Shear-Wave Elastography of the Achilles tendon: reliability analysis and impact of parameters modulating elasticity values
Journal of Ultrasound · 2024
DOI: 10.1007/s40477-024-00877-w ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Ultrasound, 2024. Verfasst von Michael-Andrei Pelea und 4 weiteren Autor:innen. Beteiligte Einrichtung: 2nd Internal Medicine Department, "Iuliu Hatieganu" University of Medicine and Pharmacy, 2-4 Clinicilor Street, 400006, Cluj-Napoca Napoca, Romania. Hauptthemen laut MEDLINE: Achilles Tendon, Elasticity Imaging Techniques. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
PURPOSE: Shear wave elastography (SWE) has seen many advancements in Achilles tendon evaluation in recent years, yet standardization of this technique is still problematic due to the lack of knowledge regarding the optimal way to perform the examination. The purpose of this study was to evaluate the effects of ankle position, probe frequency and physical effort on the shear modulus of the Achilles tendon, but also to determine the intra and inter-observer reliability of the technique. METHODS: 37 healthy volunteers were included; SWE protocol was performed by two examiners. We analyzed the shear modulus of the tendon with the ankle in neutral, maximum dorsiflexion and maximum plantar flexion using two different high frequency probes. Afterwards, the subjects performed a brief physical exercise and SWE measurements were repeated. RESULTS: The L18-5 probe showed the highest ICC values (ICC = 0.798, 95% CI 0.660-0.880, p < 0.001) when positioned at 2 cm from the calcaneal insertion with the ankle in a neutral state. Conversely, utilizing the same L18-5 probe at 1 cm from the insertion during maximum plantar flexion of the ankle resulted in the lowest ICC (ICC = 0.422, 95% CI 0.032-0.655, p = 0.019). Significant variations in elasticity values were noted among different ankle positions and probe types, while no significant changes in elasticity were observed post-physical exercise. CONCLUSION: Ankle position and probe frequency are factors that influence elasticity values of the Achilles tendon. An ankle position between 10 and 20 degrees of plantar flexion is the most suitable for SWE evaluation. However, more research focusing on Achilles tendon SWE is essential due to the challenges encountered in standardizing this region.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- 2nd Internal Medicine Department, "Iuliu Hatieganu" University of Medicine and Pharmacy, 2-4 Clinicilor Street, 400006, Cluj-Napoca Napoca, Romania.
- 2nd Internal Medicine Department, "Iuliu Hatieganu" University of Medicine and Pharmacy, 2-4 Clinicilor Street, 400006, Cluj-Napoca Napoca, Romania. dfodor@umfcluj.ro.
Zitieren
APA
Pelea, M. A., Serban, O., Badarinza, M., Gutiu, R., & Fodor, D. (2024). Shear-Wave Elastography of the Achilles tendon: reliability analysis and impact of parameters modulating elasticity values. Journal of Ultrasound. https://doi.org/10.1007/s40477-024-00877-w
Vancouver
Pelea MA, Serban O, Badarinza M, Gutiu R, Fodor D. Shear-Wave Elastography of the Achilles tendon: reliability analysis and impact of parameters modulating elasticity values. Journal of Ultrasound. 2024. doi:10.1007/s40477-024-00877-w
BibTeX
@article{Pelea2024,
author = {Pelea, Michael-Andrei and Serban, Oana and Badarinza, Maria and Gutiu, Roxana and Fodor, Daniela},
title = {Shear-Wave Elastography of the Achilles tendon: reliability analysis and impact of parameters modulating elasticity values},
journal = {Journal of Ultrasound},
year = {2024},
doi = {10.1007/s40477-024-00877-w}
}
EUVIMED