Optimizing the Radiopacity of an Injectable Polymer on Fluoroscopy used for Treatment of Type II Endoleak After Endovascular Aneurysm Repair

Jeffrey R Nagel iD, Erik Groot Jebbink iD, Stefan P M Smorenburg iD, Arjan W J Hoksbergen iD, Rutger J Lely iD, Michel Versluis iD, Michel M P J Reijnen iD

Cardiovascular Engineering and Technology · 2025

DOI: 10.1007/s13239-025-00779-w ↗

Kurz gefasst

Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, 2025. Verfasst von Jeffrey R Nagel und 6 weiteren Autor:innen. Beteiligte Einrichtung: Multi-Modality Medical Imaging group, University of Twente, Enschede, The Netherlands. Hauptthemen laut MEDLINE: Endoleak, Endovascular Procedures, Blood Vessel Prosthesis Implantation, Tantalum. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Health~Holland.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Type II endoleaks (T2EL) are a common complication after endovascular aneurysm repair. AneuFix is a newly designed elastic polymer for T2EL. AneuFix contains tantalum for visualization during fluoroscopy, which is crucial for monitoring the polymer in the side branches. The purpose of this study was to find the lowest concentration tantalum that is sufficient for safe injection in the aneurysmal sac.

Methods

AneuFix polymer with tantalum concentrations between 0 and 30% was injected into endoleak phantoms, connected to a pulsatile flow setup and with a realistic background for fluoroscopy. Furthermore, the radiopacity was investigated on fluoroscopic systems from three different vendors, using static phantoms. Results from both the dynamic and static phantoms were qualitatively evaluated by 10 clinical experts.

Results

Concentrations of ≥ 20% tantalum were consistently detected within the first 5 mm after entering the side branch, with a corresponding contrast-to-noise ratio of 2.23 ± 0.21. Furthermore, sufficient detectability scores (of at least 3 out of 5) were given to ≥ 15% tantalum. Significant differences were found in detectability scores on different fluoroscopic systems, using the default lowest-radiation-dose scan protocol for each system.

Conclusions

This study showed that tantalum concentrations ≥ 20% are consistently detected on fluoroscopy in the specified region. Compared to the original 30%, this would reduce imaging artifacts from high attenuation and scattering on follow-up imaging, while retaining sufficient detectability during injection. However, because of differences in fluoroscopic systems and scan protocols between hospitals, the combination of tantalum concentration and scan protocol should be optimized for each clinical setting.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

EndoleakEndovascular ProceduresBlood Vessel Prosthesis ImplantationTantalumContrast MediaRadiography, InterventionalPolymersAortographyAortic Aneurysm, AbdominalBlood Vessel ProsthesisHumansFluoroscopyPhantoms, ImagingRadiation DosagePredictive Value of TestsEndovascular Aneurysm Repair

Stichwörter der Autor:innen

Abdominal aortic aneurysmEndoleakInjectable embolic agentRadiopacityTantalum

Substanzen

TantalumContrast MediaPolymers

Autor:innen und Institutionen

  • Multi-Modality Medical Imaging group, University of Twente, Enschede, The Netherlands.
  • Physics of Fluids group, University of Twente, Enschede, The Netherlands.
  • Multi-Modality Medical Imaging group, University of Twente, Enschede, The Netherlands. e.grootjebbink@utwente.nl.
  • Department of Surgery, Rijnstate, Arnhem, The Netherlands. e.grootjebbink@utwente.nl.
  • Department of Surgery, Amsterdam UMC location Vrije Universiteit, Amsterdam, The Netherlands.
  • Department of Radiology, Amsterdam UMC location Vrije Universiteit, Amsterdam, The Netherlands.

ORCID-Profile: Jeffrey R Nagel, Erik Groot Jebbink, Stefan P M Smorenburg, Arjan W J Hoksbergen, Rutger J Lely, Michel Versluis, Michel M P J Reijnen

Förderung

  • Health~Holland

Zitieren

APA

Nagel, J. R., Groot Jebbink, E., Smorenburg, S. P. M., Hoksbergen, A. W. J., Lely, R. J., Versluis, M., & Reijnen, M. M. P. J. (2025). Optimizing the Radiopacity of an Injectable Polymer on Fluoroscopy used for Treatment of Type II Endoleak After Endovascular Aneurysm Repair. Cardiovascular Engineering and Technology. https://doi.org/10.1007/s13239-025-00779-w

Vancouver

Nagel JR, Groot Jebbink E, Smorenburg SPM, Hoksbergen AWJ, Lely RJ, Versluis M, et al. Optimizing the Radiopacity of an Injectable Polymer on Fluoroscopy used for Treatment of Type II Endoleak After Endovascular Aneurysm Repair. Cardiovascular Engineering and Technology. 2025. doi:10.1007/s13239-025-00779-w

BibTeX
@article{Nagel2025,
  author = {Nagel, Jeffrey R and Groot Jebbink, Erik and Smorenburg, Stefan P M and Hoksbergen, Arjan W J and Lely, Rutger J and Versluis, Michel and Reijnen, Michel M P J},
  title = {Optimizing the Radiopacity of an Injectable Polymer on Fluoroscopy used for Treatment of Type II Endoleak After Endovascular Aneurysm Repair},
  journal = {Cardiovascular Engineering and Technology},
  year = {2025},
  doi = {10.1007/s13239-025-00779-w}
}
Kostenlos starten →