Investigating the role of imaging factors in the variability of CT‐based texture analysis metrics
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14192 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Bino Abel Varghese und 8 weiteren Autor:innen. Beteiligte Einrichtung: Keck Medical Center, Department of Radiology, University of Southern California, Los Angeles, California, USA. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Algorithms. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.
Abstract (Originalsprache)
Objective
This study assesses the robustness of first‐order radiomic texture features namely interquartile range (IQR), coefficient of variation (CV) and standard deviation (SD) derived from computed tomography (CT) images by varying dose, reconstruction algorithms and slice thickness using scans of a uniform water phantom, a commercial anthropomorphic liver phantom, and a human liver in‐vivo. Materials and Methods Scans were acquired on a 16 cm detector GE Revolution Apex Edition CT scanner with variations across three different nominal slice thicknesses: 0.625, 1.25, and 2.5 mm, three different dose levels: CTDIvol of 13.86 mGy for the standard dose, 40% reduced dose and 60% reduced dose and two different reconstruction algorithms: a deep learning image reconstruction (DLIR‐high) algorithm and a hybrid iterative reconstruction (IR) algorithm ASiR‐V50% (AV50) were explored, varying one at a time. To assess the effect of non‐linear modifications of images by AV50 and DLIR‐high, images of the water phantom were also reconstructed using filtered back projection (FBP). Quantitative measures of IQR, CV and SD were extracted from twelve pre‐selected, circular (1 cm diameter) regions of interest (ROIs) capturing different texture patterns across all scans.
Results
Across all scans, imaging, and reconstruction settings, CV, IQR and SD were observed to increase with reduction in dose and slice thickness. An exception to this observation was found when using FBP reconstruction. Lower values of CV, IQR and SD were observed in DLIR‐high reconstructions compared to AV50 and FBP. The Poisson statistics were more stringently noted in FBP than DLIR‐high and AV50, due to the non‐linear nature of the latter two algorithms.
Conclusion
Variation in image noise due to dose reduction algorithms, tube current, and slice thickness show a consistent trend across phantom and patient scans. Prospective evaluation across multiple centers, scanners and imaging protocols is needed for establishing quality assurance standards of radiomics.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Keck Medical Center, Department of Radiology, University of Southern California, Los Angeles, California, USA.
- Department of Physics and Computational Radiology, Oslo, Norway.
- The Phantom Laboratory, Greenwich, New York, USA.
- Department of Radiology and Nuclear Medicine, Oslo University Hospital, Oslo, Norway.
- Department of Radiology, George Washington University, Washington, District of Columbia, USA.
ORCID-Profile: Bino Abel Varghese
Förderung
- NCI NIH HHS
Zitieren
APA
Varghese, B. A., Cen, S. Y., Jensen, K., Levy, J., Andersen, H. K., Schulz, A., Lei, X., Duddalwar, V. A., & Goodenough, D. J. (2024). Investigating the role of imaging factors in the variability of CT‐based texture analysis metrics. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14192
Vancouver
Varghese BA, Cen SY, Jensen K, Levy J, Andersen HK, Schulz A, et al. Investigating the role of imaging factors in the variability of CT‐based texture analysis metrics. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14192
BibTeX
@article{Varghese2024,
author = {Varghese, Bino Abel and Cen, Steven Yong and Jensen, Kristin and Levy, Joshua and Andersen, Hilde Kjernlie and Schulz, Anselm and Lei, Xiaomeng and Duddalwar, Vinay Anant and Goodenough, David John},
title = {Investigating the role of imaging factors in the variability of CT‐based texture analysis metrics},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14192}
}
EUVIMED