Technical performance of dual‐contrast‐agent K‐edge imaging with four energy thresholds on a commercial photon counting detector CT

Kevin J Treb, Jeffrey F Marsh, Elisabeth R Shanblatt, Tristan Nowak, Chelsea A S Dunning, George S K Fung, Bernhard T Schmidt, Cynthia H McCollough, Shuai Leng

Medical Physics · 2025

DOI: 10.1002/mp.70063 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, 2025. Verfasst von Kevin J Treb und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Photons, Contrast Media. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, NIH HHS, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Photon counting detector CT (PCD‐CT) has the potential to sort detected x‐ray photons into more than two energy bins to perform multi‐material decomposition using K‐edge materials without additional mathematical constraints/assumptions. The availability of multiple energy thresholds in PCD‐CT opens the possibility for unique clinical applications such as simultaneous multi‐contrast‐agent imaging.

Purpose

To evaluate the technical performance of a PCD‐CT system with four energy thresholds for conducting multi‐material decomposition of iodine, gadolinium, and water after a comprehensive system calibration procedure.

Methods

Experiments were performed on a commercial PCD‐CT scanner (NAEOTOM Alpha, Siemens Healthineers) using a research mode with four energy thresholds. Material calibration was performed by scanning contrast materials with known concentration placed in various sizes of cylindrical water‐equivalent phantoms at various tube currents and positions. From these data, per‐voxel contrast models were derived for each material by multiple linear regressions up to second order, where the predictor variables were combinations of tube current, phantom diameter, and distance to isocenter. After calibration, small (20 cm diameter) and large (30 cm × 40 cm) uniform multi‐energy CT phantoms and an anthropomorphic thorax phantom (31 cm × 39 cm), each containing bone, soft tissue, and several quantitative inserts of iodine (0.2–5.0 mg/ml), gadolinium (0.3–5.0 mg/ml), and a mixture of both (5.0 mg/ml each) were scanned using the four‐energy threshold mode at 140 kV with energy thresholds set at 20, 52, 75, and 82 keV. Images for each energy threshold were reconstructed with a filtered back‐projection algorithm and a quantitative kernel (Qr44). Subsequent image‐based material decomposition was performed on individual voxels of the energy threshold images using the models derived from the calibration process to generate water, iodine, and gadolinium basis images. Quantitative material concentrations and image noise magnitudes, and noise power spectra (NPS) were measured.

Results

Iodine and gadolinium concentrations above 1.25 mg/ml were visible in the material basis images in all phantoms. Averaged quantitative accuracy for iodine and gadolinium in the small, large, and thorax phantoms were 0.39 ± 0.30 mg/ml, 0.96 ± 0.53 mg/ml, and 0.53 ± 0.35 mg/ml, respectively. In the water basis images, accuracy was within 24 HU in all phantoms. Noise magnitudes in iodine and gadolinium basis images were 2.91 and 3.52 times that of the lowest energy threshold image for the small phantom, and were 6.92 and 7.17 times that of the lowest threshold image for the large phantom. Noise texture did not change between the energy threshold and material basis images for either size phantom as measured by the NPS.

Conclusions

After a comprehensive system calibration, a commercial PCD‐CT with four energy thresholds performed multiple material decomposition of iodine, gadolinium, and water for concentrations above 1.25 mg/ml in human‐sized phantoms. Noise amplification and quantitative accuracy from the decomposition greatly depended on object size.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedPhotonsContrast MediaPhantoms, ImagingCalibrationGadoliniumWaterIodine

Stichwörter der Autor:innen

K‐edge imagingmulti‐energy CTphoton counting detector CTquantitative CT

Substanzen

Contrast MediaGadoliniumWaterIodine

Autor:innen und Institutionen

  • Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
  • Siemens Medical Solutions USA, Inc., Malvern, Pennsylvania, USA.
  • Siemens Healthineers, Forchheim, Germany.

Förderung

  • NIBIB NIH HHS
  • NIH HHS
  • National Institutes of Health

Zitieren

APA

Treb, K. J., Marsh, J. F., Shanblatt, E. R., Nowak, T., Dunning, C. A. S., Fung, G. S. K., Schmidt, B. T., McCollough, C. H., & Leng, S. (2025). Technical performance of dual‐contrast‐agent K‐edge imaging with four energy thresholds on a commercial photon counting detector CT. Medical Physics. https://doi.org/10.1002/mp.70063

Vancouver

Treb KJ, Marsh JF, Shanblatt ER, Nowak T, Dunning CAS, Fung GSK, et al. Technical performance of dual‐contrast‐agent K‐edge imaging with four energy thresholds on a commercial photon counting detector CT. Medical Physics. 2025. doi:10.1002/mp.70063

BibTeX
@article{Treb2025,
  author = {Treb, Kevin J and Marsh, Jeffrey F and Shanblatt, Elisabeth R and Nowak, Tristan and Dunning, Chelsea A S and Fung, George S K and Schmidt, Bernhard T and McCollough, Cynthia H and Leng, Shuai},
  title = {Technical performance of dual‐contrast‐agent K‐edge imaging with four energy thresholds on a commercial photon counting detector CT},
  journal = {Medical Physics},
  year = {2025},
  doi = {10.1002/mp.70063}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche