDs with simulations and benchtop testing, and to compare the results with maximum values defined in the ISO 14117 electromagnetic interference standard for these devices. METHODS: The interference at pacing electrodes was determined by simulations on a male and a female computable model. A benchtop evaluation of representative CIEDs from 3 different manufacturers as specified in the ISO 14117 standard also was performed. RESULTS: Simulations showed evidence of interference with voltage values exceeding threshold values defined in the ISO 14117 standard. The level of interference varied with the frequency and amplitude of the bioimpedance signal, and between male and female models. The level of interference generated with smart scale and smart rings simulations was lower than with smart watches. Across device manufacturers, generators demonstrated susceptibility to oversensing and pacing inhibition at different signal amplitudes and frequencies. CONCLUSIONS: This study evaluated the safety of smart scales, smart watches, and smart rings with bioimpedance technology via simulation and testing. Our results indicate that these consumer electronic devices could interfere in patients with CIEDs. The present findings do not recommend the use of these devices in this population due to potential interference.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Electrical and Computer Engineering, University of Utah, Salt Lake City, Utah.
- Department of Medicine, University of Utah Health Sciences Center, Salt Lake City, Utah.
- Department of Cardiology, Heart Institute, Hospital Universitari Germans Trias i Pujol, Badalona, Spain; Centro de Investigación Biomédica en Red Enfermedades Cardiovasculares (CIBERCV), Madrid, Spain; Autonomous Univer…
- Department of Electrical and Computer Engineering, University of Utah, Salt Lake City, Utah. Electronic address: benjamin.sanchez@utah.edu.
Förderung
- NHLBI NIH HHS
- National Heart, Lung, and Blood Institute
- National Heart Lung and Blood Institute Division of Intramural Research
- National Institutes of Health
Zitieren
APA
Ha, G. B., Steinberg, B. A., Freedman, R., Bayés-Genís, A., & Sanchez, B. (2023). Safety evaluation of smart scales, smart watches, and smart rings with bioimpedance technology shows evidence of potential interference in cardiac implantable electronic devices. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2022.11.026
Vancouver
Ha GB, Steinberg BA, Freedman R, Bayés-Genís A, Sanchez B. Safety evaluation of smart scales, smart watches, and smart rings with bioimpedance technology shows evidence of potential interference in cardiac implantable electronic devices. Heart Rhythm. 2023. doi:10.1016/j.hrthm.2022.11.026
BibTeX
@article{Ha2023,
author = {Ha, Gia-Bao and Steinberg, Benjamin A and Freedman, Roger and Bayés-Genís, Antoni and Sanchez, Benjamin},
title = {Safety evaluation of smart scales, smart watches, and smart rings with bioimpedance technology shows evidence of potential interference in cardiac implantable electronic devices},
journal = {Heart Rhythm},
year = {2023},
doi = {10.1016/j.hrthm.2022.11.026}
}
EUVIMED