Abstract (Originalsprache)
Impact loads frequently cause skull fractures or other head injuries that call for a craniectomy. Despite the availability of many synthetic and natural material options, implant-related issues remain common. Repair materials should be highly biocompatible, mechanically strong, and easy to shape. However, no substance currently fully integrates with human bone tissue and meets all clinical requirements. Thus, this study explores the potential application of a relatively novel biocoated 316L stainless steel implant. We used chitosan and gelatine blends, enhanced with various stiff and soft reinforcements (Al_2O_3, ZrO_2, and Biotin) and (PCL, Clove, and PMMA), in the coating process. Hardness, wettability, FTIR, surface roughness, MTT, antibacterial activity, immersion, coating thickness, adhesion strength, and SEM were evaluated. SEM data confirmed a uniform coating with uniformly spaced reinforcement and crack-free layers. The clove-reinforced coating had the highest roughness (2.7 µm) and hardness (118 MPa). Clove-containing coatings increased surface hydrophilicity, thereby promoting cell adhesion. ZrO_2 exhibited antibacterial activity, with inhibition zones of 25 and 22 mm against Staphylococcus aureus and Pseudomonas aeruginosa, respectively. According to the MTT results, the substrate surfaces became more reactive after 72 h of interaction with MG-63 cells. Ni, Cr, and Fe ions released during the immersion tests were not contained within the Ringer's solution. However, other groups contained the released Al, Zr, and C ions. The strongest adherence was exhibited by the dispersed ceramic particles (Al_2O_3) in the matrix. The FTIR data indicated a clear attraction between the additives and the matrix, with no observable wavelength shifts. A significant change in mean cell viability across the groups was found by ANOVA ( p -value < 0.001). According to the study, the suggested implants for cranial reconstruction meet the need for implants with sufficient mechanical durability, thereby improving patients' function and quality of life.
Zitieren
APA
Faheed, N. K., Kamil, G. A., & Najm, N. (2026). Experimental Investigation of the Modification of Biomechanical and Corrosion Behavior of a Novel Skull Implant. Biomedical Materials & Devices. https://doi.org/10.1007/s44174-026-00807-w
Vancouver
Faheed NK, Kamil GA, Najm N. Experimental Investigation of the Modification of Biomechanical and Corrosion Behavior of a Novel Skull Implant. Biomedical Materials & Devices. 2026. doi:10.1007/s44174-026-00807-w
BibTeX
@article{Faheed2026,
author = {Faheed, Noor K. and Kamil, Ghadeer Ahmed and Najm, Noor},
title = {Experimental Investigation of the Modification of Biomechanical and Corrosion Behavior of a Novel Skull Implant},
journal = {Biomedical Materials & Devices},
year = {2026},
doi = {10.1007/s44174-026-00807-w}
}