Minimal effect of scanning parameters on ultrasound shear wave elastography variability in tendons
Knee Surgery, Sports Traumatology, Arthroscopy · 2024
DOI: 10.1002/ksa.12325 ↗Kurz gefasst
Fachartikel in der Zeitschrift Knee Surgery, Sports Traumatology, Arthroscopy, 2024. Verfasst von Jennifer A Zellers und 9 weiteren Autor:innen. Beteiligte Einrichtung: Program in Physical Therapy, Washington University School of Medicine in St. Louis, St. Louis, Missouri, USA. Hauptthemen laut MEDLINE: Elasticity Imaging Techniques, Achilles Tendon. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIAMS NIH HHS, National Institutes of Health/National Institute of Arthritis and Musculoskeletal and Skin Diseases.
Abstract (Originalsprache)
Purpose
Ultrasound shear wave elastography has potential use in assessing tendon tissue; however, reducing measurement variability remains challenging. The primary purpose of this study was to identify the amount of variability accounted for by ultrasound parameter (frequency, harmonics and CrossXBeam) settings on shear wave speed at two testing sites.
Methods
Shear wave elastography images of the Achilles tendon were obtained from individuals with healthy tendons ( n = 28) at two testing sites with standardised image acquisition/postprocessing protocols. Images were acquired at a range of frequencies (7–15 MHz) with CrossXBeam (a filtering technique) and harmonics settings toggled on and off. Variance decomposition analysis was performed to identify the amount of variability in shear wave speed accounted for by scan acquisition settings and testing sites.
Results
Shear wave speed variance was primarily attributed to participants (56.87% of variance; residual error: 35%). All scanning parameters, testing site and interaction terms each accounted for less than 2.5% of the variance. A statistically significant, negative relationship was observed between shear wave speed and image quality ( p = 0.001) suggesting poor image quality yields higher shear wave speed estimates.
Conclusion
The findings of this study suggest that natural variation in Achilles tendon mechanics between individuals without tendon pathology accounts for most of the shear wave speed variability. Optimising image quality, which may be observed in higher frequencies, should be considered to improve shear wave speed estimation. Clinically, this study highlights the need to take multiple images, maintain consistent ultrasound settings when tracking patient progress over time and use caution when comparing raw values from tendon scans performed in different clinics with shear wave elastography. Level of Evidence Level III.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Program in Physical Therapy, Washington University School of Medicine in St. Louis, St. Louis, Missouri, USA.
- Department of Orthopaedic Surgery, Washington University School of Medicine in St. Louis, St. Louis, Missouri, USA.
- Mallinkrodt Institute of Radiology, Washington University School of Medicine in St. Louis, St. Louis, Missouri, USA.
- Department of Neurology, Washington University in St. Louis, St. Louis, Missouri, USA.
- Department of Orthopaedic Surgery, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
- Department of Biology, Washington University in St. Louis, St. Louis, Missouri, USA.
Förderung
- NIAMS NIH HHS
- National Institutes of Health/National Institute of Arthritis and Musculoskeletal and Skin Diseases
Zitieren
APA
Zellers, J. A., Li, R., Vaidya, R., Lohse, K., North, A., Cui, S., Houston, B., Chen, M., Zheng, J., & Baxter, J. (2024). Minimal effect of scanning parameters on ultrasound shear wave elastography variability in tendons. Knee Surgery, Sports Traumatology, Arthroscopy. https://doi.org/10.1002/ksa.12325
Vancouver
Zellers JA, Li R, Vaidya R, Lohse K, North A, Cui S, et al. Minimal effect of scanning parameters on ultrasound shear wave elastography variability in tendons. Knee Surgery, Sports Traumatology, Arthroscopy. 2024. doi:10.1002/ksa.12325
BibTeX
@article{Zellers2024,
author = {Zellers, Jennifer A and Li, Ran and Vaidya, Rachana and Lohse, Keith and North, Andrew and Cui, Stephane and Houston, Bryson and Chen, Menghan and Zheng, Jie and Baxter, Josh},
title = {Minimal effect of scanning parameters on ultrasound shear wave elastography variability in tendons},
journal = {Knee Surgery, Sports Traumatology, Arthroscopy},
year = {2024},
doi = {10.1002/ksa.12325}
}
EUVIMED