Validierungsstudie in der Zeitschrift Cardiovascular Engineering and Technology, 2025. Verfasst von Roberta Scuoppo und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Engineering, Università degli Studi di Palermo, Viale delle Scienze Ed.8, Palermo, Italy. Hauptthemen laut MEDLINE: Patient-Specific Modeling, Aortic Aneurysm, Thoracic, Models, Cardiovascular, Hemodynamics. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Università degli Studi di Palermo.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Computational modeling holds promise in predicting patient-specific outcomes and guiding clinical decision-making. The patient-specific model forming the basis of a digital twin can be considered biomedical software, thereby necessitating trust in its predictive accuracy.

Methods

This study applies the ASME V&V40 framework to demonstrate the credibility of patient-specific models of aneurysmal thoracic ascending aorta (ATAA) biomechanics. A comprehensive verification, validation, and uncertainty quantification process was performed to evaluate the accuracy of the patient-specific ATAA model.

Results

After implementing the ASME V&V40 standard, the verification errors on the model inputs (i.e., material parameters and hemodynamic variables) resulted in relative errors (RE) < 1%. Validation and its uncertainty quantification of the output aneurysm diameter response showed area metric errors below 5% in the majority of cases, highlighting the accuracy of the patient-specific ATAA model against the clinical comparator. Uncertainties in wall stress predictions due to model inputs were also quantified by probability density functions. Sensitivity analysis revealed that the unknown value of aneurysm wall thickness drives the model output at the highest extent.

Conclusions

These findings contribute to a standardized methodology for evaluating the credibility of patient-specific models, enhancing their utility in computer-based clinical decision support systems for managing patients with ATAAs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Patient-Specific ModelingAortic Aneurysm, ThoracicModels, CardiovascularHemodynamicsAorta, ThoracicComputed Tomography AngiographyHumansUncertaintyReproducibility of ResultsPredictive Value of TestsBiomechanical PhenomenaAortographyMaleStress, Mechanical

Stichwörter der Autor:innen

ASME VV40; aneurysm of ascending aortaComputational fluid dynamicsFinite-element analysisUncertainty quantification and validation

Autor:innen und Institutionen

  • Department of Engineering, Università degli Studi di Palermo, Viale delle Scienze Ed.8, Palermo, Italy.
  • Department of Engineering, Università degli Studi di Palermo, Viale delle Scienze Ed.8, Palermo, Italy. salvatore.pasta@unipa.it.
  • Department of Research, IRCCS ISMETT via Tricomi, Palermo, Italy. salvatore.pasta@unipa.it.

ORCID-Profile: Salvatore Pasta

Förderung

  • Università degli Studi di Palermo

Zitieren

APA

Scuoppo, R., Catalano, C., Crascì, F., & Pasta, S. (2025). Credibility Assessment of the Patient-Specific Modeling of the Aneurysmal Ascending Thoracic Aorta: Verification, Validation and Uncertainty Quantification. Cardiovascular Engineering and Technology. https://doi.org/10.1007/s13239-025-00801-1

Vancouver

Scuoppo R, Catalano C, Crascì F, Pasta S. Credibility Assessment of the Patient-Specific Modeling of the Aneurysmal Ascending Thoracic Aorta: Verification, Validation and Uncertainty Quantification. Cardiovascular Engineering and Technology. 2025. doi:10.1007/s13239-025-00801-1

BibTeX
@article{Scuoppo2025,
  author = {Scuoppo, Roberta and Catalano, Chiara and Crascì, Fabrizio and Pasta, Salvatore},
  title = {Credibility Assessment of the Patient-Specific Modeling of the Aneurysmal Ascending Thoracic Aorta: Verification, Validation and Uncertainty Quantification},
  journal = {Cardiovascular Engineering and Technology},
  year = {2025},
  doi = {10.1007/s13239-025-00801-1}
}
Kostenlos starten →