Hemodynamic effects of pulsatile frequency of right ventricular assist device (RVAD) on pulmonary perfusion: a simulation study

Fan Meng, Yuanfei Zhu, Ming Yang

Medical & Biological Engineering & Computing · 2024

DOI: 10.1007/s11517-024-03174-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical & Biological Engineering & Computing, 2024. Verfasst von Fan Meng und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Instrument Science and Engineering, Shanghai Jiao Tong University, Shanghai, China. Hauptthemen laut MEDLINE: Heart-Assist Devices, Hemodynamics, Pulmonary Artery, Pulsatile Flow. Die Arbeit nennt 47 Literaturangaben. Gefördert von: National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Right ventricular assist devices (RVADs) have been extensively used to provide hemodynamic support for patients with end-stage right heart (RV) failure. However, conventional in-parallel RVADs can lead to an elevation of pulmonary artery (PA) pressure, consequently increasing the right ventricular (RV) afterload, which is unfavorable for the relaxation of cardiac muscles and reduction of valve complications. The aim of this study is to investigate the hemodynamic effects of the pulsatile frequency of the RVAD on pulmonary artery. Firstly, a mathematical model incorporating heart, systemic circulation, pulmonary circulation, and RVAD is developed to simulate the cardiovascular system. Subsequently, the frequency characteristics of the pulmonary circulation system are analyzed, and the calculated results demonstrate that the pulsatile frequency of the RVAD has a substantive impact on the pulmonary artery pressure. Finally, to verify the analysis results, the hemodynamic effects of the pulsatile frequency of the RVAD on pulmonary artery are compared under diffident support modes. It is found that the pulmonary artery pressure decreases by approximately 6% when the pulsatile frequency changes from 1 to 3 Hz. The increased pulsatile frequency of RA-PA support mode may facilitate the opening of the pulmonary valve, while the RV-PA support mode can more effectively reduce the load of RV. This work provides a useful method to decrease the pulmonary artery pressure during the RVAD supports and may be beneficial for improving myocardial function in patients with end-stage right heart failure, especially those with pulmonary hypertension.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart-Assist DevicesHemodynamicsPulmonary ArteryPulsatile FlowModels, CardiovascularComputer SimulationHumansPulmonary CirculationHeart Ventricles

Stichwörter der Autor:innen

Pulmonary circulationPulmonary hypertensionPulsatile frequencyRight ventricular assist devices

Autor:innen und Institutionen

  • Department of Instrument Science and Engineering, Shanghai Jiao Tong University, Shanghai, China.
  • Department of Instrument Science and Engineering, Shanghai Jiao Tong University, Shanghai, China. zhuyuanfei@sjtu.edu.cn.
  • Department of Instrument Science and Engineering, Shanghai Jiao Tong University, Shanghai, China. myang@sjtu.edu.cn.

Förderung

  • National Natural Science Foundation of China

Zitieren

APA

Meng, F., Zhu, Y., & Yang, M. (2024). Hemodynamic effects of pulsatile frequency of right ventricular assist device (RVAD) on pulmonary perfusion: a simulation study. Medical & Biological Engineering & Computing. https://doi.org/10.1007/s11517-024-03174-0

Vancouver

Meng F, Zhu Y, Yang M. Hemodynamic effects of pulsatile frequency of right ventricular assist device (RVAD) on pulmonary perfusion: a simulation study. Medical & Biological Engineering & Computing. 2024. doi:10.1007/s11517-024-03174-0

BibTeX
@article{Meng2024,
  author = {Meng, Fan and Zhu, Yuanfei and Yang, Ming},
  title = {Hemodynamic effects of pulsatile frequency of right ventricular assist device (RVAD) on pulmonary perfusion: a simulation study},
  journal = {Medical & Biological Engineering & Computing},
  year = {2024},
  doi = {10.1007/s11517-024-03174-0}
}
Kostenlos starten →