EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

In Silico Evaluation of a Self-powered Venous Ejector Pump for Fontan Patients

Reza Rasooli iD, Knut Erik Teigen Giljarhus, Aksel Hiorth, Ingunn Westvik Jolma, Jan Ludvig Vinningland, Charlotte de Lange, Henrik Brun, Henrik Holmstrom

Cardiovascular Engineering and Technology · 2023

DOI: 10.1007/s13239-023-00663-5 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, 2023. Verfasst von Reza Rasooli und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Energy Resources, Faculty of Science and Technology, University of Stavanger, 4036, Stavanger, Norway. reza.rasooli@uis.no. Hauptthemen laut MEDLINE: Fontan Procedure, Heart-Assist Devices, Heart Defects, Congenital. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 70 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Equinor, EkstraStiftelsen Helse og Rehabilitering, University of Stavanger & Stavanger University Hospital.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The Fontan circulation carries a dismal prognosis in the long term due to its peculiar physiology and lack of a subpulmonic ventricle. Although it is multifactorial, elevated IVC pressure is accepted to be the primary cause of Fontan's high mortality and morbidity. This study presents a self-powered venous ejector pump (VEP) that can be used to lower the high IVC venous pressure in single-ventricle patients.

Methods

A self-powered venous assist device that exploits the high-energy aortic flow to lower IVC pressure is designed. The proposed design is clinically feasible, simple in structure, and is powered intracorporeally. The device's performance in reducing IVC pressure is assessed by conducting comprehensive computational fluid dynamics simulations in idealized total cavopulmonary connections with different offsets. The device was finally applied to complex 3D reconstructed patient-specific TCPC models to validate its performance.

Results

The assist device provided a significant IVC pressure drop of more than 3.2 mm Hg in both idealized and patient-specific geometries, while maintaining a high systemic oxygen saturation of more than 90%. The simulations revealed no significant caval pressure rise (< 0.1 mm Hg) and sufficient systemic oxygen saturation (> 84%) in the event of device failure, demonstrating its fail-safe feature.

Conclusions

A self-powered venous assist with promising in silico performance in improving Fontan hemodynamics is proposed. Due to its passive nature, the device has the potential to provide palliation for the growing population of patients with failing Fontan.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Fontan ProcedureHeart-Assist DevicesHeart Defects, CongenitalHumansPulmonary ArteryHemodynamicsVena Cava, InferiorHeart VentriclesModels, Cardiovascular

Stichwörter der Autor:innen

Computational fluid dynamicsCongenital heart diseaseFontan hemodynamicsFontan venous assistSingle-ventricle physiologySystemic venous hypertension

Autor:innen und Institutionen

  • Department of Energy Resources, Faculty of Science and Technology, University of Stavanger, 4036, Stavanger, Norway. reza.rasooli@uis.no.
  • Department of Mechanical and Structural Engineering and Materials Science, University of Stavanger, 4036, Stavanger, Norway.
  • Department of Energy Resources, Faculty of Science and Technology, University of Stavanger, 4036, Stavanger, Norway.
  • Department of Chemistry, Bioscience and Environmental Engineering, University of Stavanger, 4036, Stavanger, Norway.
  • Norwegian Research Center (NORCE), Oslo, Norway.
  • Department of Paediatric Radiology, Sahlgrenska University Hospital, Gothenburg, Sweden.

ORCID-Profile: Reza Rasooli

Förderung

  • Equinor
  • EkstraStiftelsen Helse og Rehabilitering
  • University of Stavanger & Stavanger University Hospital

Zitieren

APA

Rasooli, R., Giljarhus, K. E. T., Hiorth, A., Jolma, I. W., Vinningland, J. L., de Lange, C., Brun, H., & Holmstrom, H. (2023). In Silico Evaluation of a Self-powered Venous Ejector Pump for Fontan Patients. Cardiovascular Engineering and Technology. https://doi.org/10.1007/s13239-023-00663-5

Vancouver

Rasooli R, Giljarhus KET, Hiorth A, Jolma IW, Vinningland JL, de Lange C, et al. In Silico Evaluation of a Self-powered Venous Ejector Pump for Fontan Patients. Cardiovascular Engineering and Technology. 2023. doi:10.1007/s13239-023-00663-5

BibTeX
@article{Rasooli2023,
  author = {Rasooli, Reza and Giljarhus, Knut Erik Teigen and Hiorth, Aksel and Jolma, Ingunn Westvik and Vinningland, Jan Ludvig and de Lange, Charlotte and Brun, Henrik and Holmstrom, Henrik},
  title = {In Silico Evaluation of a Self-powered Venous Ejector Pump for Fontan Patients},
  journal = {Cardiovascular Engineering and Technology},
  year = {2023},
  doi = {10.1007/s13239-023-00663-5}
}
Kostenlos starten →