EUVIMEDEUVIMED entdecken
Multizentrische StudieOpen Access · CC BY 4.0

Performance of the digital cell morphology analyzer MC-100i in a multicenter study in tertiary hospitals in China

Hong Jiang, Wei Xu, Wei Chen, Jun He, Haoqin Jiang, Zhigang Mao, Min Liu, Mianyang Li, Dandan Liu, Yuling Pan, Chenxue Qu, Linlin Qu und 10 weitere

Clinica Chimica Acta · 2024

DOI: 10.1016/j.cca.2024.117801 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Clinica Chimica Acta, 2024. Verfasst von Hong Jiang und 21 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, West China Hospital of Sichuan University, Chengdu 610044, China. Hauptthemen laut MEDLINE: Leukocytes, Hematology. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: This study investigated the performance of the MC-100i, a pre-commercial digital morphology analyzer utilizing a convolutional neural network algorithm, in a multicentric setting involving up to 11 tertiary hospitals in China. METHODS: Blood smears were analyzed by MC-100i, verified by morphologists, and manually differentiated. The classification performance on WBCs and RBCs was evaluated by comparing the classification results using different methods. The PLT and PLT clump counting performance was also assessed. The total assay time including hands-on time was evaluated. RESULTS: The agreements between pre- and post-classification were high for normal WBCs (κ > 0.96) and lower for overall abnormal WBCs (κ = 0.90). The post-classification results correlated well with manual differentials for both normal and abnormal WBCs (r > 0.93), except for basophils (r = 0.8480) and atypical lymphocytes (r = 0.8211). The clinical sensitivity and specificity of each RBC abnormality after verification were above 90 % using microscopy reviews as the reference. The PLTs counted by the MC-100i before and after verification correlated well with those measured by the PLT-O mode (r = 0.98). Moreover, PLT clumps were successfully classified by the analyzer in EDTA-dependent pseudothrombocytopenia blood samples. CONCLUSIONS: The MC-100i is an accurate and reliable digital cell morphology analyzer, offering another intelligent option for hematology laboratories.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LeukocytesHematologyHumansTertiary Care CentersErythrocytesChinaReproducibility of Results

Stichwörter der Autor:innen

Digital morphologyMindray MC-100iMulticenterPerformanceWhite blood cell differentials

Autor:innen und Institutionen

  • Department of Laboratory Medicine, West China Hospital of Sichuan University, Chengdu 610044, China.
  • Department of Laboratory Medicine, The First Bethune Hospital of Jilin University, Jilin 130061, China.
  • Department of Laboratory Medicine, The First Affiliated Hospital of Xi'an Jiaotong University, Xi'an 710061, China.
  • Department of Laboratory Medicine, The First Affiliated Hospital of Soochow University, Suzhou 215006, China.
  • Department of Laboratory Medicine, Huashan Hospital Fudan University, Shanghai 200040, China.
  • Department of Laboratory Medicine, The First Affiliated Hospital, Sun Yat-sen University, Guangzhou 510062, China.

Zitieren

APA

Jiang, H., Xu, W., Chen, W., He, J., Jiang, H., Mao, Z., Liu, M., Li, M., Liu, D., Pan, Y., Qu, C., Qu, L., Sun, Z., Sun, D., Wang, X., Wang, J., Wu, W., Xing, Y., Zhang, S., …, & Guan, M. (2024). Performance of the digital cell morphology analyzer MC-100i in a multicenter study in tertiary hospitals in China. Clinica Chimica Acta. https://doi.org/10.1016/j.cca.2024.117801

Vancouver

Jiang H, Xu W, Chen W, He J, Jiang H, Mao Z, et al. Performance of the digital cell morphology analyzer MC-100i in a multicenter study in tertiary hospitals in China. Clinica Chimica Acta. 2024. doi:10.1016/j.cca.2024.117801

BibTeX
@article{Jiang2024,
  author = {Jiang, Hong and Xu, Wei and Chen, Wei and He, Jun and Jiang, Haoqin and Mao, Zhigang and Liu, Min and Li, Mianyang and Liu, Dandan and Pan, Yuling and Qu, Chenxue and Qu, Linlin and Sun, Ziyong and Sun, Dehua and Wang, Xuefeng and Wang, Jianbiao and Wu, Wenjing and Xing, Ying and Zhang, Shihong and Zhang, Chi and Zheng, Lei and Guan, Ming},
  title = {Performance of the digital cell morphology analyzer MC-100i in a multicenter study in tertiary hospitals in China},
  journal = {Clinica Chimica Acta},
  year = {2024},
  doi = {10.1016/j.cca.2024.117801}
}
Kostenlos starten →