Quantitation of glyphosate, glufosinate, and AMPA in drinking water and surface waters using direct injection and charged-surface ultra-high performance liquid chromatography-tandem mass spectrometry
Chemosphere · 2024-02 · 349 · 140924
DOI: 10.1016/j.chemosphere.2023.140924 ↗Kurz gefasst
Fachartikel in der Zeitschrift Chemosphere, 2024-02. Verfasst von Alistair K Brown und 1 weiteren Autor:innen. Beteiligte Einrichtung: University of Manitoba, Department of Soil Science, Winnipeg, MB, R3T 2N2, Canada. Electronic address: Alistair.brown@umanitoba.ca. Hauptthemen laut MEDLINE: Drinking Water, Herbicides. Die Arbeit wurde bisher 26-mal zitiert (Crossref) und nennt 61 Literaturangaben. Gefördert von: Natural Sciences and Engineering Research Council of Canada, Agilent Technologies.
Abstract (Originalsprache)
Herbicides glyphosate (N-(phosphonomethyl)glycine) and glufosinate (2-amino-4-(hydroxymethylphosphinyl)butanoic acid) and the main transformation product of glyphosate, aminomethanephosphonic acid (AMPA), are challenging to analyze for in environmental samples. The quantitative method developed by this study adapts previously standardized dechlorination procedures coupled to a novel charged surface C18 column, ultra-high performance liquid chromatography-tandem mass spectrometry, polarity switching, and direct injection. The method was applied to chlorinated tap water, as well as river samples, collected in the City of Winnipeg and rural Manitoba, Canada. Using only syringe filtration without derivatization, the validated method resulted in good accuracies in both tap and surface water, at both 2 and 20 μg L-1. Method limits of detection (MLD) and quantification (MLQ) ranged from 0.022/0.074 to 0.11/0.36 μg L-1, with precisions of 0.46-2.2% (intraday) and 1.3-7.3% (interday). The mean (SEM) of the pesticides in μg L-1 for tap water were 0.11 (0.007) (AMPA), glufosinate and glyphosate MLD but < MLQ once and that was for Shannon Creek at 0.2 μg L-1. For the remaining rivers, the mean concentrations ranged from 0.31 to 3.1 μg L-1 for AMPA, and 0.087-0.53 μg L-1 for glyphosate. The method will be ideal for supporting monitoring and risk assessment programs that require high throughput sampling and quantitative methods capable of producing robust results that leverages chromatographic and mass spectrometric paradigms instead of being extraction technology focused.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- University of Manitoba, Department of Soil Science, Winnipeg, MB, R3T 2N2, Canada. Electronic address: Alistair.brown@umanitoba.ca.
- University of Manitoba, Department of Soil Science, Winnipeg, MB, R3T 2N2, Canada.
Förderung
- Natural Sciences and Engineering Research Council of Canada
- Agilent Technologies
Zitieren
APA
Brown, A. K., & Farenhorst, A. (2024). Quantitation of glyphosate, glufosinate, and AMPA in drinking water and surface waters using direct injection and charged-surface ultra-high performance liquid chromatography-tandem mass spectrometry. Chemosphere, 349, 140924. https://doi.org/10.1016/j.chemosphere.2023.140924
Vancouver
Brown AK, Farenhorst A. Quantitation of glyphosate, glufosinate, and AMPA in drinking water and surface waters using direct injection and charged-surface ultra-high performance liquid chromatography-tandem mass spectrometry. Chemosphere. 2024;349:140924. doi:10.1016/j.chemosphere.2023.140924
BibTeX
@article{Brown2024,
author = {Brown, Alistair K and Farenhorst, Annemieke},
title = {Quantitation of glyphosate, glufosinate, and AMPA in drinking water and surface waters using direct injection and charged-surface ultra-high performance liquid chromatography-tandem mass spectrometry},
journal = {Chemosphere},
year = {2024},
volume = {349},
pages = {140924},
doi = {10.1016/j.chemosphere.2023.140924}
}
EUVIMED