On the measurement of scaling factors in the RW3 plastic phantom during high energy electron beam dosimetry

Hamid Reza Baghani, Stefano Andreoli, Mostafa Robatjazi

Physical and Engineering Sciences in Medicine · 2023

DOI: 10.1007/s13246-022-01209-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physical and Engineering Sciences in Medicine, 2023. Verfasst von Hamid Reza Baghani und 2 weiteren Autor:innen. Beteiligte Einrichtung: Physics Department, Hakim Sabzevari University, Sabzevar, Iran. Hauptthemen laut MEDLINE: Plastics, Electrons. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 33 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Ionometric electron dosimetry inside water-equivalent plastic phantoms demands special considerations including determination of depth scaling and fluence scaling factors (cpl and hpl) to shift from in-phantom measurements to those relevant to water. This study evaluates these scaling factors for RW3 slab phantom and also introduces a new coefficient, k(RW3), for direct conversion from RW3 measurements to water without involving scaling factors. The RW3 solid phantom developed by the PTW Company was used and the corresponding scaling factors including cpl, hpl, and k(RW3) were measured for conventional electron energies of 4, 6, 9, 12, and 16 MeV. Separate measurements were performed in water and the RW3 slab phantom using the Advanced Markus chamber. The validity of the reported scaling factors was confirmed by comparing the direct and indirect percentage depth dose (PDD) measurements in water and in the RW3 phantom. The cpl values for the RW3 phantom were respectively equal to 0.915, 0.927, 0.934, 0.937, and 0.937 for 4, 6, 9, 12, and 16 MeV electron energies. The hpl and k(RW3) values were dependent on the depth of investigation and electron energy. Application of the cpl-hpl factors and k(RW3) coefficients to measured data inside the RW3 can reliably reproduce the measured PDD curves in water. The mean difference between the PDDs measured directly and indirectly in water and in the RW3 phantom was less than 1.2% in both approaches for PDD conversion (cpl-hpl coupling and the use of k(RW3)). The measured scaling factors and k(RW3) coefficients are sufficiently relevant to mimic water-based dosimetry results through indirect measurements inside the RW3 slab phantom. Nevertheless, employing k(RW3) is more straightforward than the cpl-hpl approach because it does not involve scaling and it is also less time-consuming.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PlasticsElectronsRadiometryRadiotherapy DosageWater

Stichwörter der Autor:innen

Depth scaling factorDosimetryElectron beamFluence scaling factorRW3 solid phantomk(RW3) coefficient

Substanzen

PlasticsWater

Autor:innen und Institutionen

  • Physics Department, Hakim Sabzevari University, Sabzevar, Iran.
  • Fisica Scanitaria, ASST Papa Giovanni XXIII, Bergamo, Italy. sandreoli@asst-pg23.it.
  • Medical Physics and Radiological Sciences Department, Sabzevar University of Medical Sciences, Sabzevar, Iran.

Zitieren

APA

Baghani, H. R., Andreoli, S., & Robatjazi, M. (2023). On the measurement of scaling factors in the RW3 plastic phantom during high energy electron beam dosimetry. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-022-01209-0

Vancouver

Baghani HR, Andreoli S, Robatjazi M. On the measurement of scaling factors in the RW3 plastic phantom during high energy electron beam dosimetry. Physical and Engineering Sciences in Medicine. 2023. doi:10.1007/s13246-022-01209-0

BibTeX
@article{Baghani2023,
  author = {Baghani, Hamid Reza and Andreoli, Stefano and Robatjazi, Mostafa},
  title = {On the measurement of scaling factors in the RW3 plastic phantom during high energy electron beam dosimetry},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2023},
  doi = {10.1007/s13246-022-01209-0}
}
Kostenlos starten →