Performance Comparison of Centered and Tilted Blunt and Lighthouse Tip Cannulae for Drainage in Extracorporeal Life Support
Cardiovascular Engineering and Technology · 2025
DOI: 10.1007/s13239-024-00770-x ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Cardiovascular Engineering and Technology, 2025. Verfasst von Federico Rorro und 2 weiteren Autor:innen. Beteiligte Einrichtung: FLOW, Department of Engineering Mechanics, KTH Royal Institute of Technology, Stockholm, Sweden. rorro@kth.se. Hauptthemen laut MEDLINE: Extracorporeal Membrane Oxygenation, Drainage, Cannula. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Research Council, Royal Institute of Technology.
Abstract (Originalsprache)
Introduction
Extracorporeal membrane oxygenation is a lifesaving treatment for patients with refractory acute respiratory, circulatory, or combined cardiopulmonary failure. The patient is cannulated with one or two cannulae for drainage and reinfusion of blood. Blood is drained from the patient, pumped through a membrane lung for oxygenation and returned to the patient. Treatment efficiency depends on correct cannula positioning and interactions between drainage and reinfusion cannula.
Methods
An experimental setup was built to study the isolated drainage performance of 24 Fr rigid models of a blunt and lighthouse tip cannula, both when centered and when tilted towards the vessel wall. Planar particle image velocimetry was used to investigate the flow field with water as the fluid medium.
Results
For similar flow configuration, higher shear stresses were recorded in the blunt tip rather than lighthouse tip cannula. Moreover, in the lighthouse tip cannula, side-holes furthest from the tip ( proximal side-holes) had the highest drainage. Results did not change substantially when the cannula was tilted towards the vessel wall.
Conclusions
The effective drainage point of the lighthouse tip cannula was located near the proximal side-holes. Lower shear stresses were recorded in the lighthouse tip cannula when compared with the blunt tip cannula, for all considered flow rate ratios and cannula positions.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- FLOW, Department of Engineering Mechanics, KTH Royal Institute of Technology, Stockholm, Sweden. rorro@kth.se.
- ECMO Centre Karolinska, Pediatric Perioperative Medicine and Intensive Care, Karolinska University Hospital, Stockholm, Sweden.
- Department of Physiology and Pharmacology, Karolinska Institutet, Stockholm, Sweden.
- FLOW, Department of Engineering Mechanics, KTH Royal Institute of Technology, Stockholm, Sweden.
ORCID-Profile: Federico Rorro
Förderung
- European Research Council
- Royal Institute of Technology
Zitieren
APA
Rorro, F., Broman, L. M., & Prahl Wittberg, L. (2025). Performance Comparison of Centered and Tilted Blunt and Lighthouse Tip Cannulae for Drainage in Extracorporeal Life Support. Cardiovascular Engineering and Technology. https://doi.org/10.1007/s13239-024-00770-x
Vancouver
Rorro F, Broman LM, Prahl Wittberg L. Performance Comparison of Centered and Tilted Blunt and Lighthouse Tip Cannulae for Drainage in Extracorporeal Life Support. Cardiovascular Engineering and Technology. 2025. doi:10.1007/s13239-024-00770-x
BibTeX
@article{Rorro2025,
author = {Rorro, Federico and Broman, Lars Mikael and Prahl Wittberg, Lisa},
title = {Performance Comparison of Centered and Tilted Blunt and Lighthouse Tip Cannulae for Drainage in Extracorporeal Life Support},
journal = {Cardiovascular Engineering and Technology},
year = {2025},
doi = {10.1007/s13239-024-00770-x}
}
EUVIMED