Quantifying Cardiothoracic Variation with Posture and Respiration to Inform Cardiac Device Design

Geeth A Kondaveeti, Varun A Bhatia, Ryan P Lahm, Megan L Harris, James P Gaewsky, F Scott Gayzik, John F Greenhalgh, Craig A Hamilton, R Brandon Stacey, Ashley A Weaver iD

Cardiovascular Engineering and Technology · 2023-02 · 14 (1) · 13-24

DOI: 10.1007/s13239-022-00631-5 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, online erschienen am 26.05.2022. Verfasst von Geeth A Kondaveeti und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, Wake Forest School of Medicine, 575 N. Patterson Ave. Suite 530, Winston-Salem, NC, 27101, USA. Hauptthemen laut MEDLINE: Defibrillators, Implantable. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIA NIH HHS, Medtronic, National Institute on Aging.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: With extravascular implantable cardioverter defibrillator leads placed beneath the sternum, it is important to quantify heart motion relative to the rib cage with postural changes and respiration. METHODS: MRI scans from five males and five females were collected in upright and supine postures at end inspiration [n = 10 each]. Left and right decubitus [n = 8 each] and prone [n = 5] MRIs at end inspiration and supine MRIs at end expiration [n = 5] were collected on a subset. Four cardiothoracic measurements, six cardiac measurements, and six cardiac landmarks were collected to measure changes across different postures and stages of respiration. RESULTS: The relative location of the LV apex to the nearest intercostal space was significantly different between the supine and decubitus postures (average ± SD difference: - 15.7 ± 11.4 mm; p < 0.05). The heart centroid to xipho-sternal junction distance was 9.7 ± 7.9 mm greater in the supine posture when compared to the upright posture (p < 0.05). Cardiac landmark motion in the lateral direction was largest due to postural movement (range 23-50 mm) from the left decubitus to the right decubitus posture, and less influenced by respiration (5-17 mm). Caudal-cranial displacement was generally larger due to upright posture (13-23 mm caudal) and inspiration (7-20 mm cranial). CONCLUSIONS: This study demonstrates that the location of the heart with respect to the rib cage varies with posture and respiration. The gravitational effects of postural shifts on the heart position are roughly 2-3 times larger than the effects of normal respiration.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Defibrillators, ImplantableMaleFemaleHumansRespirationHeartPosture

Stichwörter der Autor:innen

Extravascular implantable cardioverter defibrillator (EV ICD)Landmark motionNonischemic cardiomyopathySudden cardiac death

Autor:innen und Institutionen

  • Department of Biomedical Engineering, Wake Forest School of Medicine, 575 N. Patterson Ave. Suite 530, Winston-Salem, NC, 27101, USA.
  • Cardiac Rhythm Management, Medtronic Inc., 8200 Coral Sea Street NE, Mounds View, MN, 55112, USA.
  • Elemance LLC, 3540 Clemmons Rd #127, Clemmons, NC, 27012, USA.
  • FONAR Corporation, 110 Marcus Drive, Melville, NY, 11747, USA.
  • Department of Cardiology, Wake Forest University School of Medicine, Winston-Salem, NC, 27157, USA.
  • Department of Biomedical Engineering, Wake Forest School of Medicine, 575 N. Patterson Ave. Suite 530, Winston-Salem, NC, 27101, USA. asweaver@wakehealth.edu.

ORCID-Profile: Ashley A Weaver

Förderung

  • NIA NIH HHS
  • Medtronic
  • National Institute on Aging

Zitieren

APA

Kondaveeti, G. A., Bhatia, V. A., Lahm, R. P., Harris, M. L., Gaewsky, J. P., Gayzik, F. S., Greenhalgh, J. F., Hamilton, C. A., Stacey, R. B., & Weaver, A. A. (2023). Quantifying Cardiothoracic Variation with Posture and Respiration to Inform Cardiac Device Design. Cardiovascular Engineering and Technology, 14(1), 13-24. https://doi.org/10.1007/s13239-022-00631-5

Vancouver

Kondaveeti GA, Bhatia VA, Lahm RP, Harris ML, Gaewsky JP, Gayzik FS, et al. Quantifying Cardiothoracic Variation with Posture and Respiration to Inform Cardiac Device Design. Cardiovascular Engineering and Technology. 2023;14(1):13-24. doi:10.1007/s13239-022-00631-5

BibTeX
@article{Kondaveeti2023,
  author = {Kondaveeti, Geeth A and Bhatia, Varun A and Lahm, Ryan P and Harris, Megan L and Gaewsky, James P and Gayzik, F Scott and Greenhalgh, John F and Hamilton, Craig A and Stacey, R Brandon and Weaver, Ashley A},
  title = {Quantifying Cardiothoracic Variation with Posture and Respiration to Inform Cardiac Device Design},
  journal = {Cardiovascular Engineering and Technology},
  year = {2023},
  volume = {14},
  number = {1},
  pages = {13--24},
  doi = {10.1007/s13239-022-00631-5}
}
Kostenlos starten →