Multicenter study on mechanical accuracy of short dwell time and source transit for HDR Ir-192 brachytherapy treatment machine
Brachytherapy · 2026
DOI: 10.1016/j.brachy.2025.09.018 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Brachytherapy, 2026. Verfasst von Yoshihiro Ueda und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Osaka International Cancer Institute, Chuo-ku, Osaka 537-8567, Japan. Hauptthemen laut MEDLINE: Brachytherapy, Iridium Radioisotopes. Die Arbeit nennt 22 Literaturangaben. Gefördert von: Showa University, Health Sciences, University of Arizona, Tokyo Medical and Dental University.
Abstract (Originalsprache)
PURPOSE: To evaluate the accuracy of the effective transit and dwell times for high-dose-rate (HDR) brachytherapy machines across multiple institutions. The aim was to establish reference levels for quality assurance (QA) in MicroSelectron and Flexitron afterloaders. METHODS: The effective transit time of the amount of charge from the moving source divided by the current value and the short dwell times were measured using institution-owned well-type ionization chambers for 11 treatment machines from 10 institutions, including MicroSelectron and Flexitron afterloaders. Measurements were performed under standardized conditions to minimize variability, including controlled distances from walls and floors to reduce scattered radiation. The proportionality constants and effective transit times were calculated and the dwell times were measured for settings ranging from 0.1 to 30 s. For each afterloader, effective transit time and dwell time accuracy were measured five times, and the averages were used for analysis. RESULTS: MicroSelectron afterloaders exhibited mean effective transit times of 0.90-1.05 s, whereas Flexitron afterloaders showed shorter times of 0.54-0.62 s. For the dwell time accuracy, MicroSelectron afterloaders exhibited a mean error of -0.048 to 0.104 s, with larger variability across institutions. In contrast, Flexitron afterloaders showed a mean error consistently below ±0.01 s, even at shorter dwell times. The variability in dwell time was notably significant higher for MicroSelectron than for Flexitron afterloaders (p ≤ 0.002). CONCLUSIONS: This study demonstrates clear performance differences between MicroSelectron and Flexitron afterloaders in HDR brachytherapy and provides reference levels that support device-specific QA protocols to improve treatment precision and ensure consistent outcomes across institutions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiation Oncology, Osaka International Cancer Institute, Chuo-ku, Osaka 537-8567, Japan.
- Radiation Safety and Quality Assurance Division, National Cancer Center Hospital, Chuo-ku, Tokyo 104-0045, Japan. Electronic address: hiokamot@ncc.go.jp.
- Division of Radiology, Department of Clinical Technology, Kagoshima University Hospital, Kagoshima-City, Kagoshima 890-8520, Japan.
- Division of Radiology, Department of Clinical Technology, Tottori University Hospital, Yonago-City, Tottori 683-8504, Japan.
- Department of Radiation Oncology, Faculty of Medicine, Juntendo University, Bunkyo-ku, Tokyo 113-0033, Japan.
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Higashi- ku, Fukuoka, 812-8582, Japan.
Förderung
- Showa University
- Health Sciences, University of Arizona
- Tokyo Medical and Dental University
Zitieren
APA
Ueda, Y., Okamoto, H., Oku, Y., Ono, Y., Takatsu, J., Fukunaga, J. I., Ushijima, T., Toyota, M., Iijima, K., Murakami, N., & Ohno, T. (2026). Multicenter study on mechanical accuracy of short dwell time and source transit for HDR Ir-192 brachytherapy treatment machine. Brachytherapy. https://doi.org/10.1016/j.brachy.2025.09.018
Vancouver
Ueda Y, Okamoto H, Oku Y, Ono Y, Takatsu J, Fukunaga JI, et al. Multicenter study on mechanical accuracy of short dwell time and source transit for HDR Ir-192 brachytherapy treatment machine. Brachytherapy. 2026. doi:10.1016/j.brachy.2025.09.018
BibTeX
@article{Ueda2026,
author = {Ueda, Yoshihiro and Okamoto, Hiroyuki and Oku, Yoshifumi and Ono, Yasushi and Takatsu, Jun and Fukunaga, Jun-Ichi and Ushijima, Takahiro and Toyota, Masahiko and Iijima, Kotaro and Murakami, Naoya and Ohno, Tatsuya},
title = {Multicenter study on mechanical accuracy of short dwell time and source transit for HDR Ir-192 brachytherapy treatment machine},
journal = {Brachytherapy},
year = {2026},
doi = {10.1016/j.brachy.2025.09.018}
}
EUVIMED