Evaluating CT Dose Variation Across Scanner Technologies: Implications for Compliance with New CMS CT Radiation Dose Measure
Academic Radiology · 2025
DOI: 10.1016/j.acra.2025.03.026 ↗Kurz gefasst
Fachartikel in der Zeitschrift Academic Radiology, 2025. Verfasst von Madan M Rehani und 2 weiteren Autor:innen. Beteiligte Einrichtung: Massachusetts General Hospital, 55 Fruit Str. Boston, Massachusetts. Electronic address: mrehani@mgh.harvard.edu. Hauptthemen laut MEDLINE: Radiation Dosage, Tomography, X-Ray Computed, Tomography Scanners, X-Ray Computed, Guideline Adherence. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 21 Literaturangaben.
Abstract (Originalsprache)
RATIONALE AND OBJECTIVES: In 2025, the Centers for Medicare and Medicaid Services introduced a computed tomography (CT) dose measure for pay-for-performance programs. Hospitals employ diverse scanner fleets, but the impact of scanner technologies on dose benchmarking remains unclear. This study evaluates dose variation across scanner models and its benchmarking implications. MATERIALS AND METHODS: A retrospective analysis examined CT exams from January to December 2023 at a quaternary-care hospital, focusing on median-sized adults (water-equivalent diameter: 16-19cm head, 18-22cm neck, 29-33cm torso). Dose indices from seven scanner models across eight adult exams were evaluated. The 50th and 75th percentile doses were calculated per exam and scanner model combination, compared to American College of Radiology achievable doses and diagnostic reference levels. RESULTS: Analyzing 34,166 studies, significant dose variations with scanner models emerged. Head without contrast (N=21,654) had median volume CT-dose-index (CTDIvol) of 36.1-68.3mGy and dose-length-product (DLP) 704-1307.8mGy·cm; 75th percentiles were 43.1-69.1mGy and 838.2-1378.1mGy·cm. Chest with contrast (N=3065) showed median CTDIvol of 6.7-16.1mGy and DLP 263.8-579.7mGy·cm; 75th percentiles were 8.2-19.5mGy and 329-713.7mGy·cm. Abdomen/pelvis with contrast (N=2740) had median CTDIvol of 8.8-15.2mGy and DLP 429.3-782.1mGy·cm; 75th percentiles were 10-18.5mGy and 533.4-941.5mGy·cm. While the number of studies was smaller, five other exams also showed large dose variations across scanner models. CONCLUSION: Single-value dose benchmarks ignoring scanner technology may be inadequate, even for similar-sized patients, potentially requiring scanner removal. Incorporating benchmarks with diverse technologies could prevent increased healthcare costs and patient care disruptions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Massachusetts General Hospital, 55 Fruit Str. Boston, Massachusetts. Electronic address: mrehani@mgh.harvard.edu.
- Massachusetts General Hospital, 55 Fruit Str. Boston, Massachusetts.
Zitieren
APA
Rehani, M. M., Mataac, M. T., & Li, X. (2025). Evaluating CT Dose Variation Across Scanner Technologies: Implications for Compliance with New CMS CT Radiation Dose Measure. Academic Radiology. https://doi.org/10.1016/j.acra.2025.03.026
Vancouver
Rehani MM, Mataac MT, Li X. Evaluating CT Dose Variation Across Scanner Technologies: Implications for Compliance with New CMS CT Radiation Dose Measure. Academic Radiology. 2025. doi:10.1016/j.acra.2025.03.026
BibTeX
@article{Rehani2025,
author = {Rehani, Madan M and Mataac, Maria T and Li, Xinhua},
title = {Evaluating CT Dose Variation Across Scanner Technologies: Implications for Compliance with New CMS CT Radiation Dose Measure},
journal = {Academic Radiology},
year = {2025},
doi = {10.1016/j.acra.2025.03.026}
}
EUVIMED