How Real Are Computed Tomography Low Dose Simulations? An Investigational In-Vivo Large Animal Study
Academic Radiology · 2023
DOI: 10.1016/j.acra.2022.11.008 ↗Kurz gefasst
Fachartikel in der Zeitschrift Academic Radiology, 2023. Verfasst von Andreas S Brendlin und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, Eberhard-Karls University, D-72076 - Tuebingen, Germany. Electronic address: andreas.brendlin@…. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 29 Literaturangaben.
Abstract (Originalsprache)
OBJECTIVES: CT low-dose simulation methods have gained significant traction in protocol development, as they lack the risk of increased patient exposure. However, in-vivo validations of low-dose simulations are as uncommon as prospective low-dose image acquisition itself. Therefore, we investigated the extent to which simulated low-dose CT datasets resemble their real-dose counterparts. MATERIALS AND METHODS: Fourteen veterinarian-sedated alive pigs underwent three CT scans on the same third generation dual-source scanner with 2 months between each scan. At each time, three additional scans ensued, with mAs reduced to 50%, 25%, and 10%. All scans were reconstructed using wFBP and ADMIRE levels 1-5. Matching low-dose datasets were generated from the 100% scans using reconstruction-based and DICOM-based simulations. Objective image quality (CT numbers stability, noise, and signal-to-noise ratio) was measured via consistent regions of interest. Three radiologists independently rated all possible dataset combinations per time point for subjective image quality (-1=inferior, 0=equal, 1=superior). The points were averaged for a semiquantitative score, and inter-rater-agreement was measured using Spearman's correlation coefficient. A structural similarity index (SSIM) analyzed the voxel-wise similarity of the volumes. Adequately corrected mixed-effects analysis compared objective and subjective image quality. Multiple linear regression with three-way interactions measured the contribution of dose, reconstruction mode, simulation method, and rater to subjective image quality. RESULTS: There were no significant differences between objective and subjective image quality of reconstruction-based and DICOM-based simulation on all dose levels (p≥0.137). However, both simulation methods produced significantly lower objective image quality than real-dose images below 25% mAs due to noise overestimation (p
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Diagnostic and Interventional Radiology, Eberhard-Karls University, D-72076 - Tuebingen, Germany. Electronic address: andreas.brendlin@med.uni-tuebingen.de.
- Department of Diagnostic and Interventional Radiology, Eberhard-Karls University, D-72076 - Tuebingen, Germany.
- Institute for Nuclear Medicine and Clinical Molecular Imaging, Eberhard-Karls University, Tuebingen, Germany.
- Department of General, Visceral and Transplant Surgery, Eberhard-Karls University, Tuebingen, Germany.
- Department of Diagnostic and Interventional Radiology, Eberhard-Karls University, D-72076 - Tuebingen, Germany; Department of Neuroradiology, University Medical Center, Mainz, Germany.
Zitieren
APA
Brendlin, A. S., Wrazidlo, R., Almansour, H., Estler, A., Plajer, D., Vega, S. G. C., Klingert, W., Bertolani, E., Othman, A. E., Schenk, M., & Afat, S. (2023). How Real Are Computed Tomography Low Dose Simulations? An Investigational In-Vivo Large Animal Study. Academic Radiology. https://doi.org/10.1016/j.acra.2022.11.008
Vancouver
Brendlin AS, Wrazidlo R, Almansour H, Estler A, Plajer D, Vega SGC, et al. How Real Are Computed Tomography Low Dose Simulations? An Investigational In-Vivo Large Animal Study. Academic Radiology. 2023. doi:10.1016/j.acra.2022.11.008
BibTeX
@article{Brendlin2023,
author = {Brendlin, Andreas S and Wrazidlo, Robin and Almansour, Haidara and Estler, Arne and Plajer, David and Vega, Salvador Guillermo Castaneda and Klingert, Wilfried and Bertolani, Elisa and Othman, Ahmed E and Schenk, Martin and Afat, Saif},
title = {How Real Are Computed Tomography Low Dose Simulations? An Investigational In-Vivo Large Animal Study},
journal = {Academic Radiology},
year = {2023},
doi = {10.1016/j.acra.2022.11.008}
}
EUVIMED