Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer
Physica Medica · 2024
DOI: 10.1016/j.ejmp.2023.103203 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physica Medica, 2024. Verfasst von Annele Heikkilä und 6 weiteren Autor:innen. Beteiligte Einrichtung: BioMediTech, Faculty of Medicine and Health Technology, Tampere University, Korkeakoulunkatu 3, 33720 Tampere, Finland; Department of Medical Physics…. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Unilateral Breast Neoplasms, Breast Neoplasms. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of ≥1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- BioMediTech, Faculty of Medicine and Health Technology, Tampere University, Korkeakoulunkatu 3, 33720 Tampere, Finland; Department of Medical Physics, Tampere University Hospital, P.O. Box 2000, 33521 Tampere, Finland.…
- Department of Medical Physics, Tampere University Hospital, P.O. Box 2000, 33521 Tampere, Finland; Department of Oncology, Tampere University Hospital, P.O. Box 2000, 33521 Tampere, Finland.
- Department of Medical Physics, Central Finland Health Care District, Hoitajantie 3, 40620, Jyväskylä, Finland.
- BioMediTech, Faculty of Medicine and Health Technology, Tampere University, Korkeakoulunkatu 3, 33720 Tampere, Finland; School of Electrical and Information Engineering, University of the Witwatersrand, Johannesburg, 1…
Zitieren
APA
Heikkilä, A., Boman, E., Rossi, M., Vanhanen, A., Mankinen, M., Postema, M., & Koivumäki, T. (2024). Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer. Physica Medica. https://doi.org/10.1016/j.ejmp.2023.103203
Vancouver
Heikkilä A, Boman E, Rossi M, Vanhanen A, Mankinen M, Postema M, et al. Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer. Physica Medica. 2024. doi:10.1016/j.ejmp.2023.103203
BibTeX
@article{Heikkil2024,
author = {Heikkilä, Annele and Boman, Eeva and Rossi, Maija and Vanhanen, Antti and Mankinen, Mikko and Postema, Michiel and Koivumäki, Tuomas},
title = {Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer},
journal = {Physica Medica},
year = {2024},
doi = {10.1016/j.ejmp.2023.103203}
}
EUVIMED