Dose-Dependent Analysis of Image Quality in Pediatric Head CT Scans Across Different Scanners to Optimize Clinical Protocols Using Phantom-Based Assessment
Tomography · 2025
DOI: 10.3390/tomography11110119 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Tomography, 2025. Verfasst von Hiroshi Kuwahara und 7 weiteren Autor:innen. Beteiligte Einrichtung: Doctoral Course of Health Science, Graduate School of Nursing, Oita University of Nursing and Health Sciences, 2944-9 Megusuno, Oita City 870-1201, O…. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Head, Tomography Scanners, X-Ray Computed. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: This research was funded by the Japan Society for the Promotion of Science and internal research funding from Oita University of Nursing and Health Sciences., Graduate School of Nursing Science, Oita University of Nursing and Health Sciences.
Abstract (Originalsprache)
Background/Objectives: Optimization of pediatric head computed tomography (CT) protocols is essential to minimize radiation exposure while maintaining diagnostic image quality. Previous studies mainly relied on phantom-based measurements or visual assessments, and validation using clinical images remains limited. This study aimed to establish quantitative thresholds for noise and contrast-to-noise ratio (CNR) in pediatric head CT by integrating multicenter clinical data with phantom evaluations.
Methods
A multicenter retrospective study was conducted using CT systems from eight hospitals, combined with Catphan phantom experiments and pediatric head CT data. Scan parameters, automatic exposure control settings, and reconstruction methods were collected. Image quality was quantified by the standard deviation (SD) of noise and CNR obtained from regions of interest in gray and white matter. Radiation dose was represented by CTDIvol. Relationships among CTDIvol, SD, and CNR were analyzed across scanners from three manufacturers (Canon, FUJI, and GE).
Results
Consistent dose–response trends were observed across institutions and manufacturers. Image noise decreased as CTDIvol increased, but reached a plateau at higher doses. CNR improved with dose escalation, then stabilized. Both phantom experiments and clinical analyses identified a target SD of 5 and CNR of 2 as optimal indicators for pediatric head CT.
Conclusions
Quantitative thresholds were determined as practical indicators for balancing diagnostic image quality with dose reduction. Further reduction may be achieved through advanced reconstruction methods, such as deep learning-based algorithms. These findings may contribute to standardizing pediatric head CT protocols and supporting safer and more effective diagnostic imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Doctoral Course of Health Science, Graduate School of Nursing, Oita University of Nursing and Health Sciences, 2944-9 Megusuno, Oita City 870-1201, Oita Prefecture, Japan.
- Department of Environmental Health Science, Oita University of Nursing and Health Sciences, 2944-9 Megusuno, Oita City 870-1201, Oita Prefecture, Japan.
- Radiology Department, Nakatsu Municipal Hospital, 173-1 Shimoikenaga, Nakatsu City 871-8511, Oita Prefecture, Japan.
- Radiology Department, Seichokai Social Medical Corporation Belland General Hospital, 500-3 Higashiyama, Naka Ward, Sakai City 599-8247, Osaka Prefecture, Japan.
- Department of Radiology, Oita Children's Hospital, Fujimoto Ikuseikai Social Medical Corporation, 83-7 Katajima, Oita City 870-0943, Oita Prefecture, Japan.
- Department of Radiology, Medical Technology Division, Nagoya Daini Red Cross Hospital, Japanese Red Cross Aichi Medical Center, 2-9 Myoken-cho, Showa-ku, Nagoya City 466-8650, Aichi Prefecture, Japan.
ORCID-Profile: Hiroshi Kuwahara, Mitsuaki Ojima
Förderung
- This research was funded by the Japan Society for the Promotion of Science and internal research funding from Oita University of Nursing and Health Sciences.
- Graduate School of Nursing Science, Oita University of Nursing and Health Sciences
Zitieren
APA
Kuwahara, H., Ojima, M., Kawamura, T., Saitou, D., Andou, K., Ariga, E., Hasegawa, K., & Kai, M. (2025). Dose-Dependent Analysis of Image Quality in Pediatric Head CT Scans Across Different Scanners to Optimize Clinical Protocols Using Phantom-Based Assessment. Tomography. https://doi.org/10.3390/tomography11110119
Vancouver
Kuwahara H, Ojima M, Kawamura T, Saitou D, Andou K, Ariga E, et al. Dose-Dependent Analysis of Image Quality in Pediatric Head CT Scans Across Different Scanners to Optimize Clinical Protocols Using Phantom-Based Assessment. Tomography. 2025. doi:10.3390/tomography11110119
BibTeX
@article{Kuwahara2025,
author = {Kuwahara, Hiroshi and Ojima, Mitsuaki and Kawamura, Tsuneko and Saitou, Daisuke and Andou, Kazunari and Ariga, Eiji and Hasegawa, Kotaro and Kai, Michiaki},
title = {Dose-Dependent Analysis of Image Quality in Pediatric Head CT Scans Across Different Scanners to Optimize Clinical Protocols Using Phantom-Based Assessment},
journal = {Tomography},
year = {2025},
doi = {10.3390/tomography11110119}
}
EUVIMED