EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Microfluidic contact lens for continuous monitoring of ocular oxidative stress

Yuqi Shi, Yubing Hu, Yihan Zhang, Angie Davina Tjandra, Seojung Park, Rona Chandrawati, Savas Tasoglu, Nan Jiang, Ali K Yetisen

Biosensors and Bioelectronics · 2025

DOI: 10.1016/j.bios.2025.117427 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2025. Verfasst von Yuqi Shi und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Chemical Engineering, Imperial College London, South Kensington, London, SW7 2BU, UK. Hauptthemen laut MEDLINE: Biosensing Techniques, Oxidative Stress, Glutathione, Contact Lenses. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 57 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Fundamental Research Funds for the Central Universities, Engineering and Physical Sciences Research Council.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Oxidative stress-induced ocular dysfunctions are a significant global health concern. Glutathione (GSH), an abundant antioxidant in tears, can serve as an index for oxidative stress (OS). A coumarin-based fluorescent probe named CCAE was synthesized for GSH monitoring, functioning through structural changes via Michael's addition with GSH and elimination by H2O2, which restores the conjugation structure to assess the severity of OS-induced ocular diseases. CCAE demonstrated a high sensitivity with a detection limit (LOD) of 0.12 mM and reversibility for 15 cycles. A wearable contact lens sensor was developed featuring a microfluidic lens patterned with a 365 nm pulsed laser, requiring 6 μL of tears. CCAE was encapsulated in citric acid-crosslinked poly(vinyl alcohol) film and embedded in poly((hydroxyethyl)methacrylate-co-ethylene glycol)/polyvinylpyrrolidone (poly(HEMA-co-EG)/PVP) lenses. A customized smartphone readout device enabled quantitative GSH readings for point-of-care applications. Tested on an ex vivo porcine anterior eye model, the sensor achieved an LOD of 0.204 mM, within a detection range of 0.62-1.17 mM and 0.13-0.73 mM under mild and severe OS conditions respectively. The sensors maintained operational stability for 2 days and storage stability for 1 week. This reversible GSH contact lens sensor offers a unique platform for diagnosing and monitoring OS-related ocular conditions at point-of-care settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Biosensing TechniquesOxidative StressGlutathioneContact LensesAnimalsSwineTearsHumansLimit of DetectionFluorescent DyesCoumarinsEquipment DesignWearable Electronic DevicesHydrogen Peroxide

Stichwörter der Autor:innen

Dry eye diseaseMedical diagnosticsOcular sensorsOxidative stress conditionsPoint-of-care diagnosisSmartphone readout device

Substanzen

GlutathioneFluorescent DyesCoumarinsHydrogen Peroxide

Autor:innen und Institutionen

  • Department of Chemical Engineering, Imperial College London, South Kensington, London, SW7 2BU, UK.
  • School of Chemical Engineering, The University of New South Wales (UNSW Sydney), Sydney, NSW, 2052, Australia; Australian Centre for Nanomedicine (ACN), The University of New South Wales (UNSW Sydney), Sydney, NSW, 2052…
  • Department of Mechanical Engineering, Koc University, Sarıyer, Istanbul, 34450, Türkiye.
  • West China School of Basic Medical Sciences & Forensic Medicine, Sichuan University, Chengdu, 610041, China; Jinfeng Laboratory, Chongqing, 401329, China.
  • Department of Chemical Engineering, Imperial College London, South Kensington, London, SW7 2BU, UK. Electronic address: a.yetisen@imperial.ac.uk.

Förderung

  • Fundamental Research Funds for the Central Universities
  • Engineering and Physical Sciences Research Council

Zitieren

APA

Shi, Y., Hu, Y., Zhang, Y., Tjandra, A. D., Park, S., Chandrawati, R., Tasoglu, S., Jiang, N., & Yetisen, A. K. (2025). Microfluidic contact lens for continuous monitoring of ocular oxidative stress. Biosensors and Bioelectronics. https://doi.org/10.1016/j.bios.2025.117427

Vancouver

Shi Y, Hu Y, Zhang Y, Tjandra AD, Park S, Chandrawati R, et al. Microfluidic contact lens for continuous monitoring of ocular oxidative stress. Biosensors and Bioelectronics. 2025. doi:10.1016/j.bios.2025.117427

BibTeX
@article{Shi2025,
  author = {Shi, Yuqi and Hu, Yubing and Zhang, Yihan and Tjandra, Angie Davina and Park, Seojung and Chandrawati, Rona and Tasoglu, Savas and Jiang, Nan and Yetisen, Ali K},
  title = {Microfluidic contact lens for continuous monitoring of ocular oxidative stress},
  journal = {Biosensors and Bioelectronics},
  year = {2025},
  doi = {10.1016/j.bios.2025.117427}
}
Kostenlos starten →