Non-invasive, ultrasensitive detection of glucose in saliva using metal oxide transistors
Biosensors and Bioelectronics · 2023
DOI: 10.1016/j.bios.2023.115448 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2023. Verfasst von Abhinav Sharma und 10 weiteren Autor:innen. Beteiligte Einrichtung: King Abdullah University of Science and Technology (KAUST), KAUST Solar Center (KSC), Thuwal 23955-6900, Saudi Arabia. Electronic address: abhinav.sh…. Hauptthemen laut MEDLINE: Biosensing Techniques, Zinc Oxide. Die Arbeit wurde bisher 48-mal zitiert (Crossref) und nennt 56 Literaturangaben.
Abstract (Originalsprache)
Transistor-based biosensors represent an emerging technology for inexpensive point-of-care testing (POCT) applications. However, the limited sensitivity of the current transistor technologies hinders their practical deployment. In this study, we developed tri-channel In2O3/ZnO heterojunction thin-film transistors (TFTs) featuring the surface-immobilized enzyme glucose oxidase to detect glucose in various biofluids. This unusual channel design facilitates strong coupling between the electrons transported along the buried In2O3/ZnO heterointerface and the electrostatic perturbations caused by the interactions between glucose and surface-immobilized glucose oxidase. The enzyme selectively binds to glucose, causing a change in charge density on the channel surface. By exploring this effect, the solid-state biosensing TFT (BioTFT) can selectively detect glucose in artificial and real saliva over a wide range of concentrations from 500 nM to 20 mM with limits of detection of ∼365 pM (artificial saliva) and ∼416 nM (real saliva) in less than 60 s. The specificity of the sensor towards glucose has been demonstrated against various interfering species in artificial saliva, further highlighting its unique capabilities. Moreover, the BioTFTs exhibited good operating stability upon storage for up to two weeks, with relative standard deviation (RSD) values ranging from 2.36% to 6.39% for 500 nM glucose concentration. Our BioTFTs are easy to manufacture with reliable operation, making them ideal for non-invasive POCT applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- King Abdullah University of Science and Technology (KAUST), KAUST Solar Center (KSC), Thuwal 23955-6900, Saudi Arabia. Electronic address: abhinav.sharma@kaust.edu.sa.
- King Abdullah University of Science and Technology (KAUST), KAUST Solar Center (KSC), Thuwal 23955-6900, Saudi Arabia.
- Department of Electronic and Computer Engineering, The Hong Kong University of Science and Technology, Clear Water Bay, Kowloon, Hong Kong.
- Department of Mechanical Engineering, National Taiwan University, Taipei 10617, Taiwan.
- King Abdullah University of Science and Technology (KAUST), KAUST Solar Center (KSC), Thuwal 23955-6900, Saudi Arabia. Electronic address: thomas.anthopoulos@kaust.edu.sa.
Zitieren
APA
Sharma, A., AlGhamdi, W. S., Faber, H., Lin, Y. H., Liu, C. H., Hsu, E. K., Lin, W. Z., Naphade, D., Mandal, S., Heeney, M., & Anthopoulos, T. D. (2023). Non-invasive, ultrasensitive detection of glucose in saliva using metal oxide transistors. Biosensors and Bioelectronics. https://doi.org/10.1016/j.bios.2023.115448
Vancouver
Sharma A, AlGhamdi WS, Faber H, Lin YH, Liu CH, Hsu EK, et al. Non-invasive, ultrasensitive detection of glucose in saliva using metal oxide transistors. Biosensors and Bioelectronics. 2023. doi:10.1016/j.bios.2023.115448
BibTeX
@article{Sharma2023,
author = {Sharma, Abhinav and AlGhamdi, Wejdan S and Faber, Hendrik and Lin, Yen-Hung and Liu, Chien-Hao and Hsu, En-Kai and Lin, Wei-Zhi and Naphade, Dipti and Mandal, Suman and Heeney, Martin and Anthopoulos, Thomas D},
title = {Non-invasive, ultrasensitive detection of glucose in saliva using metal oxide transistors},
journal = {Biosensors and Bioelectronics},
year = {2023},
doi = {10.1016/j.bios.2023.115448}
}
EUVIMED