Impact of the measurement conditions and compression paddle on mammography dosimeter response
Physica Medica · 2024-07 · 123 · 103405
DOI: 10.1016/j.ejmp.2024.103405 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physica Medica, 2024-07. Verfasst von Maysa Costa de Castro und 8 weiteren Autor:innen. Beteiligte Einrichtung: STUK - Radiation and Nuclear Authority, Jokiniemenkuja 1, 01370 Vantaa, Finland. Hauptthemen laut MEDLINE: Mammography. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
The effect of mammography measurement conditions was investigated to evaluate their impact on measurement uncertainties in clinical practice. The most prominent physical X-ray beam quantities i.e., - air kerma, half-value layer, and X-ray tube voltage - were examined by measuring the response of two ionization chambers and six X-ray multimeters (XMMs) of different models. Measurements were performed using several anode/filter combinations and both with and without the compression paddle in the X-ray beam. Maximum differences of higher than 6 % were found for all quantities when the dosimeter displayed value was compared with the reference value or the variation within the clinical anode/filter combinations Mo/Mo and Mo/Rh were considered. The study showed that the calibration procedure with the W/Al anode/filter combination was reliable only for ionization chambers, and the response of XMMs varies in such a way that the calibration coefficient cannot be predicted between various measurement conditions used in calibration and clinical practices. XMM calibrations are typically performed without a compression paddle in the beam, and the response of the XMM changes when radiation quality is slightly altered. If XMM specific data is not available, based on this study, an additional uncertainty of 2 % (k = 1) could be used as a typical estimate, at least for air kerma measurements. XMMs should be used for clinical measurements in mammography only with correct settings. If the correct settings are not available, the XMMs should not be used or used only with extreme caution.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- STUK - Radiation and Nuclear Authority, Jokiniemenkuja 1, 01370 Vantaa, Finland.
- STUK - Radiation and Nuclear Authority, Jokiniemenkuja 1, 01370 Vantaa, Finland; HUS HUS Diagnostic Center, Radiology, University of Helsinki and Helsinki University Hospital, Finland. Electronic address: paula.toroi@st…
- HUS HUS Diagnostic Center, Radiology, University of Helsinki and Helsinki University Hospital, Finland.
- ENEA-INMRI, Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, Via Anguillarese 301, 00123 Santa Maria di Galeria, (RM), Italy.
- Vinca Institute of Nuclear Sciences - National Institute of the Republic of Serbia, University of Belgrade, Mike Petrovica Alasa 12-14, 11351 Vinča, Belgrade, Serbia.
Zitieren
APA
Costa de Castro, M., Toroi, P., Kortesniemi, M., Lindholm, C., Pinto, M., Rinaldi, L., Tietäväinen, A., Tikkanen, J., & Živanović, M. (2024). Impact of the measurement conditions and compression paddle on mammography dosimeter response. Physica Medica, 123, 103405. https://doi.org/10.1016/j.ejmp.2024.103405
Vancouver
Costa de Castro M, Toroi P, Kortesniemi M, Lindholm C, Pinto M, Rinaldi L, et al. Impact of the measurement conditions and compression paddle on mammography dosimeter response. Physica Medica. 2024;123:103405. doi:10.1016/j.ejmp.2024.103405
BibTeX
@article{CostadeCastro2024,
author = {Costa de Castro, Maysa and Toroi, Paula and Kortesniemi, Mika and Lindholm, Carita and Pinto, Massimo and Rinaldi, Luigi and Tietäväinen, Aino and Tikkanen, Joonas and Živanović, Miloš},
title = {Impact of the measurement conditions and compression paddle on mammography dosimeter response},
journal = {Physica Medica},
year = {2024},
volume = {123},
pages = {103405},
doi = {10.1016/j.ejmp.2024.103405}
}
EUVIMED