EUVIMEDEUVIMED entdecken
Validierungsstudie

Computer Aided Intracranial Aneurysm Treatment Based on 2D/3D Mapping, Virtual Deployment and Online Distal Marker Detection

Nicolas Dazeo iD, José Ignacio Orlando, Camila García, Romina Muñoz, Laura Obrado, Hector Fernandez, Jordi Blasco, Luis San Román, Juan M Macho, Andreas Ding, Raphael Utz, Ignacio Larrabide

Cardiovascular Engineering and Technology · 2024

DOI: 10.1007/s13239-024-00745-y ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Cardiovascular Engineering and Technology, 2024. Verfasst von Nicolas Dazeo und 11 weiteren Autor:innen. Beteiligte Einrichtung: Yatiris Research Group, PLADEMA Institute, CONICET-UNICEN, Campus Universitario, Tandil, Argentina. ndazeo@pladema.exa.unicen.edu.ar. Hauptthemen laut MEDLINE: Intracranial Aneurysm, Imaging, Three-Dimensional, Radiographic Image Interpretation, Computer-Assisted, Endovascular Procedures. Die Arbeit nennt 40 Literaturangaben. Gefördert von: Eurostars.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To introduce a computational tool for peri-interventional intracranial aneurysm treatment guidance that maps preoperative planning information from simulation onto real-time X-Ray imaging. METHODS: Preoperatively, multiple flow diverter (FD) devices are simulated based on the 3D mesh of the vessel to treat, to choose the optimal size and location. In the peri-operative stage, this 3D information is aligned and mapped to the continuous 2D-X-Ray scan feed from the operating room. The current flow diverter position in the 3D model is estimated by automatically detecting the distal FD marker locations and mapping them to the treated vessel. This allows to visually assess the possible outcome of releasing the device at the current position, and compare it with the one chosen pre-operatively. RESULTS: The full pipeline was validated using retrospectively collected biplane images from four different patients (5 3D-DSA datasets in total). The distal FD marker detector obtained an average F1-score of 0.67 ( ± 0.224 ) in 412 2D-X-Ray scans. After aligning 3D-DSA + 2D-X-Ray datasets, the average difference between simulated and deployed positions was 0.832 mm ( ± 0.521 mm). Finally, we qualitatively show that the proposed approach is able to display the current location of the FD compared to their pre-operatively planned position. CONCLUSIONS: The proposed method allows to support the FD deployment procedure by merging and presenting preoperative simulation information to the interventionists, aiding them to make more accurate and less risky decisions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intracranial AneurysmImaging, Three-DimensionalRadiographic Image Interpretation, Computer-AssistedEndovascular ProceduresPatient-Specific ModelingCerebral AngiographyHumansRetrospective StudiesPredictive Value of TestsAngiography, Digital SubtractionReproducibility of ResultsTreatment OutcomeModels, CardiovascularProsthesis DesignStents

Stichwörter der Autor:innen

2D-X-ray3D-DSAAlignmentCerebral aneurysmReal timeTreatment guidance

Autor:innen und Institutionen

  • Yatiris Research Group, PLADEMA Institute, CONICET-UNICEN, Campus Universitario, Tandil, Argentina. ndazeo@pladema.exa.unicen.edu.ar.
  • Yatiris Research Group, PLADEMA Institute, CONICET-UNICEN, Campus Universitario, Tandil, Argentina.
  • Mentice S. L., Barcelona, Spain.
  • Department of Neuroradiology, Hospital Clinic, Barcelona, Spain.
  • Acandis, GmbH, Pforzheim, Germany.

ORCID-Profile: Nicolas Dazeo

Förderung

  • Eurostars

Zitieren

APA

Dazeo, N., Orlando, J. I., García, C., Muñoz, R., Obrado, L., Fernandez, H., Blasco, J., Román, L. S., Macho, J. M., Ding, A., Utz, R., & Larrabide, I. (2024). Computer Aided Intracranial Aneurysm Treatment Based on 2D/3D Mapping, Virtual Deployment and Online Distal Marker Detection. Cardiovascular Engineering and Technology. https://doi.org/10.1007/s13239-024-00745-y

Vancouver

Dazeo N, Orlando JI, García C, Muñoz R, Obrado L, Fernandez H, et al. Computer Aided Intracranial Aneurysm Treatment Based on 2D/3D Mapping, Virtual Deployment and Online Distal Marker Detection. Cardiovascular Engineering and Technology. 2024. doi:10.1007/s13239-024-00745-y

BibTeX
@article{Dazeo2024,
  author = {Dazeo, Nicolas and Orlando, José Ignacio and García, Camila and Muñoz, Romina and Obrado, Laura and Fernandez, Hector and Blasco, Jordi and Román, Luis San and Macho, Juan M and Ding, Andreas and Utz, Raphael and Larrabide, Ignacio},
  title = {Computer Aided Intracranial Aneurysm Treatment Based on 2D/3D Mapping, Virtual Deployment and Online Distal Marker Detection},
  journal = {Cardiovascular Engineering and Technology},
  year = {2024},
  doi = {10.1007/s13239-024-00745-y}
}
Kostenlos starten →