In vitro investigation of axial mechanical support devices implanted in the novel convergent cavopulmonary connection Fontan
European Journal of Cardio-Thoracic Surgery · 02.01.2024 · 65 (1)
DOI: 10.1093/ejcts/ezad413 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Cardio-Thoracic Surgery, online erschienen am 05.01.2024. Verfasst von Vincent Cleveland und 11 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Children's National Hospital, Washington, DC, USA. Hauptthemen laut MEDLINE: Fontan Procedure, Heart Defects, Congenital. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: NHLBI NIH HHS, American Heart Association National Innovation.
Abstract (Originalsprache)
OBJECTIVES The 2 opposing inflows and 2 outflows in a total cavopulmonary connection make mechanical circulatory support (MCS) extremely challenging. We have previously reported a novel convergent cavopulmonary connection (CCPC) Fontan design that improves baseline characteristics and provides a single inflow and outflow, thus simplifying MCS. This study aims to assess the feasibility of MCS of this novel configuration using axial flow pumps in an in vitro benchtop model. METHODS Three-dimensional segmentations of 12 single-ventricle patients (body surface area 0.5–1.75 m2) were generated from cardiovascular magnetic resonance images. The CCPC models were designed by connecting the inferior vena cava and superior vena cava to a shared conduit ascending to the pulmonary arteries, optimized in silico. The 12 total cavopulmonary connection and their corresponding CCPC models underwent in vitro benchtop characterization. Two MCS devices were used, the Impella RP® and the PediPump. RESULTS MCS successfully and symmetrically reduced the pressure in both vena cavae by >20 mmHg. The devices improved the hepatic flow distribution balance of all CCPC models (Impella RP®P = 0.045, PediPump P = 0.055). CONCLUSIONS The CCPC Fontan design provides a feasible MCS solution for a failing Fontan by balancing hepatic flow distribution and symmetrically decompressing the central venous pressure. Cardiac index may also improve with MCS. Additional studies are needed to evaluate this concept for managing Fontan failure.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Cardiology, Children's National Hospital, Washington, DC, USA.
- Department of Mechanical Engineering, Johns Hopkins University, Baltimore, MD, USA.
- Department of Mechanical Engineering, Texas Tech University, Lubbock, TX, USA.
- Biostatistical Design and Analysis Center, Clinical and Translational Science Institute, University of Minnesota, Minneapolis, MN, USA.
- Division of Pediatric Cardiology, University of Pittsburgh Medical Center, Pittsburgh, PA, USA.
- Department of Pediatric Cardiac Surgery, M Health Fairview University of Minnesota, Minneapolis, MN, USA.
ORCID-Profile: Pranava Sinha
Förderung
- NHLBI NIH HHS
- American Heart Association National Innovation
Zitieren
APA
Cleveland, V., Contento, J., Mass, P., Hardikar, P., Wu, Q., Liu, X., Aslan, S., Loke, Y. H., Krieger, A., Lunos, S., Olivieri, L., & Sinha, P. (2024). In vitro investigation of axial mechanical support devices implanted in the novel convergent cavopulmonary connection Fontan. European Journal of Cardio-Thoracic Surgery, 65(1). https://doi.org/10.1093/ejcts/ezad413
Vancouver
Cleveland V, Contento J, Mass P, Hardikar P, Wu Q, Liu X, et al. In vitro investigation of axial mechanical support devices implanted in the novel convergent cavopulmonary connection Fontan. European Journal of Cardio-Thoracic Surgery. 2024;65(1). doi:10.1093/ejcts/ezad413
BibTeX
@article{Cleveland2024,
author = {Cleveland, Vincent and Contento, Jacqueline and Mass, Paige and Hardikar, Priyanka and Wu, Qiyuan and Liu, Xiaolong and Aslan, Seda and Loke, Yue-Hin and Krieger, Axel and Lunos, Scott and Olivieri, Laura and Sinha, Pranava},
title = {In vitro investigation of axial mechanical support devices implanted in the novel convergent cavopulmonary connection Fontan},
journal = {European Journal of Cardio-Thoracic Surgery},
year = {2024},
volume = {65},
number = {1},
doi = {10.1093/ejcts/ezad413}
}
EUVIMED