Modeling scattered radiation from multi-leaf collimators (MLCs) to improve calculation accuracy of in-air output ratio
Australasian Physical & Engineering Sciences in Medicine · 2019-09 · 42 (3) · 719-731
DOI: 10.1007/s13246-019-00781-2 ↗Kurz gefasst
Fachartikel in der Zeitschrift Australasian Physical & Engineering Sciences in Medicine, online erschienen am 22.07.2019. Verfasst von So-Yeon Park und 3 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea. Hauptthemen laut MEDLINE: Air, Models, Theoretical, Scattering, Radiation. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea.
Abstract (Originalsprache)
This study aims to model an extra-focal source for the scattered radiation from multi-leaf collimators (MLCs), namely an MLC scatter source, and to correct in-air output ratio (Sc) calculated using the conventional dual source model (DSM) to achieve better accuracy of point dose calculation. To develop the MLC scatter source, a 6 MV photon beam from a Varian Clinac® iX linear accelerator with millennium 120 MLCs was used. It was assumed that the position for the MLC scatter source was located at the center of the MLC, consisting of line-based and area-based sources to consider the characteristics of the scattered radiation from the MLCs empirically. Based on the measured Sc values for MLC-defined fields, optimal parameters for the line-based and area-based sources were calculated using optimization process. For evaluation of proposed method, measurements were taken for various MLC-defined square and irregular fields. The Sc values calculated using the proposed MLC scatter source and conventional DSM were compared with the measured data. For MLC-defined square fields, the measured Sc values showed better agreement with those calculated using the MLC scatter source (the mean difference was - 0.03%) compared with those calculated using the DSM (the mean difference was 0.18%). For MLC-defined irregular fields, the maximum dose differences between measurements and calculations using the MLC scatter source and DSM were 0.54% and 1.45%, respectively. The developed MLC scatter source could improve the accuracy of Sc calculation for both square and irregular fields defined by MLCs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul, Republic of Korea.
- Department of Radiation Oncology, Veterans Health Service Medical Center, Seoul, Republic of Korea.
- Department of Radiation Oncology, Virginia Commonwealth University, Richmond, VA, USA.
- Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA.
- Radiation Protection and Emergency Preparedness Bureau, Nuclear Safety and Security Commission, Seoul, Republic of Korea. st.kim7@daum.net.
Förderung
- National Research Foundation of Korea
Zitieren
APA
Park, S. Y., Kim, S., Sung, W., & Kim, S. T. (2019). Modeling scattered radiation from multi-leaf collimators (MLCs) to improve calculation accuracy of in-air output ratio. Australasian Physical & Engineering Sciences in Medicine, 42(3), 719-731. https://doi.org/10.1007/s13246-019-00781-2
Vancouver
Park SY, Kim S, Sung W, Kim ST. Modeling scattered radiation from multi-leaf collimators (MLCs) to improve calculation accuracy of in-air output ratio. Australasian Physical & Engineering Sciences in Medicine. 2019;42(3):719-731. doi:10.1007/s13246-019-00781-2
BibTeX
@article{Park2019,
author = {Park, So-Yeon and Kim, Siyong and Sung, Wonmo and Kim, Sang-Tae},
title = {Modeling scattered radiation from multi-leaf collimators (MLCs) to improve calculation accuracy of in-air output ratio},
journal = {Australasian Physical & Engineering Sciences in Medicine},
year = {2019},
volume = {42},
number = {3},
pages = {719--731},
doi = {10.1007/s13246-019-00781-2}
}
EUVIMED