Bone remodeling of dental implant comparing between the customized titanium abutment and commercialized stock abutment and positioning of the implant placement depth: A 3D finite element analysis
Digital Dentistry Journal · 2025
DOI: 10.1016/j.ddj.2025.100018 ↗Kurz gefasst
Fachartikel in der Zeitschrift Digital Dentistry Journal, 2025. Verfasst von Poosit Thongsamak und 4 weiteren Autor:innen. Die Arbeit nennt 65 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Chiang Mai University.
Abstract (Originalsprache)
Implant failure has been associated with the distribution of occlusal load and improper biomechanics of the dental implant. Although subcrestal implant placement has been shown to provide several benefits, there is still some disagreement when equicrestal implant placement is considered. Another factor were types of abutments: a prefabricated stock abutment with a standard cylindrical form, and a custom-made CAD/CAM abutment designed with individualized emergence profile and angulation to better adapt to the prosthetic requirementsThis study aims to compare and analyze the stress distribution across implants components. Additionally, this study evaluates peri-implant bone remodeling in relation to implant placement depth and abutment design.
Methods
Four models of different implant abutments and different implant placement depths (SAE: Commercialized stock abutment with equicrestal implant placement, SAS: Commercialized stock abutment with subcrestal implant placement, CAE: Customized titanium abutment with equicrestal implant placement and CAS: Customized titanium abutment with subcrestal implant placement) were analyzed using Finite element analysis (FEA). A 200 N force was loaded as a mechanical factor. The stress, strain, and strain energy density (SED) were calculated in 12 months of bone remodeling.
Results
CAS and CAE had the greatest increase in cortical and cancellous bone density, respectively. CAS showed the greatest maximum von Mises stress at the cancellous bone, implant fixture, and screw, while SAS demonstrated the greatest maximum von Mises stress at abutment and crown. The maximum von Mises stress in the cortical bone were similar in all models. However, in the cancellous bone, SAS and CAS showed the values slightly exceeded the natural bony structure boundary.
Conclusions
The implant placement depth had a more significant impact than abutment design. CAS exhibited the most rapid remodeling period in the first few months.
Förderung
- Chiang Mai University
Zitieren
APA
Thongsamak, P., Rungsiyakull, C., Rungsiyakull, P., Somngam, C., & Khongkhunthian, P. (2025). Bone remodeling of dental implant comparing between the customized titanium abutment and commercialized stock abutment and positioning of the implant placement depth: A 3D finite element analysis. Digital Dentistry Journal. https://doi.org/10.1016/j.ddj.2025.100018
Vancouver
Thongsamak P, Rungsiyakull C, Rungsiyakull P, Somngam C, Khongkhunthian P. Bone remodeling of dental implant comparing between the customized titanium abutment and commercialized stock abutment and positioning of the implant placement depth: A 3D finite element analysis. Digital Dentistry Journal. 2025. doi:10.1016/j.ddj.2025.100018
BibTeX
@article{Thongsamak2025,
author = {Thongsamak, Poosit and Rungsiyakull, Chaiy and Rungsiyakull, Pimduen and Somngam, Chutikarn and Khongkhunthian, Pathawee},
title = {Bone remodeling of dental implant comparing between the customized titanium abutment and commercialized stock abutment and positioning of the implant placement depth: A 3D finite element analysis},
journal = {Digital Dentistry Journal},
year = {2025},
doi = {10.1016/j.ddj.2025.100018}
}
EUVIMED