First evaluation of color K-edge image quality using spectral photon-counting CT combined with two contrast agents: A phantom study

Ramata Coulibaly, Antoine Robert, Angèle Houmeau, Christian A Hernandez, Agnieszka Gutwinska, Maria Nicole Antonuccio, Simon Rit, Joël Greffier, Salim A Si-Mohamed

Diagnostic and Interventional Imaging · 2026

DOI: 10.1016/j.diii.2025.11.005 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostic and Interventional Imaging, 2026. Verfasst von Ramata Coulibaly und 8 weiteren Autor:innen. Beteiligte Einrichtung: Université de Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint Etienne, CNRS, INSERM, CREATIS UMR 5220, U1206, 69100 Villeurbanne, France. Hauptthemen laut MEDLINE: Phantoms, Imaging, Contrast Media, Tomography, X-Ray Computed. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Research Council, H2020 European Research Council.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: The purpose of this study was to assess the image quality of color K-edge imaging obtained with a spectral photon-counting CT (SPCCT) scanner using a spectral phantom with a mixture of iodine-based and gadolinium-based contrast agents. MATERIALS AND METHODS: A clinical SPCCT scanner prototype was used to scan a spectral phantom. Three dedicated cavities were filled with three contrast agents including iodine alone, gadolinium alone and a mixture of both. Two concentrations of 0.5 and 2 mg/mL were evaluated using nine helical PCCT scans at 120 kVp and 150 mAs. Conventional, color iodine and color K-edge gadolinium images were obtained through a material decomposition algorithm using three basis materials (water, iodine, gadolinium). Attenuation (in Hounsfield unit [HU]), iodine and gadolinium concentrations and task-based transfer function (TTF) were measured on each cavity and image. The noise power spectrum (NPS) was calculated on the phantom's background. RESULTS: Color K-edge imaging differentiated iodine and gadolinium but underestimated their concentrations. Gadolinium concentrations were underestimated by 9.4 ± 2.2 (standard deviation [SD]) % and 9.2 ± 1.0 (SD) % for gadolinium alone, 14.9 ± 2.3 (SD) % and 11.4 ± 1.2 (SD) % for the mixture, at 0.5 and 2 mg/mL, respectively. Similar TTF values at 50 % were found for color iodine (0.43 ± 0.01 [SD] mm-1) and color K-edge gadolinium (0.45 ± 0.03 (SD) mm-1) images for respective cavities at 2 mg/mL but the lowest values were found for color K-edge gadolinium images (0.43 ± 0.01 [SD] mm-1vs. 0.29 ± 0.01 [SD] mm-1) at 0.5 mg/mL. The value of noise magnitude was 24.75 HU, 0.06 mg/mL and 0.03 mg/mL for conventional, color iodine and color K-edge gadolinium images, respectively. CONCLUSION: Color K-edge imaging helps distinguish between contrast agents while being associated with lownoise magnitude, high-frequency spatial noise and high spatial resolution.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingContrast MediaTomography, X-Ray ComputedGadoliniumIodinePhotonsHumansColorAlgorithms

Stichwörter der Autor:innen

Color K-edgeContrast agentsImage qualityPhantom studySpectral photon counting CT

Substanzen

Contrast MediaGadoliniumIodine

Autor:innen und Institutionen

  • Université de Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint Etienne, CNRS, INSERM, CREATIS UMR 5220, U1206, 69100 Villeurbanne, France.
  • CT Clinical Science, Philips, 92150 Suresnes, France.
  • Department of Medical Imaging, CHU Nîmes, Univ Montpellier, Nîmes Medical Imaging Group, EA 2992, 30029 Nîmes, France.
  • Université de Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint Etienne, CNRS, INSERM, CREATIS UMR 5220, U1206, 69100 Villeurbanne, France; Department of Radiology, Hôpital Louis Pradel, Hospices Civils de Ly…

Förderung

  • European Research Council
  • H2020 European Research Council

Zitieren

APA

Coulibaly, R., Robert, A., Houmeau, A., Hernandez, C. A., Gutwinska, A., Antonuccio, M. N., Rit, S., Greffier, J., & Si-Mohamed, S. A. (2026). First evaluation of color K-edge image quality using spectral photon-counting CT combined with two contrast agents: A phantom study. Diagnostic and Interventional Imaging. https://doi.org/10.1016/j.diii.2025.11.005

Vancouver

Coulibaly R, Robert A, Houmeau A, Hernandez CA, Gutwinska A, Antonuccio MN, et al. First evaluation of color K-edge image quality using spectral photon-counting CT combined with two contrast agents: A phantom study. Diagnostic and Interventional Imaging. 2026. doi:10.1016/j.diii.2025.11.005

BibTeX
@article{Coulibaly2026,
  author = {Coulibaly, Ramata and Robert, Antoine and Houmeau, Angèle and Hernandez, Christian A and Gutwinska, Agnieszka and Antonuccio, Maria Nicole and Rit, Simon and Greffier, Joël and Si-Mohamed, Salim A},
  title = {First evaluation of color K-edge image quality using spectral photon-counting CT combined with two contrast agents: A phantom study},
  journal = {Diagnostic and Interventional Imaging},
  year = {2026},
  doi = {10.1016/j.diii.2025.11.005}
}
Kostenlos starten →