Simulation of thoracic endovascular aortic repair in a perfused patient-specific model of type B aortic dissection
International Journal of Computer Assisted Radiology and Surgery · 2024
DOI: 10.1007/s11548-024-03190-3 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Computer Assisted Radiology and Surgery, 2024. Verfasst von Lukas Mohl und 11 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiac Surgery, Heidelberg University Hospital, Heidelberg, Germany. l.mohl@stud.uni-heidelberg.de. Hauptthemen laut MEDLINE: Aortic Dissection, Endovascular Procedures, Aortic Aneurysm, Thoracic, Aorta, Thoracic. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Heidelberger Stiftung Chirurgie, Universitätsklinikum Heidelberg.
Abstract (Originalsprache)
Purpose
Complicated type B Aortic dissection is a severe aortic pathology that requires treatment through thoracic endovascular aortic repair (TEVAR). During TEVAR a stentgraft is deployed in the aortic lumen in order to restore blood flow. Due to the complicated pathology including an entry, a resulting dissection wall with potentially several re-entries, replicating this structure artificially has proven to be challenging thus far.
Methods
We developed a 3d printed, patient-specific and perfused aortic dissection phantom with a flexible dissection flap and all major branching vessels. The model was segmented from CTA images and fabricated out of a flexible material to mimic aortic wall tissue. It was placed in a pulsatile hemodynamic flow loop. Hemodynamics were investigated through pressure and flow measurements and doppler ultrasound imaging. Surgeons performed a TEVAR intervention including stentgraft deployment under fluoroscopic guidance.
Results
The flexible aortic dissection phantom was successfully incorporated in the hemodynamic flow loop, a systolic pressure of 112 mmHg and physiological flow of 4.05 L per minute was reached. Flow velocities were higher in true lumen with a up to 35.7 cm/s compared to the false lumen with a maximum of 13.3 cm/s, chaotic flow patterns were observed on main entry and reentry sights. A TEVAR procedure was successfully performed under fluoroscopy. The position of the stentgraft was confirmed using CTA imaging.
Conclusions
This perfused in-vitro phantom allows for detailed investigation of the complex inner hemodynamics of aortic dissections on a patient-specific level and enables the simulation of TEVAR procedures in a real endovascular operating environment. Therefore, it could provide a dynamic platform for future surgical training and research.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiac Surgery, Heidelberg University Hospital, Heidelberg, Germany. l.mohl@stud.uni-heidelberg.de.
- Department of Internal Medicine III, Department of Cardiology, Heidelberg University Hospital, Im Neuenheimer Feld 410, 69120, Heidelberg, Germany. l.mohl@stud.uni-heidelberg.de.
- German Centre for Cardiovascular Research (DZHK), Partner Site Heidelberg/Mannheim, Heidelberg, Germany. l.mohl@stud.uni-heidelberg.de.
- Department of Cardiac Surgery, Heidelberg University Hospital, Heidelberg, Germany.
- Department of Internal Medicine III, Department of Cardiology, Heidelberg University Hospital, Im Neuenheimer Feld 410, 69120, Heidelberg, Germany.
- German Centre for Cardiovascular Research (DZHK), Partner Site Heidelberg/Mannheim, Heidelberg, Germany.
Förderung
- Heidelberger Stiftung Chirurgie
- Universitätsklinikum Heidelberg
Zitieren
APA
Mohl, L., Karl, R., Hagedorn, M. N., Runz, A., Skornitzke, S., Toelle, M., Bergt, C. S., Hatzl, J., Uhl, C., Böckler, D., Meisenbacher, K., & Engelhardt, S. (2024). Simulation of thoracic endovascular aortic repair in a perfused patient-specific model of type B aortic dissection. International Journal of Computer Assisted Radiology and Surgery. https://doi.org/10.1007/s11548-024-03190-3
Vancouver
Mohl L, Karl R, Hagedorn MN, Runz A, Skornitzke S, Toelle M, et al. Simulation of thoracic endovascular aortic repair in a perfused patient-specific model of type B aortic dissection. International Journal of Computer Assisted Radiology and Surgery. 2024. doi:10.1007/s11548-024-03190-3
BibTeX
@article{Mohl2024,
author = {Mohl, Lukas and Karl, Roger and Hagedorn, Matthias N and Runz, Armin and Skornitzke, Stephan and Toelle, Malte and Bergt, C Soeren and Hatzl, Johannes and Uhl, Christian and Böckler, Dittmar and Meisenbacher, Katrin and Engelhardt, Sandy},
title = {Simulation of thoracic endovascular aortic repair in a perfused patient-specific model of type B aortic dissection},
journal = {International Journal of Computer Assisted Radiology and Surgery},
year = {2024},
doi = {10.1007/s11548-024-03190-3}
}
EUVIMED